How it works · DUV-001

DUV-001: a blue-dye dilution made by the robot

Which part of the experiment does what, where, why, on what evidence, and what it cannot do. Every number the experiment reports is traced to the parts it rests on; a number is no stronger than its weakest part, nor than the unmeasured values it rests on.

The same as a report, with the equations, the evidence audit and the positioning: the scientific basis.

The workflow

11 of its 14 parts rest on exact laws, published data, standards or checks against independent data; 2 calibrated, 1 estimate. Each estimate or placeholder names its source and how it would be measured.

the experiment, step by step (click a step for its parts)1. Goalthe goal and the rule,frozenThe goal and the decision rule, frozen: established1 part2. Planthe recipe, theengine's plan and itspredictionThe recipe meets the goal (G1): exact lawThe prediction and its uncertainty (G4): establishedThe engine's plan (G2): calibratedThe stock's strength: estimate4 parts1 to measure3. Executethe robot makes thetubeThe robot (VialBot) and its notebook (G3): established1 partsimulated labthe readings come from the test scenario (the simulated laboratory), not from the model4. Verifythe balance and thespectrophotometerAnalytical balance: establishedUV-Vis spectrophotometer: establishedThe measured strength (G5): establishedBrilliant Blue FCF's absorptivity, calibrated: calibrated4 parts5. Decidethe rule; the cause;the next tubeThe decision (D): establishedThe stock standardised: established2 parts6. Reportthe report, theexport, the evidencepackageThe evidence package (G6): exact lawReport and export: established2 partsnot accepted: the stock is measured through the tube, and the next tube is planned from itthe experiment, step by step (tap a step)1. Goalthe goal and the rule, frozenThe goal and the decision rule, frozen: established2. Planthe recipe, the engine's plan and its…The recipe meets the goal (G1): exact lawThe prediction and its uncertainty (G4): establishedThe engine's plan (G2): calibratedThe stock's strength: estimate3. Executethe robot makes the tubeThe robot (VialBot) and its notebook (G3): establishedsimulated labthe readings come from the test scenario (the simulated laboratory), not from the model4. Verifythe balance and the spectrophotometerAnalytical balance: establishedUV-Vis spectrophotometer: establishedThe measured strength (G5): establishedBrilliant Blue FCF's absorptivity, calibrated: calibrated5. Decidethe rule; the cause; the next tubeThe decision (D): establishedThe stock standardised: established6. Reportthe report, the export, the evidence packageThe evidence package (G6): exact lawReport and export: establishednot accepted: the next tube
The evidence scale: what each word means

measured here measured in this project on physical equipment (none yet)
exact law an exact law or a definition: conservation of matter, molar masses from atomic weights, the relative centrifugal force, a dilution's mass balance
checked published data or a published method, checked here against independent published measurements it was not built from
established a published method, standard or instrument specification, used inside its stated domain; not checked here against independent data
calibrated a parameter fitted to named observations: valid in the fitted domain only
estimate a value or range from a datasheet, a publication or practice, not measured on this bench: a Type B estimate (GUM, JCGM 100, 4.3), its source named and how to measure it said
placeholder a value with no source yet, stated so that a measurement can replace it; how to measure it is said
not modelled absent from the model: a stated limitation
test scenario the simulated laboratory the instruments read: by design not the model, so not scored; it tests the procedure against a laboratory that differs from the model

Each step's parts

The parts that act in each step, with their evidence. Click a part for its principle, its evidence and its limits.

Goal

the goal and the rule, frozen

The goal and the decision rule, frozen established

gates G0

Decide

the rule; the cause; the next tube

The decision (D) establishedThe stock standardised established

gates D

Report

the report, the export, the evidence package

Report and export establishedThe evidence package (G6) exact law

gates G6

What each reported number rests on

NumberGoverned byWeakest partUncertainty resting on estimates or placeholdersVerdictInfluences not measured here
the predicted strength (G4)The recipe meets the goal (G1), The engine's plan (G2), The prediction and its uncertainty (G4), The stock's strengthestimate86 %
the stock's strength (estimated ±0.5 %) 86 % (estimate); the transfers (virtual calibration records) 14 % (established)
uncertainthe stock's estimated ±0.5 % and the virtual calibration records
the tube's fill by weighingAnalytical balanceestablished—supportedthe balance's estimated noise and a perfect tare
the measured strength (G5)UV-Vis spectrophotometer, Brilliant Blue FCF's absorptivity, calibrated, The measured strength (G5)calibrated85 %
absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established)
uncertainthe calibrated absorptivity's uncertainty (mostly the lot's assay, an estimate), the estimated path tolerance and stray fraction
the decision (D)UV-Vis spectrophotometer, Brilliant Blue FCF's absorptivity, calibrated, The measured strength (G5), The decision (D)calibrated85 %
absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established)
uncertainas the measured strength
the cause found (the stock's strength)The measured strength (G5), Brilliant Blue FCF's absorptivity, calibrated, The stock standardisedcalibrated85 %
absorptivity 67 % (calibrated); stray-light correction 16 % (estimate); readings (repeatability of the sample and blank means) 10 % (established)
uncertainas the measured strength

Rule: a quantity is supported when its weakest governing part is at least calibrated and estimates or placeholders carry less than half of its uncertainty; uncertain when a governing part is only an estimate or a placeholder, or when they carry half or more of its uncertainty; unsupported when a part it needs is not modelled.

What the run actually used

the reference run, scenario 'nominal', seed 4242; the dye bench (no chemistry beyond dilution and Beer-Lambert) · The stock is what its label says · The stock is 5 % weaker than its label. The branch each part took in this run: never a value of the hidden laboratory.

Tube 1 (V01)

What the instruments showed

Each reading of the run set against the published data or the specification that gives it meaning, as an analyst would plot it; the numbers are under each figure. The readings are simulated; the curves are published data or the model's.

tube 1's spectrum, 400-750 nm: Brilliant Blue FCF's band at 630 nmthe spectrophotometer's scan of the tube against a water blank; the strength is read at the fixed wavelength, 630 nm
0.00.51.01.5absorbance (1 cm)400500600700wavelength (nm)630 nm: the strength is read here630 nm: the strength is read hereBrilliant Blue FCF at the measured 10.035 µM, by Beer-Lambert (published maxima, modelled shape)tube 1 scanned against a water blank (the instrument's display, simulated)the scan's maximum: A 1.304 at 630 nm630 nm, A 1.304
  • Brilliant Blue FCF at the measured 10.035 µM, by Beer-Lambert (published maxima, modelled shape)
  • tube 1 scanned against a water blank (the instrument's display, simulated)

The instrument showed A 1.304 at 630 nm; the band peaks at 630 nm (Brilliant Blue FCF's published maximum: 627-631 nm); at 630 nm the scan shows A 1.304; the case's method read A 1.303 there (three readings, the blank subtracted, stray light corrected): 10.035 µM ± 0.079 µM (k = 2).

The numbers
nmscan (A)Brilliant Blue FCF at the measured strength (A)
4000.1460.146
4100.1550.153
4200.1290.128
4300.0880.088
4400.0510.051
4500.0290.025
4600.0120.011
4700.0140.011
4800.0180.015
4900.0230.020
5000.0290.027
5100.0480.043
5200.0670.066
5300.0910.091
5400.1310.131
5500.1820.183
5600.2620.263
5700.3680.369
5800.4770.478
5900.5900.586
6000.7230.720
6100.9270.928
6201.1641.166
6301.3041.305
6401.1701.173
6500.8300.830
6600.5020.502
6700.2710.273
6800.1250.121
6900.0550.053
7000.0130.015
710-0.0020.000
720-0.0030.000
730-0.0010.000
7400.0020.000
750-0.0020.000

Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5) · simulated: a virtual instrument's scan of the simulated tube; the dye's spectrum published in its maxima, modelled between

tube 1 on the balance: 1,200.4 µL against the predicted fillthe weighed volume (mass × Z, ISO 8655-6) against the engine's predicted fill and the goal's volume tolerance
the goal's volume, ±1.5 %the engine's predicted fill, 1,201.4 ± 2.8 µL (k = 2)11801190120012101220volume (µL)weighed 1,200.4 ± 0.1 µL (k = 2)1,200.4 µL
  • the balance: 1,200.4 µL (± 2u)
  • the goal's volume, ±1.5 %
  • the engine's predicted fill, 1,201.4 ± 2.8 µL (k = 2)

ζ -0.7: the fill agrees with the prediction (|ζ| < 3 is no action signal).

The numbers
µL
weighed1,200.39
u (weighing)0.058
predicted fill1,201.42
u (prediction)1.391
the goal's volume1,182.0-1,218.0
ζ-0.74

ISO 8655-6:2022 (the Z factor); the engine's dispensing model on its (virtual) calibration records; ISO 13528:2022 (ζ) · simulated: a virtual balance's reading of the simulated tube

The instrument's response at 630 nm: Beer-Lambert, stray light and the working rangeabsorbance against strength: the published absorptivity's straight line, what the instrument reads (stray radiation bends it down), and where each tube's reading lies
the working range, A 0.1-1.50123absorbance at 630 nm (1 cm)0510152025Brilliant Blue FCF (µM)the goal, 10 µMthe goal, 10 µMBeer-Lambert: A = ε l c, ε = 130.258 per mM per cm (published)what the instrument reads (stray light 0.0005)T1: A 1.303 → 10.035 µM ± 0.079 µMT1
  • Beer-Lambert: A = ε l c, ε = 130.258 per mM per cm (published)
  • what the instrument reads (stray light 0.0005)
  • the tubes' readings (± U, k = 2)
  • the working range, A 0.1-1.5

at the goal the instrument reads -0.32 % below Beer-Lambert (A 1.298 for 1.303); the method inverts the stray-light equation, so the strength is Beer-Lambert's; T1 read A 1.303.

The numbers
µMBeer-Lambert (A)the instrument reads (A)
00.0000.000
2.50.3260.325
50.6510.651
7.50.9770.975
101.3031.298
12.51.6281.619
151.9541.935
17.52.2802.240
202.6052.526
22.52.9312.777
253.2562.977

Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5); the working range: the stated working range: above A 1.5 the stray-light correction exceeds 0.5 %, below A 0.1 one reading's noise exceeds 2 %; outside it the reading is not used (dilute, or prepare a stronger sample) · published absorptivity; the instrument's stated stray light; simulated readings

Every part in detail

The goal and the decision rule, frozen established procedure

What it does
10.00 µM Brilliant Blue FCF in 1,200 uL, within ±2 % at P >= 0.95
Principle
pre-registration: the goal and the rule are fixed, and their digests recorded, before anything is made
Where
Goal
Why
a tube is judged against the goal, never against its own recipe (a recipe consistent with itself but wrong for the goal once became its own target and passed: G1 stops that)
Evidence
established a stated goal and a decision rule after JCGM 106 and ILAC G8, frozen by sha256 digests (G0)
References
  • JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
  • ILAC G8:09/2019, Guidelines on decision rules and statements of conformity, 4.2.3: a non-binary statement with a guard band (conditional pass, conditional fail)
Checked by
  • G0: the digests are the frozen ones
Limitations
  • ±2 % on the strength and P ≥ 0.95 are this demonstration's requirements (the Q1 case's), and ±1.5 % on the final volume is the robot's transfer tolerance applied to the total; none is a universal acceptance limit.
Code
web/reference.py

The recipe meets the goal (G1) exact law model

What it does
the stock and water volumes that give the goal's strength and volume
Principle
the dilution law, by an exact mass balance in rationals, independent of the engine's floats
Equation
c_tube = C_stock V_stock / (V_stock + V_water)
Where
Plan
Why
the recipe is checked against the goal before the robot moves
Evidence
exact law exact arithmetic on the requested volumes and the declared stock strength
References
—
Checked by
  • G1; tests/test_reference_case.py
Limitations
—
Code
web/reference.py

The stock's strength estimate data

What it does
the dye stock's declared strength and how well it is known
Principle
the label's strength, with a stated relative uncertainty
Where
Plan
Why
every tube's strength is proportional to it
Evidence
estimate the label (50 µM) with ±0.5 % relative standard uncertainty, assumed: an assumption, stated: about five times the 0.09 % reached by a carefully prepared standard in the Eurachem/CITAC Guide QUAM:2012, Example A1
References
—
Checked by
—
Limitations
  • not a certified reference: its ±0.5 % dominates the prediction's uncertainty
  • in one scenario the stock is not what its label says: the loop must find it
Code
web/reference.py
Parameters (2)
ParameterValueEvidenceSource
declared strength50.0 µMestimatethe label; measure: weigh the dye for the stock (a gravimetric standard) or standardise it against the calibrated absorptivity
relative standard uncertainty0.005 estimatean assumption, stated: about five times the 0.09 % reached by a carefully prepared standard in the Eurachem/CITAC Guide QUAM:2012, Example A1

The engine's plan (G2) calibrated method

What it does
how the robot makes each transfer: pump pulses or pipette strokes, tips, mixing
Principle
the engine's planner (B0-B5) allocates each transfer to the device whose calibrated error model meets the tolerance, or refuses with the reason
Where
Plan
Why
a recipe the robot cannot make within tolerance is refused before anything is made
Evidence
calibrated the error models come from calibration records; here virtual gravimetric records (weighings on the simulator, not on the real devices)
References
  • ISO 8655-6 (gravimetric calibration of piston-operated volumetric apparatus)
Checked by
  • G2 the plan or its refusal; the engine's B0-B5 gates and their tests
Limitations
  • the calibration records are virtual: a real robot needs its own gravimetric calibration
Code
vialbot/pipeline, vialbot/examples/calibration_virtual_gravimetric.json

The prediction and its uncertainty (G4) established method

What it does
the strength the tube should have, with its uncertainty and the probability of meeting the goal
Principle
the GUM's law of propagation with correlated strokes, validated by the adaptive Monte Carlo method; each transfer from the engine's predictive model of its device
Where
Plan
Why
the plan is accepted only when the goal is met with P >= 0.95
Evidence
established JCGM 100 and 101; its inputs carry their own evidence (the budget below)
References
  • JCGM 100:2008 (GUM)
  • JCGM 101:2008 (Monte Carlo)
Checked by
  • G4; tests/test_uncertainty.py
Limitations
—
Code
web/uncertainty.py

The robot (VialBot) and its notebook (G3) established instrument

What it does
moves, picks tips, pipettes and pumps; replies to every command; its notebook is checked independently
Principle
a Creality Ender-3 V3 SE motion stage with a Thermo E1-ClipTip 300 pipette head and a Lee LPM solenoid pump; the independent checker (B7) reads the robot's replies, not the tube
Where
Execute
Why
the transfers are made, acknowledged and checked
Evidence
established the devices' datasheets: the pipette 10-300 uL (Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed); the pump 10 uL per pulse, ±15 % (The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %))
References
  • Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed
  • The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %)
Checked by
  • G3: the checker's verdict on the robot's notebook (B7)
Limitations
  • every hardware number is a design assumption until the device is surveyed and calibrated
  • the pipette's error is modelled the same at every volume; real air-displacement pipettes err more at low volume
Code
vialbot/examples/hardware_q1_dye.json
Parameters (4)
ParameterValueEvidenceSource
pipette range10-300 uLestablishedThermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed
pipette tolerance0.015 estimateThermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed
pump pulse10 uLestablishedThe Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %)
pump accuracy0.15 establishedThe Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %)

The virtual laboratory (the hidden truth) test scenario virtual laboratory

What it does
the simulator that does what the robot really delivers and what the stock really holds
Principle
each device has a fixed relative bias and scatter, close to but not identical with any datasheet; the planner never sees them; in one scenario the stock is weaker than its label
Where
Execute
Why
to test the loop (plan, make, verify, decide, correct) against a laboratory that is not the plan
Evidence
test scenario simulated by design: the test bench, not the model; its values are never shown
References
—
Checked by
—
Limitations
  • it tests the procedure, not the chemistry: a pass here is not a measurement
Code
vialbot/qualification/sim.py

Analytical balance established instrument

What it does
weighs the tube: its fill volume
Principle
mass to volume by the ISO 8655-6 conversion factor Z(t, p, rh) for water; buoyancy for a stated density
Equation
V = m Z(t, p, rh)
Where
Verify
Why
the fill check: is the tube's volume what the recipe ordered?
Evidence
established ISO 8655-6's Z factor; the readability of a 4-place balance; its noise an estimate, the tare taken as perfect
References
  • ISO 8655-6:2022, gravimetric reference measurement procedure (the Z factor)
  • OIML R 111-1 (2004): weights and buoyancy
Checked by
—
Limitations
  • the tare is perfect and evaporation during weighing is neglected
  • a simpler balance model than the KCl bench's (no linearity, minimum weight, static or warm-vessel effects): one balance for the whole lab is planned
Code
vialbot/instruments/balance.py
Parameters (3)
ParameterValueEvidenceSource
readability0.1 mgestablisheda 4-place analytical balance
noise (sd)0.05 estimatea 4-place balance's typical repeatability; measure: ten weighings of one tube
Z1.003169 uL/mgestablishedISO 8655-6

UV-Vis spectrophotometer established instrument

What it does
reads the tube's absorbance at 630, 530 nm against a water blank
Principle
Beer-Lambert for additive species, with stray radiation in the sample and the blank beams
Equation
A_read = log10((1 + s) / (10^-A + s)) + noise, A = l sum(epsilon c)
Where
Verify
Why
the tube's strength is measured, not taken from the recipe
Evidence
established Beer-Lambert and the stray-light equation (Harvey 10B.5); the noise and the stray fraction are stated choices
References
  • D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray))
Checked by
—
Limitations
  • the analytical readings are at two wavelengths; the 400-750 nm scan is for seeing the spectrum, not for the decision
  • the scan's band shape between the dye's published band maxima is modelled (malachite green's measured band, the same chromophore family): no open digital spectrum of Brilliant Blue FCF was found; a measured spectrum of the laboratory's own lot replaces it
  • no named instrument's specification yet
Code
vialbot/instruments/spectrophotometer.py, vialbot/examples/dye_brilliant_blue_fcf.json, scripts/dye_spectrum.py
Parameters (4)
ParameterValueEvidenceSource
stray fraction0.0005 estimatethe instrument's stated stray fraction, known to within 50 % (assumed); measure: a cut-off filter or a strong absorber at 630 nm
noise (sd)0.002 Aestimatea typical photometric noise; measure: ten readings of one solution
resolution0.001 Aestablisheda typical instrument
range2.5 Aestablishedstated

Brilliant Blue FCF's absorptivity, calibrated calibrated data

What it does
converts an absorbance into a strength: c = A / (epsilon l)
Principle
a calibration line: five standards of the laboratory's own lot made by weight (its dye content assayed), read three times each on this spectrophotometer, the stray light corrected, A = (epsilon l) c fitted
Equation
epsilon(630 nm) = slope of A against c = 130.258 per mM per cm (JECFA's published 0.164 L/(mg cm) x 792.86 g/mol = 130)
Where
Verify
Why
the measured strength is only as good as it
Evidence
calibrated a calibration of the lot (a virtual record: simulated readings of virtual standards): 0.32 % relative standard uncertainty, 92 % of it the lot's assay and the glassware (estimates); it recovers the simulator's absorptivity (set from JECFA's) within 0.2 %: a check of the procedure, not of the value
References
  • JECFA specification for Brilliant Blue FCF (28th JECFA 1984, FNP 31/1 1984 and FNP 52 1992; metals revised at the 59th JECFA 2002): 'Absorptivity (a) = 0.164 mg/L/cm at approximately 630 nm', of the pure colouring matter; formula weight 792.86 (https://www.fao.org/fileadmin/user_upload/jecfa_additives/docs/Monograph1/additive-059-m1.pdf)
  • the laboratory's calibration record: vialbot/examples/calibration_dye_absorptivity_virtual.json
Checked by
  • tests/test_credibility.py: the record's line and budget; tests/test_duv_measurement.py
Limitations
  • the record is virtual: on a real bench run scripts/dye_calibrate.py's protocol and replace it
  • 92 % of its uncertainty is the lot's assay (a titration's typical repeatability, an estimate): replicate titrations or a certified reference material would make the dye's strengths supported
  • the readings alone cannot tell a weak stock from a wrong absorptivity: the calibration is the reference
Code
web/measurement.py, scripts/dye_calibrate.py, vialbot/examples/calibration_dye_absorptivity_virtual.json
Parameters (5)
ParameterValueEvidenceSource
epsilon at 630 nm130.258 per mM per cmcalibratedvirtual:dye-calibration:seed20261024
its relative uncertainty0.003233 calibratedthe calibration's budget
the lot's dye content (titration)0.912 estimatethe lot's total colouring matter by titration with titanous chloride (JECFA, Combined Compendium Vol. 4); its 0.3 % relative standard uncertainty is an estimate typical of a redox titration's repeatability; measure: replicate titrations of the lot (n >= 3), or a certified reference material
the dilutions' glassware0.0008 estimateclass A pipettes and flasks (ISO 648, ISO 1042); 0.08 % per standard, an estimate; measure: weigh each dilution's water
the published value (JECFA)130.0 per mM per cmestablishedJECFA specification for Brilliant Blue FCF (28th JECFA 1984, FNP 31/1 1984 and FNP 52 1992; metals revised at the 59th JECFA 2002): 'Absorptivity (a) = 0.164 mg/L/cm at approximately 630 nm', of the pure colouring matter; formula weight 792.86 (https://www.fao.org/fileadmin/user_upload/jecfa_additives/docs/Monograph1/additive-059-m1.pdf)

The measured strength (G5) established method

What it does
the tube's strength from the readings alone, with its uncertainty and the checks that make a reading usable
Principle
three readings of the sample and of a water blank; blank subtracted; stray light corrected by inverting its equation; Beer-Lambert; the GUM's propagation
Equation
c = A_corrected / (epsilon l)
Where
Verify
Why
the decision rests on it
Evidence
established established analytical practice (blank, replicates, working range) and the GUM
References
  • D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray))
  • JCGM 100:2008 (GUM): law of propagation of uncertainty (5.1.2); expanded uncertainty U = k u with k = 2 (6.2, 6.3)
Checked by
  • G5 checks: blank at zero, inside the working range, enough liquid, mixed
Limitations
—
Code
web/measurement.py
Parameters (3)
ParameterValueEvidenceSource
replicates3 establishedpractice
working range0.1-1.5 Aestablishedthe stated working range: above A 1.5 the stray-light correction exceeds 0.5 %, below A 0.1 one reading's noise exceeds 2 %; outside it the reading is not used (dilute, or prepare a stronger sample)
path1.0 cmestimateASSUMED: a 1 cm cell, 0.1 % relative standard uncertainty (a quartz cell's tolerance); measure: the cell's certificate

The decision (D) established decision

What it does
accept, reject or undecided; when not accepted, the cause and the next recipe
Principle
guarded acceptance at the limits; the measurement against the prediction (zeta); the balance against the planned fill
Where
Decide
Why
a frozen rule decides, not a person after the fact
Evidence
established JCGM 106; Eurachem/CITAC guard bands; ISO 13528 zeta
References
  • JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
  • Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification
  • ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal
Checked by
  • D; tests/test_decision.py
Limitations
—
Code
web/decision.py

The stock standardised established method

What it does
when the evidence points at the stock, its strength is measured through tube 1 and the next recipe uses it
Principle
standardising a solution against a measurement, with its uncertainty
Where
Decide
Why
the cause is found and corrected, not guessed
Evidence
established Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty)
References
  • Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty)
Checked by
—
Limitations
  • it inherits the absorptivity's assumption
Code
web/decision.py

Report and export established record

What it does
the report of the run and the workflow export, with the case's section
Principle
the record of what was done, read and decided
Where
Report
Why
anyone can read what happened and why
Evidence
established generated from the run's own record
References
—
Checked by
  • tests/test_duv_demo.py: a case report reads what the room read and keeps a hidden scenario hidden
  • tests/test_vchem_experiments.py: the report and export routes
Limitations
  • a report of a simulated run: its readings are synthetic
Code
web/report.py, web/workflow_export.py

The evidence package (G6) exact law record

What it does
the run's evidence bound by digests and verified by running it again
Principle
content hashes and a deterministic engine
Where
Report
Why
anyone can check that the record is what the engine produces
Evidence
exact law sha256 digests and reproduction
References
—
Checked by
  • G6 VALID; a tampered package is INVALID
Limitations
—
Code
web/case_package.py

The scientific basis of this experiment, as a report: /world/basis/duv-001 · the other reference experiment: VCW-001 · the whole model and its validation: the overview · this page as JSON: /api/credibility/duv-001.