Scientific basis · DUV-001
DUV-001: a blue-dye dilution made by the robot: the scientific basis
Scientific foundations, present capabilities, partial replacement of physical experiments, and the validation roadmap.
The structure and the positioning follow a colleague's technical assessment (24 October 2026); every number on this page is the credibility engine's or the reference record's, so it stays true to the code. · print this page for a PDF.
Status: a functioning reference case of the robot's workflow (plan, make, verify, decide, correct, report); its chemistry is a dilution and Beer-Lambert; not an experimentally qualified robot.
Suggested public description: VialBot is a physics-informed virtual laboratory for screening, testing and refining supported experiments before physical execution.
What 'replacement' means here: VialBot can replace dilution arithmetic, the planning of the robot's transfers within tolerance, the rehearsal of the robot's workflow and the training of the decision and its cause-finding, within its models. It cannot replace the gravimetric calibration of the real pipette and pump, the certification of a stock or a dye's absorptivity, or the physical runs that would qualify the robot.
1. Why this is a world model rather than a 3D animation
A scientific world model represents the state of the experiment and evolves it when an action is performed; its predictions generate what the displays and the instruments show:
s(t + dt) = F_theta( s(t), a(t), dt ) y(t) = H_phi( s(t) ) + noise(t)Here the state s(t) holds each tube's contents (the volume from each stock, the strengths), the robot's tips and position, the instruments' placed tubes; an action a(t) is to order a transfer (stock, tube, volume), send it, pick up a tube, place it on the balance or in the spectrophotometer, write the report; the robot's simulated devices deliver with their own bias and scatter; the balance and the spectrophotometer read the tube. F is the physical and chemical process model, H the simulated instruments. Hence the room is not programmed to 'colour tube 1 blue': the engine's dilution of what the simulated devices really delivered sets the tube's strength, Beer-Lambert sets its absorbance, and the room shows both. It is neither a universal molecular simulation nor a calibrated digital twin of specific equipment: its validity is stated part by part, quantity by quantity (below).
2. What it can replace, reduce, and cannot presently replace
| Research activity | Current practical benefit | Appropriate claim |
|---|---|---|
| Calculate a dilution and the stock's use | the exact dilution law; a recipe that misses the goal is refused (G1) | available |
| Plan the robot's transfers within tolerance | the engine allocates pump pulses and pipette strokes from its calibration records | planning supported; the records are virtual until the robot is calibrated |
| Rehearse the robot workflow | order, send, weigh, read, decide, report, with the engine's refusals and errors | virtual process training |
| Find a wrong stock from the evidence | the decision standardises the stock through tube 1 and corrects tube 2 | demonstrated in simulation |
| Obtain balance and absorbance readings | synthetic readings with stated noise and stray light; the lot's absorptivity calibrated with weighed standards (a virtual record) | synthetic readings; the calibration to be run on the real bench |
| Certify a stock or a dye's absorptivity | needs a reference material and measurements | physical work required |
| Qualify the real robot's delivery | needs a gravimetric calibration of the real pipette and pump | physical work required |
Scope of replacement: the defensible contribution is the partial substitution of exploratory physical iterations by computationally justified screening, training and protocol checking, not the elimination of real experiments. Quantified reductions of cost, waste or runs need a comparative study; none is established yet.
3. What happens when a researcher changes the experiment?
The goal is 10 µM Brilliant Blue FCF in 1,200 uL from a 50 µM stock. Changing the stock's volume changes the tube's strength by the dilution law and its absorbance by Beer-Lambert (the dye's published absorptivity, 1.30 x 10^5 per M per cm at 630 nm):
| stock (uL) | water (uL) | pump pulses (10 uL) | strength (µM) | A at 630 nm (1 cm) | against the goal |
|---|---|---|---|---|---|
| 200 | 1,000 | 20 | 8.333 | 1.085 | outside ±2 % |
| 220 | 980 | 22 | 9.167 | 1.194 | outside ±2 % |
| 240 | 960 | 24 | 10.000 | 1.303 | within ±2 % |
| 260 | 940 | 26 | 10.833 | 1.411 | outside ±2 % |
| 280 | 920 | 28 | 11.667 | 1.520 | outside ±2 % |
These are the recipe's exact consequences; what the robot delivers differs by its devices' bias and scatter (the prediction carries them), and the measurement reads what the tube really holds. The room's reference loop is a synthetic fault experiment: the stock is secretly 5 % weak, tube 1 is rejected, the stock is standardised through it, and tube 2 is accepted.
4. The scientific principles and equations, part by part
Each part distinguishes the physical principle, its actual computation and its evidence. A genuine equation does not guarantee that its parameters fit the experiment studied: the evidence level says how far it does.
The goal and the decision rule, frozen established
Principle: pre-registration: the goal and the rule are fixed, and their digests recorded, before anything is made.
Where: Goal · why: a tube is judged against the goal, never against its own recipe (a recipe consistent with itself but wrong for the goal once became its own target and passed: G1 stops that).
Implementation: web/reference.py
Confidence: established a stated goal and a decision rule after JCGM 106 and ILAC G8, frozen by sha256 digests (G0).
Limitations:
- ±2 % on the strength and P ≥ 0.95 are this demonstration's requirements (the Q1 case's), and ±1.5 % on the final volume is the robot's transfer tolerance applied to the total; none is a universal acceptance limit.
The recipe meets the goal (G1) exact law
Principle: the dilution law, by an exact mass balance in rationals, independent of the engine's floats.
c_tube = C_stock V_stock / (V_stock + V_water)Where: Plan · why: the recipe is checked against the goal before the robot moves.
Implementation: web/reference.py
Confidence: exact law exact arithmetic on the requested volumes and the declared stock strength.
The stock's strength estimate
Principle: the label's strength, with a stated relative uncertainty.
Where: Plan · why: every tube's strength is proportional to it.
Implementation: web/reference.py
Confidence: estimate the label (50 µM) with ±0.5 % relative standard uncertainty, assumed: an assumption, stated: about five times the 0.09 % reached by a carefully prepared standard in the Eurachem/CITAC Guide QUAM:2012, Example A1.
Limitations:
- not a certified reference: its ±0.5 % dominates the prediction's uncertainty
- in one scenario the stock is not what its label says: the loop must find it
The engine's plan (G2) calibrated
Principle: the engine's planner (B0-B5) allocates each transfer to the device whose calibrated error model meets the tolerance, or refuses with the reason.
Where: Plan · why: a recipe the robot cannot make within tolerance is refused before anything is made.
Implementation: vialbot/pipeline, vialbot/examples/calibration_virtual_gravimetric.json
Confidence: calibrated the error models come from calibration records; here virtual gravimetric records (weighings on the simulator, not on the real devices).
Limitations:
- the calibration records are virtual: a real robot needs its own gravimetric calibration
The prediction and its uncertainty (G4) established
Principle: the GUM's law of propagation with correlated strokes, validated by the adaptive Monte Carlo method; each transfer from the engine's predictive model of its device.
Where: Plan · why: the plan is accepted only when the goal is met with P >= 0.95.
Implementation: web/uncertainty.py
Confidence: established JCGM 100 and 101; its inputs carry their own evidence (the budget below).
The virtual laboratory (the hidden truth) test scenario
Principle: each device has a fixed relative bias and scatter, close to but not identical with any datasheet; the planner never sees them; in one scenario the stock is weaker than its label.
Where: Execute · why: to test the loop (plan, make, verify, decide, correct) against a laboratory that is not the plan.
Implementation: vialbot/qualification/sim.py
Confidence: test scenario simulated by design: the test bench, not the model; its values are never shown.
Limitations:
- it tests the procedure, not the chemistry: a pass here is not a measurement
Brilliant Blue FCF's absorptivity, calibrated calibrated
Principle: a calibration line: five standards of the laboratory's own lot made by weight (its dye content assayed), read three times each on this spectrophotometer, the stray light corrected, A = (epsilon l) c fitted.
epsilon(630 nm) = slope of A against c = 130.258 per mM per cm (JECFA's published 0.164 L/(mg cm) x 792.86 g/mol = 130)Where: Verify · why: the measured strength is only as good as it.
Implementation: web/measurement.py, scripts/dye_calibrate.py, vialbot/examples/calibration_dye_absorptivity_virtual.json
Confidence: calibrated a calibration of the lot (a virtual record: simulated readings of virtual standards): 0.32 % relative standard uncertainty, 92 % of it the lot's assay and the glassware (estimates); it recovers the simulator's absorptivity (set from JECFA's) within 0.2 %: a check of the procedure, not of the value.
Limitations:
- the record is virtual: on a real bench run scripts/dye_calibrate.py's protocol and replace it
- 92 % of its uncertainty is the lot's assay (a titration's typical repeatability, an estimate): replicate titrations or a certified reference material would make the dye's strengths supported
- the readings alone cannot tell a weak stock from a wrong absorptivity: the calibration is the reference
The measured strength (G5) established
Principle: three readings of the sample and of a water blank; blank subtracted; stray light corrected by inverting its equation; Beer-Lambert; the GUM's propagation.
c = A_corrected / (epsilon l)Where: Verify · why: the decision rests on it.
Implementation: web/measurement.py
Confidence: established established analytical practice (blank, replicates, working range) and the GUM.
The decision (D) established
Principle: guarded acceptance at the limits; the measurement against the prediction (zeta); the balance against the planned fill.
Where: Decide · why: a frozen rule decides, not a person after the fact.
Implementation: web/decision.py
Confidence: established JCGM 106; Eurachem/CITAC guard bands; ISO 13528 zeta.
The stock standardised established
Principle: standardising a solution against a measurement, with its uncertainty.
Where: Decide · why: the cause is found and corrected, not guessed.
Implementation: web/decision.py
Confidence: established Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty).
Limitations:
- it inherits the absorptivity's assumption
Report and export established
Principle: the record of what was done, read and decided.
Where: Report · why: anyone can read what happened and why.
Implementation: web/report.py, web/workflow_export.py
Confidence: established generated from the run's own record.
Limitations:
- a report of a simulated run: its readings are synthetic
The evidence package (G6) exact law
Principle: content hashes and a deterministic engine.
Where: Report · why: anyone can check that the record is what the engine produces.
Implementation: web/case_package.py
Confidence: exact law sha256 digests and reproduction.
5. The instruments
Virtual instruments respond to the simulated state plus the modelled instrument's errors; they never display the target. Their readings are synthetic.
| Instrument | Evidence | Specification from | Estimated parameters (and placeholders) |
|---|---|---|---|
| The robot (VialBot) and its notebook (G3) moves, picks tips, pipettes and pumps; replies to every command; its notebook is checked independently | established | Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed; The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %) | pipette tolerance |
| Analytical balance weighs the tube: its fill volume | established | ISO 8655-6:2022, gravimetric reference measurement procedure (the Z factor); OIML R 111-1 (2004): weights and buoyancy | noise (sd) |
| UV-Vis spectrophotometer reads the tube's absorbance at 630, 530 nm against a water blank | established | D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray)) | stray fraction; noise (sd) |
What the instruments showed in the reference run
the reference run, scenario 'stock-error', seed 4242: each reading set against the published data or the specification that gives it meaning; the numbers are under each figure.
- Brilliant Blue FCF at the measured 9.534 µM, by Beer-Lambert (published maxima, modelled shape)
- tube 1 scanned against a water blank (the instrument's display, simulated)
The instrument showed A 1.240 at 630 nm; the band peaks at 630 nm (Brilliant Blue FCF's published maximum: 627-631 nm); at 630 nm the scan shows A 1.240; the case's method read A 1.238 there (three readings, the blank subtracted, stray light corrected): 9.534 µM ± 0.074 µM (k = 2).
The numbers
| nm | scan (A) | Brilliant Blue FCF at the measured strength (A) |
|---|---|---|
| 400 | 0.139 | 0.138 |
| 410 | 0.147 | 0.146 |
| 420 | 0.123 | 0.122 |
| 430 | 0.084 | 0.083 |
| 440 | 0.049 | 0.048 |
| 450 | 0.028 | 0.024 |
| 460 | 0.012 | 0.010 |
| 470 | 0.013 | 0.011 |
| 480 | 0.018 | 0.014 |
| 490 | 0.022 | 0.019 |
| 500 | 0.028 | 0.026 |
| 510 | 0.046 | 0.041 |
| 520 | 0.064 | 0.062 |
| 530 | 0.087 | 0.086 |
| 540 | 0.124 | 0.124 |
| 550 | 0.173 | 0.174 |
| 560 | 0.249 | 0.250 |
| 570 | 0.350 | 0.351 |
| 580 | 0.453 | 0.454 |
| 590 | 0.561 | 0.557 |
| 600 | 0.687 | 0.684 |
| 610 | 0.881 | 0.882 |
| 620 | 1.106 | 1.108 |
| 630 | 1.240 | 1.239 |
| 640 | 1.112 | 1.114 |
| 650 | 0.788 | 0.789 |
| 660 | 0.477 | 0.477 |
| 670 | 0.257 | 0.259 |
| 680 | 0.119 | 0.115 |
| 690 | 0.052 | 0.050 |
| 700 | 0.012 | 0.014 |
| 710 | -0.002 | 0.000 |
| 720 | -0.003 | 0.000 |
| 730 | -0.001 | 0.000 |
| 740 | 0.002 | 0.000 |
| 750 | -0.002 | 0.000 |
Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5) · simulated: a virtual instrument's scan of the simulated tube; the dye's spectrum published in its maxima, modelled between
- the balance: 1,200.4 µL (± 2u)
- the goal's volume, ±1.5 %
- the engine's predicted fill, 1,201.4 ± 2.8 µL (k = 2)
ζ -0.7: the fill agrees with the prediction (|ζ| < 3 is no action signal).
The numbers
| µL | |
|---|---|
| weighed | 1,200.39 |
| u (weighing) | 0.058 |
| predicted fill | 1,201.42 |
| u (prediction) | 1.391 |
| the goal's volume | 1,182.0-1,218.0 |
| ζ | -0.74 |
ISO 8655-6:2022 (the Z factor); the engine's dispensing model on its (virtual) calibration records; ISO 13528:2022 (ζ) · simulated: a virtual balance's reading of the simulated tube
- Brilliant Blue FCF at the measured 9.981 µM, by Beer-Lambert (published maxima, modelled shape)
- tube 2 scanned against a water blank (the instrument's display, simulated)
The instrument showed A 1.291 at 630 nm; the band peaks at 630 nm (Brilliant Blue FCF's published maximum: 627-631 nm); at 630 nm the scan shows A 1.291; the case's method read A 1.296 there (three readings, the blank subtracted, stray light corrected): 9.981 µM ± 0.079 µM (k = 2).
The numbers
| nm | scan (A) | Brilliant Blue FCF at the measured strength (A) |
|---|---|---|
| 400 | 0.145 | 0.145 |
| 410 | 0.152 | 0.152 |
| 420 | 0.129 | 0.127 |
| 430 | 0.088 | 0.087 |
| 440 | 0.052 | 0.050 |
| 450 | 0.023 | 0.025 |
| 460 | 0.012 | 0.011 |
| 470 | 0.015 | 0.011 |
| 480 | 0.015 | 0.015 |
| 490 | 0.020 | 0.020 |
| 500 | 0.029 | 0.027 |
| 510 | 0.041 | 0.042 |
| 520 | 0.062 | 0.065 |
| 530 | 0.089 | 0.090 |
| 540 | 0.133 | 0.130 |
| 550 | 0.184 | 0.182 |
| 560 | 0.260 | 0.261 |
| 570 | 0.365 | 0.367 |
| 580 | 0.476 | 0.475 |
| 590 | 0.583 | 0.583 |
| 600 | 0.717 | 0.716 |
| 610 | 0.919 | 0.923 |
| 620 | 1.158 | 1.160 |
| 630 | 1.291 | 1.298 |
| 640 | 1.168 | 1.166 |
| 650 | 0.828 | 0.826 |
| 660 | 0.500 | 0.499 |
| 670 | 0.273 | 0.272 |
| 680 | 0.120 | 0.121 |
| 690 | 0.052 | 0.052 |
| 700 | 0.013 | 0.015 |
| 710 | -0.004 | 0.000 |
| 720 | -0.002 | 0.000 |
| 730 | 0.003 | 0.000 |
| 740 | 0.001 | 0.000 |
| 750 | -0.001 | 0.000 |
Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5) · simulated: a virtual instrument's scan of the simulated tube; the dye's spectrum published in its maxima, modelled between
- the balance: 1,193.8 µL (± 2u)
- the goal's volume, ±1.5 %
- the engine's predicted fill, 1,193.5 ± 2.8 µL (k = 2)
ζ +0.2: the fill agrees with the prediction (|ζ| < 3 is no action signal).
The numbers
| µL | |
|---|---|
| weighed | 1,193.77 |
| u (weighing) | 0.058 |
| predicted fill | 1,193.46 |
| u (prediction) | 1.376 |
| the goal's volume | 1,182.0-1,218.0 |
| ζ | +0.22 |
ISO 8655-6:2022 (the Z factor); the engine's dispensing model on its (virtual) calibration records; ISO 13528:2022 (ζ) · simulated: a virtual balance's reading of the simulated tube
- Beer-Lambert: A = ε l c, ε = 130.258 per mM per cm (published)
- what the instrument reads (stray light 0.0005)
- the tubes' readings (± U, k = 2)
- the working range, A 0.1-1.5
at the goal the instrument reads -0.32 % below Beer-Lambert (A 1.298 for 1.303); the method inverts the stray-light equation, so the strength is Beer-Lambert's; T1 read A 1.238, T2 read A 1.296.
The numbers
| µM | Beer-Lambert (A) | the instrument reads (A) |
|---|---|---|
| 0 | 0.000 | 0.000 |
| 2.5 | 0.326 | 0.325 |
| 5 | 0.651 | 0.651 |
| 7.5 | 0.977 | 0.975 |
| 10 | 1.303 | 1.298 |
| 12.5 | 1.628 | 1.619 |
| 15 | 1.954 | 1.935 |
| 17.5 | 2.280 | 2.240 |
| 20 | 2.605 | 2.526 |
| 22.5 | 2.931 | 2.777 |
| 25 | 3.256 | 2.977 |
Brilliant Blue FCF: JECFA's absorptivity, 0.164 L/(mg cm) at about 630 nm (1.30 x 10^5 L/(mol cm)); band positions from the supplier's specification; the band shape modelled from malachite green's measured band (scripts/dye_spectrum.py). The instrument: Beer-Lambert with stray radiation (D. Harvey, Analytical Chemistry 2.1, 10B.5); the working range: the stated working range: above A 1.5 the stray-light correction exceeds 0.5 %, below A 0.1 one reading's noise exceeds 2 %; outside it the reading is not used (dilute, or prepare a stronger sample) · published absorptivity; the instrument's stated stray light; simulated readings
6. Evidence audit: what is actually supported?
| Evidence | Present? | What it supports |
|---|---|---|
| The sources and the equations can be inspected | yes | every part names its method, equation and code (above) |
| Automated tests | yes | 0 test functions in 0 files (test_reference*.py, test_uncertainty.py, test_measurement.py, test_decision.py, test_duv_*.py, test_credibility.py); the whole suite runs before every commit |
| The recipe's mass balance checked exactly | yes | G1 in exact rationals, independent of the engine's floats |
| Comparison with independent published data | no | not yet: the calibration is virtual and recovers the simulator's own absorptivity (set from JECFA's), so their agreement proves nothing; a calibration on the real spectrophotometer, compared with JECFA's value, and a measured spectrum are to come |
| Synthetic hidden-laboratory trials, a fault and its correction | yes | a stock 5 % weak is found and corrected; the evidence packages verify (G6) |
| A gravimetric calibration of the real pipette and pump | no | the calibration records are virtual (weighings on the simulator) |
| Real robot runs of the case | no | physical end-to-end validation not established |
And the credibility engine's verdict on each reported number (its weakest governing part; the share of its uncertainty that rests on estimates or placeholders):
| Number | Weakest part | On estimates | Verdict |
|---|---|---|---|
| the predicted strength (G4) | estimate | 86 % | uncertain |
| the tube's fill by weighing | established | — | supported |
| the measured strength (G5) | calibrated | 83 % | uncertain |
| the decision (D) | calibrated | 83 % | uncertain |
| the cause found (the stock's strength) | calibrated | 83 % | uncertain |
Rule: a quantity is supported when its weakest governing part is at least calibrated and estimates or placeholders carry less than half of its uncertainty; uncertain when a governing part is only an estimate or a placeholder, or when they carry half or more of its uncertainty; unsupported when a part it needs is not modelled.
7. Strengths and hard boundaries, stated precisely
Strengths demonstrated in software
- The robot's own engine plans and the room drives it: the transfers, refusals and errors are the engine's, not an animation's.
- A goal and a rule frozen first: a recipe that is consistent with itself but wrong for the goal cannot pass (G1).
- Prediction and measurement kept apart: the prediction from the calibration records, the measurement from the readings alone.
- A decision that names its cause: the stock is standardised through tube 1 and the next recipe uses it.
- Reproducible evidence: digests, records and packages that verify by re-running.
Current boundaries
- ±2 % on the strength and P ≥ 0.95 are this demonstration's requirements (the Q1 case's), and ±1.5 % on the final volume is the robot's transfer tolerance applied to the total; none is a universal acceptance limit.
- not a certified reference: its ±0.5 % dominates the prediction's uncertainty
- in one scenario the stock is not what its label says: the loop must find it
- the calibration records are virtual: a real robot needs its own gravimetric calibration
- every hardware number is a design assumption until the device is surveyed and calibrated
- the pipette's error is modelled the same at every volume; real air-displacement pipettes err more at low volume
- it tests the procedure, not the chemistry: a pass here is not a measurement
- the tare is perfect and evaporation during weighing is neglected
- a simpler balance model than the KCl bench's (no linearity, minimum weight, static or warm-vessel effects): one balance for the whole lab is planned
- the analytical readings are at two wavelengths; the 400-750 nm scan is for seeing the spectrum, not for the decision
- the scan's band shape between the dye's published band maxima is modelled (malachite green's measured band, the same chromophore family): no open digital spectrum of Brilliant Blue FCF was found; a measured spectrum of the laboratory's own lot replaces it
- no named instrument's specification yet
- the record is virtual: on a real bench run scripts/dye_calibrate.py's protocol and replace it
- 92 % of its uncertainty is the lot's assay (a titration's typical repeatability, an estimate): replicate titrations or a certified reference material would make the dye's strengths supported
- the readings alone cannot tell a weak stock from a wrong absorptivity: the calibration is the reference
- it inherits the absorptivity's assumption
- a report of a simulated run: its readings are synthetic
- Visual appearance, instrument displays and a PASS in a simulated laboratory are never measured evidence.
8. How to turn 'partially replacing experiments' into a measurable claim
The publishable proposition is not a digital replica of every chemical. A testable question: for a defined class of liquid-handling and crystallisation experiments, how many exploratory physical trials, operator errors and unproductive runs can a verified chemistry world model avoid, without compromising the quality of the eventual physical outcome?
The comparison uses matched objectives and budgets: (A) conventional recipe preparation and trial-and-error; (B) virtual screening plus physical confirmation. Pre-register the scope, the materials and the endpoints; report the physical experiment count, the material consumed, the operator and instrument time, the success fraction, the endpoint error and the corrective actions. Simulated savings are never cited as laboratory savings.
Minimum next demonstration: Run the calibration of the absorptivity (scripts/dye_calibrate.py's protocol) on the real spectrophotometer with replicate assays of the lot's dye content, measure the dye's spectrum, and name the spectrophotometer whose specification the model uses; calibrate the real pipette and pump gravimetrically (ISO 8655-6); freeze the model; then run a held-out dilution series on the robot and compare the measured strengths with the predictions.
Afterwards: Two-dye mixtures (read at two wavelengths and unmixed) are the next case, once single-dye strengths are validated on the robot.
9. Recommended positioning
Headline: VialBot: explore chemistry before entering the laboratory.
One sentence: VialBot is a physics-informed virtual laboratory for screening, testing and refining supported experiments before physical execution.
Two sentences: VialBot combines chemical-state models, simulated laboratory instruments and independently checked workflows to move part of early experimental screening into a virtual environment. It helps researchers reduce avoidable physical trial-and-error while clearly reporting which outcomes are established, estimated or still require bench validation.
Manuscript framing: A provenance-aware chemistry world model for the partial substitution of laboratory trial-and-error through in-silico experiment design, simulation-based fault diagnosis, instrument-aware workflows and quantitatively benchmarked sim-to-real transfer.
Avoid: 'Replaces chemistry experiments'; 'Predicts the exact phase of any material'; 'A fully realistic equipment digital twin'; 'Verified 95 % accurate experimental outcomes' or a quantified waste reduction before measurements exist.
10. References
- JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
- ILAC G8:09/2019, Guidelines on decision rules and statements of conformity, 4.2.3: a non-binary statement with a guard band (conditional pass, conditional fail)
- ISO 8655-6 (gravimetric calibration of piston-operated volumetric apparatus)
- JCGM 100:2008 (GUM)
- JCGM 101:2008 (Monte Carlo)
- Thermo E1-ClipTip 300 datasheet (10-300 uL); plunger ratio and 1.5 % relative tolerance assumed
- The Lee Company LPM series datasheet PDS-129 (10 uL per pulse, +/-15 %)
- ISO 8655-6:2022, gravimetric reference measurement procedure (the Z factor)
- OIML R 111-1 (2004): weights and buoyancy
- D. Harvey, Analytical Chemistry 2.1 (open textbook), 10B.5 Limitations to Beer's law: A = -log((P_T + P_stray) / (P_0 + P_stray))
- JECFA specification for Brilliant Blue FCF (28th JECFA 1984, FNP 31/1 1984 and FNP 52 1992; metals revised at the 59th JECFA 2002): 'Absorptivity (a) = 0.164 mg/L/cm at approximately 630 nm', of the pure colouring matter; formula weight 792.86 (https://www.fao.org/fileadmin/user_upload/jecfa_additives/docs/Monograph1/additive-059-m1.pdf)
- the laboratory's calibration record: vialbot/examples/calibration_dye_absorptivity_virtual.json
- JCGM 100:2008 (GUM): law of propagation of uncertainty (5.1.2); expanded uncertainty U = k u with k = 2 (6.2, 6.3)
- Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification
- ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal
- Eurachem/CITAC Guide QUAM:2012, Example A2: standardising a sodium hydroxide solution (a solution's strength determined against a reference, with its uncertainty)
- ASME V&V 10-2019 and V&V 20-2009 (verification, then validation; E = S - D)
- ASME V&V 40-2018 (context of use, credibility factors)
- Shao, G.; Hightower, J.; Schindel, W. Credibility consideration for digital twins in manufacturing. Manufacturing Letters 2023, 35, 24-28
- JCGM 100:2008 (GUM), JCGM 101:2008 (Monte Carlo), JCGM 106:2012 (conformity assessment)
- ISO 13528:2022 (zeta scores)
- VialBot World Model: scientific foundations, present capabilities, partial replacement of physical experiments, and validation roadmap. A technical assessment of commit 5e1922b by a colleague, 24 October 2026 (the structure and the positioning of this report)
Assessment boundary: everything marked simulated, fitted, estimated, a placeholder or pending remains so; no physical measurement was made for this report.