How it works · DUV-001
DUV-001: a blue-dye dilution made by the robot
Which part of the experiment does what, where, why, on what evidence, and what it cannot do. Every number the experiment reports is traced to the parts it rests on; a number is no stronger than its weakest part, nor than the unmeasured values it rests on.
The same as a report, with the equations, the evidence audit and the positioning: the scientific basis.
The workflow
11 of its 14 parts rest on exact laws, published data, standards or checks against independent data; 2 calibrated, 1 estimate. Each estimate or placeholder names its source and how it would be measured.
- exact law 2
- established 9
- calibrated 2
- estimate 1
The evidence scale: what each word means
measured here measured in this project on physical equipment (none yet)
exact law an exact law or a definition: conservation of matter, molar masses from atomic weights, the relative centrifugal force, a dilution's mass balance
checked published data or a published method, checked here against independent published measurements it was not built from
established a published method, standard or instrument specification, used inside its stated domain; not checked here against independent data
calibrated a parameter fitted to named observations: valid in the fitted domain only
estimate a value or range from a datasheet, a publication or practice, not measured on this bench: a Type B estimate (GUM, JCGM 100, 4.3), its source named and how to measure it said
placeholder a value with no source yet, stated so that a measurement can replace it; how to measure it is said
not modelled absent from the model: a stated limitation
test scenario the simulated laboratory the instruments read: by design not the model, so not scored; it tests the procedure against a laboratory that differs from the model
Each step's parts
The parts that act in each step, with their evidence. Click a part for its principle, its evidence and its limits.
Plan
the recipe, the engine's plan and its prediction
The recipe meets the goal (G1) exact lawThe stock's strength estimateThe engine's plan (G2) calibratedThe prediction and its uncertainty (G4) establishedgates G1, G2, G4
Execute
the robot makes the tube
The robot (VialBot) and its notebook (G3) establishedThe virtual laboratory (the hidden truth) test scenariogates G3
Verify
the balance and the spectrophotometer
Analytical balance establishedUV-Vis spectrophotometer establishedBrilliant Blue FCF's absorptivity, calibrated calibratedThe measured strength (G5) establishedgates G5
Decide
the rule; the cause; the next tube
The decision (D) establishedThe stock standardised establishedgates D
Report
the report, the export, the evidence package
Report and export establishedThe evidence package (G6) exact lawgates G6
What each reported number rests on
| Number | Governed by | Weakest part | Uncertainty resting on estimates or placeholders | Verdict | Influences not measured here |
|---|---|---|---|---|---|
| the predicted strength (G4) | The recipe meets the goal (G1), The engine's plan (G2), The prediction and its uncertainty (G4), The stock's strength | estimate | 86 % the stock's strength (estimated ±0.5 %) 86 % (estimate); the transfers (virtual calibration records) 14 % (established) | uncertain | the stock's estimated ±0.5 % and the virtual calibration records |
| the tube's fill by weighing | Analytical balance | established | — | supported | the balance's estimated noise and a perfect tare |
| the measured strength (G5) | UV-Vis spectrophotometer, Brilliant Blue FCF's absorptivity, calibrated, The measured strength (G5) | calibrated | 85 % absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established) | uncertain | the calibrated absorptivity's uncertainty (mostly the lot's assay, an estimate), the estimated path tolerance and stray fraction |
| the decision (D) | UV-Vis spectrophotometer, Brilliant Blue FCF's absorptivity, calibrated, The measured strength (G5), The decision (D) | calibrated | 85 % absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established) | uncertain | as the measured strength |
| the cause found (the stock's strength) | The measured strength (G5), Brilliant Blue FCF's absorptivity, calibrated, The stock standardised | calibrated | 85 % absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established) | uncertain | as the measured strength |
Rule: a quantity is supported when its weakest governing part is at least calibrated and estimates or placeholders carry less than half of its uncertainty; uncertain when a governing part is only an estimate or a placeholder, or when they carry half or more of its uncertainty; unsupported when a part it needs is not modelled.
What the run actually used
the reference run, scenario 'nominal', seed 4242; the dye bench (no chemistry beyond dilution and Beer-Lambert) · The stock is what its label says · The stock is 5 % weaker than its label. The branch each part took in this run: never a value of the hidden laboratory.
- UV-Vis spectrophotometer: read at 630, 530 nm against a water blank; each read tube also scanned 400-750 nm
- Brilliant Blue FCF's absorptivity, calibrated: the lot's calibrated absorptivity (a virtual calibration record) converts the reading into a strength
Tube 1 (V01)
- plan: The engine can make the goal's recipe on this robot: blue dye by 24 pump pulses; water by 4 pipette strokes; 2 tips; mixing 3 × 300 µL.
- prediction: P(within the tolerance) = 0.9988; the stock's estimated ±0.5 % carries 86 % of the uncertainty
- readings: sample [1.303, 1.306, 1.302], blank [0.002, 0.002, -0.002] (A at the analytical wavelength)
- measurement: PASS: Blank at zero PASS; Inside the working range PASS; Enough liquid for the cell PASS; Mixed before reading PASS
- fill: the balance: 1,200.4 uL weighed against 1,201.4 uL predicted (zeta -0.74, agrees)
- decision: accept
What the instruments showed
Each reading of the run set against the published data or the specification that gives it meaning, as an analyst would plot it; the numbers are under each figure. The readings are simulated; the curves are published data or the model's.
- Brilliant Blue FCF at the measured 10.035 µM, by Beer-Lambert (published maxima, modelled shape)
- tube 1 scanned against a water blank (the instrument's display, simulated)
The instrument showed A 1.304 at 630 nm; the band peaks at 630 nm (Brilliant Blue FCF's published maximum: 627-631 nm); at 630 nm the scan shows A 1.304; the case's method read A 1.303 there (three readings, the blank subtracted, stray light corrected): 10.035 µM ± 0.079 µM (k = 2).
The numbers
| nm | scan (A) | Brilliant Blue FCF at the measured strength (A) |
|---|---|---|
| 400 | 0.146 | 0.146 |
| 410 | 0.155 | 0.153 |
| 420 | 0.129 | 0.128 |
| 430 | 0.088 | 0.088 |
| 440 | 0.051 | 0.051 |
| 450 | 0.029 | 0.025 |
| 460 | 0.012 | 0.011 |
| 470 | 0.014 | 0.011 |
| 480 | 0.018 | 0.015 |
| 490 | 0.023 | 0.020 |
| 500 | 0.029 | 0.027 |
| 510 | 0.048 | 0.043 |
| 520 | 0.067 | 0.066 |
| 530 | 0.091 | 0.091 |
| 540 | 0.131 | 0.131 |
| 550 | 0.182 | 0.183 |
| 560 | 0.262 | 0.263 |
| 570 | 0.368 | 0.369 |
| 580 | 0.477 | 0.478 |
| 590 | 0.590 | 0.586 |
| 600 | 0.723 | 0.720 |
| 610 | 0.927 | 0.928 |
| 620 | 1.164 | 1.166 |
| 630 | 1.304 | 1.305 |
| 640 | 1.170 | 1.173 |
| 650 | 0.830 | 0.830 |
| 660 | 0.502 | 0.502 |
| 670 | 0.271 | 0.273 |
| 680 | 0.125 | 0.121 |
| 690 | 0.055 | 0.053 |
| 700 | 0.013 | 0.015 |
| 710 | -0.002 | 0.000 |
| 720 | -0.003 | 0.000 |
| 730 | -0.001 | 0.000 |
| 740 | 0.002 | 0.000 |
| 750 | -0.002 | 0.000 |
Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5) · simulated: a virtual instrument's scan of the simulated tube; the dye's spectrum published in its maxima, modelled between
- the balance: 1,200.4 µL (± 2u)
- the goal's volume, ±1.5 %
- the engine's predicted fill, 1,201.4 ± 2.8 µL (k = 2)
ζ -0.7: the fill agrees with the prediction (|ζ| < 3 is no action signal).
The numbers
| µL | |
|---|---|
| weighed | 1,200.39 |
| u (weighing) | 0.058 |
| predicted fill | 1,201.42 |
| u (prediction) | 1.391 |
| the goal's volume | 1,182.0-1,218.0 |
| ζ | -0.74 |
ISO 8655-6:2022 (the Z factor); the engine's dispensing model on its (virtual) calibration records; ISO 13528:2022 (ζ) · simulated: a virtual balance's reading of the simulated tube
- Beer-Lambert: A = ε l c, ε = 130.258 per mM per cm (published)
- what the instrument reads (stray light 0.0005)
- the tubes' readings (± U, k = 2)
- the working range, A 0.1-1.5
at the goal the instrument reads -0.32 % below Beer-Lambert (A 1.298 for 1.303); the method inverts the stray-light equation, so the strength is Beer-Lambert's; T1 read A 1.303.
The numbers
| µM | Beer-Lambert (A) | the instrument reads (A) |
|---|---|---|
| 0 | 0.000 | 0.000 |
| 2.5 | 0.326 | 0.325 |
| 5 | 0.651 | 0.651 |
| 7.5 | 0.977 | 0.975 |
| 10 | 1.303 | 1.298 |
| 12.5 | 1.628 | 1.619 |
| 15 | 1.954 | 1.935 |
| 17.5 | 2.280 | 2.240 |
| 20 | 2.605 | 2.526 |
| 22.5 | 2.931 | 2.777 |
| 25 | 3.256 | 2.977 |
Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5); the working range: the stated working range: above A 1.5 the stray-light correction exceeds 0.5 %, below A 0.1 one reading's noise exceeds 2 %; outside it the reading is not used (dilute, or prepare a stronger sample) · published absorptivity; the instrument's stated stray light; simulated readings
Every part in detail
The goal and the decision rule, frozen established procedure
- What it does
- 10.00 µM Brilliant Blue FCF in 1,200 uL, within ±2 % at P >= 0.95
- Principle
- pre-registration: the goal and the rule are fixed, and their digests recorded, before anything is made
- Where
- Goal
- Why
- a tube is judged against the goal, never against its own recipe (a recipe consistent with itself but wrong for the goal once became its own target and passed: G1 stops that)
- Evidence
- established a stated goal and a decision rule after JCGM 106 and ILAC G8, frozen by sha256 digests (G0)
- References
- JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
- ILAC G8:09/2019, Guidelines on decision rules and statements of conformity, 4.2.3: a non-binary statement with a guard band (conditional pass, conditional fail)
- Checked by
- G0: the digests are the frozen ones
- Limitations
- ±2 % on the strength and P ≥ 0.95 are this demonstration's requirements (the Q1 case's), and ±1.5 % on the final volume is the robot's transfer tolerance applied to the total; none is a universal acceptance limit.
- Code
- web/reference.py
The recipe meets the goal (G1) exact law model
- What it does
- the stock and water volumes that give the goal's strength and volume
- Principle
- the dilution law, by an exact mass balance in rationals, independent of the engine's floats
- Equation
- c_tube = C_stock V_stock / (V_stock + V_water)
- Where
- Plan
- Why
- the recipe is checked against the goal before the robot moves
- Evidence
- exact law exact arithmetic on the requested volumes and the declared stock strength
- References
- —
- Checked by
- G1; tests/test_reference_case.py
- Limitations
- —
- Code
- web/reference.py
The stock's strength estimate data
- What it does
- the dye stock's declared strength and how well it is known
- Principle
- the label's strength, with a stated relative uncertainty
- Where
- Plan
- Why
- every tube's strength is proportional to it
- Evidence
- estimate the label (50 µM) with ±0.5 % relative standard uncertainty, assumed: an assumption, stated: about five times the 0.09 % reached by a carefully prepared standard in the Eurachem/CITAC Guide QUAM:2012, Example A1
- References
- —
- Checked by
- —
- Limitations
- not a certified reference: its ±0.5 % dominates the prediction's uncertainty
- in one scenario the stock is not what its label says: the loop must find it
- Code
- web/reference.py
Parameters (2)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| declared strength | 50.0 µM | estimate | the label; measure: weigh the dye for the stock (a gravimetric standard) or standardise it against the calibrated absorptivity |
| relative standard uncertainty | 0.005 | estimate | an assumption, stated: about five times the 0.09 % reached by a carefully prepared standard in the Eurachem/CITAC Guide QUAM:2012, Example A1 |
The engine's plan (G2) calibrated method
- What it does
- how the robot makes each transfer: pump pulses or pipette strokes, tips, mixing
- Principle
- the engine's planner (B0-B5) allocates each transfer to the device whose calibrated error model meets the tolerance, or refuses with the reason
- Where
- Plan
- Why
- a recipe the robot cannot make within tolerance is refused before anything is made
- Evidence
- calibrated the error models come from calibration records; here virtual gravimetric records (weighings on the simulator, not on the real devices)
- References
- ISO 8655-6 (gravimetric calibration of piston-operated volumetric apparatus)
- Checked by
- G2 the plan or its refusal; the engine's B0-B5 gates and their tests
- Limitations
- the calibration records are virtual: a real robot needs its own gravimetric calibration
- Code
- vialbot/pipeline, vialbot/examples/calibration_virtual_gravimetric.json
The prediction and its uncertainty (G4) established method
- What it does
- the strength the tube should have, with its uncertainty and the probability of meeting the goal
- Principle
- the GUM's law of propagation with correlated strokes, validated by the adaptive Monte Carlo method; each transfer from the engine's predictive model of its device
- Where
- Plan
- Why
- the plan is accepted only when the goal is met with P >= 0.95
- Evidence
- established JCGM 100 and 101; its inputs carry their own evidence (the budget below)
- References
- JCGM 100:2008 (GUM)
- JCGM 101:2008 (Monte Carlo)
- Checked by
- G4; tests/test_uncertainty.py
- Limitations
- —
- Code
- web/uncertainty.py
The robot (VialBot) and its notebook (G3) established instrument
- What it does
- moves, picks tips, pipettes and pumps; replies to every command; its notebook is checked independently
- Principle
- a Creality Ender-3 V3 SE motion stage with a Thermo E1-ClipTip 300 pipette head and a Lee LPM solenoid pump; the independent checker (B7) reads the robot's replies, not the tube
- Where
- Execute
- Why
- the transfers are made, acknowledged and checked
- Evidence
- established the devices' datasheets: the pipette 10-300 uL (Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed); the pump 10 uL per pulse, ±15 % (The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %))
- References
- Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed
- The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %)
- Checked by
- G3: the checker's verdict on the robot's notebook (B7)
- Limitations
- every hardware number is a design assumption until the device is surveyed and calibrated
- the pipette's error is modelled the same at every volume; real air-displacement pipettes err more at low volume
- Code
- vialbot/examples/hardware_q1_dye.json
Parameters (4)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| pipette range | 10-300 uL | established | Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed |
| pipette tolerance | 0.015 | estimate | Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed |
| pump pulse | 10 uL | established | The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %) |
| pump accuracy | 0.15 | established | The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %) |
The virtual laboratory (the hidden truth) test scenario virtual laboratory
- What it does
- the simulator that does what the robot really delivers and what the stock really holds
- Principle
- each device has a fixed relative bias and scatter, close to but not identical with any datasheet; the planner never sees them; in one scenario the stock is weaker than its label
- Where
- Execute
- Why
- to test the loop (plan, make, verify, decide, correct) against a laboratory that is not the plan
- Evidence
- test scenario simulated by design: the test bench, not the model; its values are never shown
- References
- —
- Checked by
- —
- Limitations
- it tests the procedure, not the chemistry: a pass here is not a measurement
- Code
- vialbot/qualification/sim.py
Analytical balance established instrument
- What it does
- weighs the tube: its fill volume
- Principle
- mass to volume by the ISO 8655-6 conversion factor Z(t, p, rh) for water; buoyancy for a stated density
- Equation
- V = m Z(t, p, rh)
- Where
- Verify
- Why
- the fill check: is the tube's volume what the recipe ordered?
- Evidence
- established ISO 8655-6's Z factor; the readability of a 4-place balance; its noise an estimate, the tare taken as perfect
- References
- ISO 8655-6:2022, gravimetric reference measurement procedure (the Z factor)
- OIML R 111-1 (2004): weights and buoyancy
- Checked by
- —
- Limitations
- the tare is perfect and evaporation during weighing is neglected
- a simpler balance model than the KCl bench's (no linearity, minimum weight, static or warm-vessel effects): one balance for the whole lab is planned
- Code
- vialbot/instruments/balance.py
Parameters (3)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| readability | 0.1 mg | established | a 4-place analytical balance |
| noise (sd) | 0.05 | estimate | a 4-place balance's typical repeatability; measure: ten weighings of one tube |
| Z | 1.003169 uL/mg | established | ISO 8655-6 |
UV-Vis spectrophotometer established instrument
- What it does
- reads the tube's absorbance at 630, 530 nm against a water blank
- Principle
- Beer-Lambert for additive species, with stray radiation in the sample and the blank beams
- Equation
- A_read = log10((1 + s) / (10^-A + s)) + noise, A = l sum(epsilon c)
- Where
- Verify
- Why
- the tube's strength is measured, not taken from the recipe
- Evidence
- established Beer-Lambert and the stray-light equation (Harvey 10B.5); the noise and the stray fraction are stated choices
- References
- D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray))
- Checked by
- —
- Limitations
- the analytical readings are at two wavelengths; the 400-750 nm scan is for seeing the spectrum, not for the decision
- the scan's band shape between the dye's published band maxima is modelled (malachite green's measured band, the same chromophore family): no open digital spectrum of Brilliant Blue FCF was found; a measured spectrum of the laboratory's own lot replaces it
- no named instrument's specification yet
- Code
- vialbot/instruments/spectrophotometer.py, vialbot/examples/dye_brilliant_blue_fcf.json, scripts/dye_spectrum.py
Parameters (4)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| stray fraction | 0.0005 | estimate | the instrument's stated stray fraction, known to within 50 % (assumed); measure: a cut-off filter or a strong absorber at 630 nm |
| noise (sd) | 0.002 A | estimate | a typical photometric noise; measure: ten readings of one solution |
| resolution | 0.001 A | established | a typical instrument |
| range | 2.5 A | established | stated |
Brilliant Blue FCF's absorptivity, calibrated calibrated data
- What it does
- converts an absorbance into a strength: c = A / (epsilon l)
- Principle
- a calibration line: five standards of the laboratory's own lot made by weight (its dye content assayed), read three times each on this spectrophotometer, the stray light corrected, A = (epsilon l) c fitted
- Equation
- epsilon(630 nm) = slope of A against c = 130.258 per mM per cm (JECFA's published 0.164 L/(mg cm) x 792.86 g/mol = 130)
- Where
- Verify
- Why
- the measured strength is only as good as it
- Evidence
- calibrated a calibration of the lot (a virtual record: simulated readings of virtual standards): 0.32 % relative standard uncertainty, 92 % of it the lot's assay and the glassware (estimates); it recovers the simulator's absorptivity (set from JECFA's) within 0.2 %: a check of the procedure, not of the value
- References
- JECFA specification for Brilliant Blue FCF (28th JECFA 1984, FNP 31/1 1984 and FNP 52 1992; metals revised at the 59th JECFA 2002): 'Absorptivity (a) = 0.164 mg/L/cm at approximately 630 nm', of the pure colouring matter; formula weight 792.86 (https://www.fao.org/fileadmin/user_upload/jecfa_additives/docs/Monograph1/additive-059-m1.pdf)
- the laboratory's calibration record: vialbot/examples/calibration_dye_absorptivity_virtual.json
- Checked by
- tests/test_credibility.py: the record's line and budget; tests/test_duv_measurement.py
- Limitations
- the record is virtual: on a real bench run scripts/dye_calibrate.py's protocol and replace it
- 92 % of its uncertainty is the lot's assay (a titration's typical repeatability, an estimate): replicate titrations or a certified reference material would make the dye's strengths supported
- the readings alone cannot tell a weak stock from a wrong absorptivity: the calibration is the reference
- Code
- web/measurement.py, scripts/dye_calibrate.py, vialbot/examples/calibration_dye_absorptivity_virtual.json
Parameters (5)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| epsilon at 630 nm | 130.258 per mM per cm | calibrated | virtual:dye-calibration:seed20261024 |
| its relative uncertainty | 0.003233 | calibrated | the calibration's budget |
| the lot's dye content (titration) | 0.912 | estimate | the lot's total colouring matter by titration with titanous chloride (JECFA, Combined Compendium Vol. 4); its 0.3 % relative standard uncertainty is an estimate typical of a redox titration's repeatability; measure: replicate titrations of the lot (n >= 3), or a certified reference material |
| the dilutions' glassware | 0.0008 | estimate | class A pipettes and flasks (ISO 648, ISO 1042); 0.08 % per standard, an estimate; measure: weigh each dilution's water |
| the published value (JECFA) | 130.0 per mM per cm | established | JECFA specification for Brilliant Blue FCF (28th JECFA 1984, FNP 31/1 1984 and FNP 52 1992; metals revised at the 59th JECFA 2002): 'Absorptivity (a) = 0.164 mg/L/cm at approximately 630 nm', of the pure colouring matter; formula weight 792.86 (https://www.fao.org/fileadmin/user_upload/jecfa_additives/docs/Monograph1/additive-059-m1.pdf) |
The measured strength (G5) established method
- What it does
- the tube's strength from the readings alone, with its uncertainty and the checks that make a reading usable
- Principle
- three readings of the sample and of a water blank; blank subtracted; stray light corrected by inverting its equation; Beer-Lambert; the GUM's propagation
- Equation
- c = A_corrected / (epsilon l)
- Where
- Verify
- Why
- the decision rests on it
- Evidence
- established established analytical practice (blank, replicates, working range) and the GUM
- References
- D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray))
- JCGM 100:2008 (GUM): law of propagation of uncertainty (5.1.2); expanded uncertainty U = k u with k = 2 (6.2, 6.3)
- Checked by
- G5 checks: blank at zero, inside the working range, enough liquid, mixed
- Limitations
- —
- Code
- web/measurement.py
Parameters (3)
| Parameter | Value | Evidence | Source |
|---|---|---|---|
| replicates | 3 | established | practice |
| working range | 0.1-1.5 A | established | the stated working range: above A 1.5 the stray-light correction exceeds 0.5 %, below A 0.1 one reading's noise exceeds 2 %; outside it the reading is not used (dilute, or prepare a stronger sample) |
| path | 1.0 cm | estimate | ASSUMED: a 1 cm cell, 0.1 % relative standard uncertainty (a quartz cell's tolerance); measure: the cell's certificate |
The decision (D) established decision
- What it does
- accept, reject or undecided; when not accepted, the cause and the next recipe
- Principle
- guarded acceptance at the limits; the measurement against the prediction (zeta); the balance against the planned fill
- Where
- Decide
- Why
- a frozen rule decides, not a person after the fact
- Evidence
- established JCGM 106; Eurachem/CITAC guard bands; ISO 13528 zeta
- References
- JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
- Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification
- ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal
- Checked by
- D; tests/test_decision.py
- Limitations
- —
- Code
- web/decision.py
The stock standardised established method
- What it does
- when the evidence points at the stock, its strength is measured through tube 1 and the next recipe uses it
- Principle
- standardising a solution against a measurement, with its uncertainty
- Where
- Decide
- Why
- the cause is found and corrected, not guessed
- Evidence
- established Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty)
- References
- Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty)
- Checked by
- —
- Limitations
- it inherits the absorptivity's assumption
- Code
- web/decision.py
Report and export established record
- What it does
- the report of the run and the workflow export, with the case's section
- Principle
- the record of what was done, read and decided
- Where
- Report
- Why
- anyone can read what happened and why
- Evidence
- established generated from the run's own record
- References
- —
- Checked by
- tests/test_duv_demo.py: a case report reads what the room read and keeps a hidden scenario hidden
- tests/test_vchem_experiments.py: the report and export routes
- Limitations
- a report of a simulated run: its readings are synthetic
- Code
- web/report.py, web/workflow_export.py
The evidence package (G6) exact law record
- What it does
- the run's evidence bound by digests and verified by running it again
- Principle
- content hashes and a deterministic engine
- Where
- Report
- Why
- anyone can check that the record is what the engine produces
- Evidence
- exact law sha256 digests and reproduction
- References
- —
- Checked by
- G6 VALID; a tampered package is INVALID
- Limitations
- —
- Code
- web/case_package.py
The scientific basis of this experiment, as a report: /world/basis/duv-001 · the other reference experiment: VCW-001 · the whole model and its validation: the overview · this page as JSON: /api/credibility/duv-001.