Reference case · VCW-001 · a crystallisation in a realistic virtual laboratory
VCW-001: recover 60 % of a KCl charge as dry crystals
The laboratory as it should be: accepted The ethanol bottle is not what its label says: accepted
300 mg of KCl in 1,000 uL of water at 25.5 C; ethanol from the bottle labelled 'Ethanol absolute' as the antisolvent; the volume of ethanol is the recipe. The crystals are spun down, the pellet dried to constant mass and weighed, corrected for the salt its mother liquor carried. Goal: 180.0 mg of dry KCl crystals within 6 %.
Record ede873f90b3e1c14 · seed 4242 · engine 0.13.0 (22a4cfa592f66fc8) · world model a8a57a3e1c3e7de2 · the record as JSON · the model's own qualification (W0-W7)
1 · The goal and the rule, fixed before any tube
Goal: 180.0 mg of dry KCl crystals from 300 mg of KCl in 1,000 uL of water, within 6 % (169.2-190.8 mg), at 25.5 C, probability 0.95. the block is set 0.5 C above the table's lowest isotherm (25.0 C) so that its +/-0.5 C accuracy keeps the solution inside the model's domain; the tolerance of 6 % is the smallest the published table's own uncertainty lets the model promise at 95 %
Assay: gravimetric, 105 C to constant mass (two weighings within 0.5 mg), after evaporating the ethanol in the fume hood; minus the KCl the retained mother liquor carried, from the decanted supernatant's measured mass fraction.
Rule: accept the tube when its measured yield lies in [180.0 - 10.8 + w, 180.0 + 10.8 - w], reject it outside [169.2 - w, 190.8 + w], and say it cannot be decided between, w = 1.645 u(Y) at the limits (guarded acceptance, P = 0.95).
Agreement: zeta scores of the measured yield and of the supernatant's mass fraction against the prediction for the tube as made: |zeta| <= 2 agree, 2-3 a warning, >= 3 an action signal.
Cause: in order: an instrument refused or aborted -> execution; constant mass not reached or a tube weighed warm -> drying, weigh again; a fill (weighed against planned addition) at an action signal -> execution; the yield low and the supernatant rich (or the reverse) at an action signal -> the solvent or the temperature: read the ethanol bottle with the densitometer and the block with the thermometer; the one at an action signal against its belief is the cause; neither -> the model's table at this composition (reported, not corrected; the loop stops); the yield alone at an action signal -> handling, repeat.
Learning: a cause found by a check replaces the belief it contradicts (the ethanol's water content from its density through the published table; the block's offset from the thermometer); a belief never changes without a check. Next recipe: the next tube's ethanol volume (or block setting) is the model's for the goal with the new belief, rounded to the instrument's step.
157.2 mg on the left, 202.8 mg on the right. Accept 169.56-190.44 mg; reject below 168.84 or above 191.16 mg; between them the tube cannot be decided (guard band w = 0.36 mg = 1.645 u, u = 0.218 mg).
Goal digest 35901e3371c97d1b, rule digest 67945473e9ac8f64, frozen 2026-10-23 (gate G0).
2 · What you need
Reagents: potassium chloride, reagent grade (assay 99.0-100.5 %, loss on drying <= 1.0 %); the bottle labelled 'Ethanol absolute' (water <= 0.2 %); purified water. Vessels: 15 mL polypropylene conical tubes (no caps in the oven), 25 mL glass weighing bottles, a counterweight tube.
| Instrument | Its specification comes from |
|---|---|
| Analytical balance, 0.1 mg | Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class); NIST SOP 2 (NISTIR 6969, 2019) and OIML R 111-1: conventional mass against 8,000 kg/m3 weights; Mettler Toledo white paper 'Electrostatic charges in weighing': errors of a few mg on plastic vessels, worst below 45 % RH |
| Pipettes P1000 and P5000 | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL; Thermo Scientific application note D00178 'Effect of liquid properties in pipetting': ethanol's vapour pressure makes an air-displacement pipette under-deliver with the forward technique; reverse pipetting reduces it (magnitude ASSUMED) |
| Thermoblock (15 mL SmartBlock) | Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes |
| Syringe pump (slow feed, optional) | Harvard Apparatus Pump 11 Elite (distributor's specification): accuracy +/-0.5 %, reproducibility +/-0.05 %; New Era NE-1000 syringe table (B-D 5 mL, 11.99 mm); start-up lag from slack: Neff et al., Eur. J. Anaesthesiol. 2007 |
| Benchtop centrifuge, swing-bucket rotor | Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm |
| Drying oven | drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096) |
| Desiccator | practice: cool in a desiccator before weighing (USP <731>) |
| Densitometer (for the decision's checks) | Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C |
| Reference thermometer | IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C |
Every specification is the analyst's knowledge; each instrument on the virtual bench has its own hidden characteristics inside it.
3 · The procedure
4 · The prediction (before any tube)
The model's recipe for the goal: 1,580 uL of ethanol (P5000, reverse). Expected 180.0 mg, u = 5.4 mg (first order); the Monte Carlo method (269,994 draws): 180.0 mg, 95 % interval 169.4-190.5 mg, 95.5 % within the tolerance.
| Input | u(Y) mg | share | basis |
|---|---|---|---|
| table_rel | 4.72 | 76 % | Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error |
| eth_sys_uL | 1.26 | 5 % | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL |
| charge_mg | 1.14 | 4 % | ASSUMED: a charge weighed in by spatula within +/-2 mg of the target |
| moisture | 1.03 | 4 % | USP Potassium Chloride: loss on drying <= 1.0 % at 105 C |
| water_sys_uL | 0.93 | 3 % | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL |
| purity | 0.86 | 3 % | USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 % |
| eth_rand_uL | 0.82 | 2 % | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL |
| water_rand_uL | 0.60 | 1 % | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL |
The published table's own uncertainty dominates: it is why the goal's tolerance is 6 % and not less.
5 · Results, scenario by scenario
The laboratory as it should be
Every bottle is what its label says. Only the ordinary scatter of reagents, instruments and hands remains, inside what the model's beliefs allow.
Tube 1 (T1): 1,575 uL of ethanol by pipette (reverse), block 25.5 C. Belief about the ethanol: ASSUMED (the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)).
As made (weighed): charge 299.9 mg, water 994.0 mg, ethanol 1,231.7 mg.
Wet pellet 286.4 mg, dry pellet 190.2 mg, supernatant 2,239.3 mg, residue 109.8 mg: w = 4.889 %, Y = 184.88 mg (U = 0.46 mg, k = 2).
| Check | Verdict | What was found | |
|---|---|---|---|
| G1 | Recipe meets the goal | PASS | With 1575 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg. the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief |
| G2 | Feasible in this laboratory | PASS | Every step is within the instruments' specifications and the model's domain. vialbot/instruments/labkit.py specifications; the table's domain |
| G3 | Performed as ordered | PASS | Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles. the instruments' own logs, never the hidden laboratory |
| G4 | Prediction and its uncertainty | PASS | The model expects 180.0 mg with a standard uncertainty of 5.4 mg (95.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 76 %, eth_sys_uL 5 %, charge_mg 4 %. The Monte Carlo method (269,994 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 169.4-190.5 mg, 95.5 % within the tolerance. JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2) |
| G5 | The instruments' evidence | PASS | Measured yield 184.88 mg (U = 0.46 mg, k = 2); the supernatant held 4.889 % KCl by mass. The supernatant residue (109.8 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %. drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed' |
| D | Decision | PASS | Measured 184.88 mg (u 0.23 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. ACCEPT. Against the prediction for the tube as made: zeta +1.0 for the yield, -0.3 for the supernatant. JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection; ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal |
Decision: accept.
Status: tube 1 (T1) accepted: 184.9 mg. G6 PASS VALID: The package verifies (VALID): every section matches its digest, the goal and the rule are the frozen ones, and running the notebook again gives the same attempts, status and claims.
What this establishes:
rests on G0, G1, G2, G3, G5, D · virtually qualified: a simulated laboratory and its virtual instruments; nothing measured
rests on G4, G5, D · virtually qualified: a simulated laboratory and its virtual instruments; nothing measured
The ethanol bottle is not what its label says (hidden)
The bottle labelled 'Ethanol absolute' holds Ethanol 96 % (Ph. Eur.), 5-7 % water by mass: a common mix-up. The analyst is not told; the readings must reveal it.
Tube 1 (T1): 1,575 uL of ethanol by pipette (reverse), block 25.5 C. Belief about the ethanol: ASSUMED (the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)).
As made (weighed): charge 299.9 mg, water 994.0 mg, ethanol 1,263.4 mg.
Wet pellet 255.6 mg, dry pellet 167.5 mg, supernatant 2,301.8 mg, residue 132.4 mg: w = 5.737 %, Y = 161.77 mg (U = 0.45 mg, k = 2).
| Check | Verdict | What was found | |
|---|---|---|---|
| G1 | Recipe meets the goal | PASS | With 1575 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg. the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief |
| G2 | Feasible in this laboratory | PASS | Every step is within the instruments' specifications and the model's domain. vialbot/instruments/labkit.py specifications; the table's domain |
| G3 | Performed as ordered | PASS | Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles. the instruments' own logs, never the hidden laboratory |
| G4 | Prediction and its uncertainty | PASS | The model expects 180.0 mg with a standard uncertainty of 5.4 mg (95.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 76 %, eth_sys_uL 5 %, charge_mg 4 %. The Monte Carlo method (269,994 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 169.4-190.5 mg, 95.5 % within the tolerance. JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2) |
| G5 | The instruments' evidence | PASS | Measured yield 161.77 mg (U = 0.45 mg, k = 2); the supernatant held 5.737 % KCl by mass. The supernatant residue (132.4 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %. drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed' |
| D | Decision | FAIL | Measured 161.77 mg (u 0.22 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. REJECT. Against the prediction for the tube as made: zeta -4.2 for the yield, +5.1 for the supernatant. The densitometer reads 0.8095 g/cm3 where absolute ethanol gives 0.7878 (zeta +27): the bottle holds 7.6 +/- 0.3 % water. The next tube takes 2170 uL of it. JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection |
Next: the bottle holds 7.6 % water, not <= 0.2 %: 2,170 uL of ethanol in the next tube.
Decision: reject.
Tube 2 (T2): 2,170 uL of ethanol by pipette (reverse), block 25.5 C. Belief about the ethanol: MEASURED (the bottle's density 0.8095 g/cm3 at 22.1 C, through the published table).
As made (weighed): charge 297.9 mg, water 994.6 mg, ethanol 1,733.7 mg.
Wet pellet 280.6 mg, dry pellet 186.3 mg, supernatant 2,745.7 mg, residue 111.9 mg: w = 4.063 %, Y = 181.93 mg (U = 0.45 mg, k = 2).
| Check | Verdict | What was found | |
|---|---|---|---|
| G1 | Recipe meets the goal | PASS | With 2170 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg. the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief |
| G2 | Feasible in this laboratory | PASS | Every step is within the instruments' specifications and the model's domain. vialbot/instruments/labkit.py specifications; the table's domain |
| G3 | Performed as ordered | PASS | Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles. the instruments' own logs, never the hidden laboratory |
| G4 | Prediction and its uncertainty | PASS | The model expects 180.0 mg with a standard uncertainty of 3.9 mg (99.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 53 %, eth_water 10 %, charge_mg 8 %. The Monte Carlo method (209,996 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 172.3-187.6 mg, 99.5 % within the tolerance. JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2) |
| G5 | The instruments' evidence | PASS | Measured yield 181.93 mg (U = 0.45 mg, k = 2); the supernatant held 4.063 % KCl by mass. The supernatant residue (111.9 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %. drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed' |
| D | Decision | PASS | Measured 181.93 mg (u 0.23 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. ACCEPT. Against the prediction for the tube as made: zeta +1.2 for the yield, -0.2 for the supernatant. JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection; ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal |
Decision: accept.
Status: tube 2 (T2) accepted: 181.9 mg. G6 PASS VALID: The package verifies (VALID): every section matches its digest, the goal and the rule are the frozen ones, and running the notebook again gives the same attempts, status and claims.
What this establishes:
rests on G5, D · virtually qualified: a simulated laboratory and its virtual instruments; nothing measured
rests on G0, G1, G2, G3, G5, D · virtually qualified: a simulated laboratory and its virtual instruments; nothing measured
rests on G4, G5, D · virtually qualified: a simulated laboratory and its virtual instruments; nothing measured
6 · Troubleshooting
| What you see | What to do |
|---|---|
| The centrifuge stops with 'Inb' | No counterweight, or one of another mass: fill a tube with water to the same mass within 0.1 g. |
| The centrifuge stops with 'SPEED' | The A-4-38 rotor stops at 4,400 rpm: 5,000 rpm, as some protocols say, is not attainable with a swing-bucket rotor for 15 mL tubes. |
| The oven refuses | A vessel still wet with ethanol: evaporate it in the fume hood first (flammable vapour). |
| No yield: 'dried to constant mass' is UNKNOWN | Dry another hour and weigh again: two weighings must agree within 0.5 mg. |
| No yield: 'cooled' is UNKNOWN | A vessel weighed warm reads light: 30 min in the desiccator before each weighing. |
| The yield is low and the supernatant rich | The solvent is wetter (or the block warmer) than believed: read the ethanol bottle with the densitometer and the block with the thermometer. |
| The pipette refuses a volume | The P5000 sets volumes in steps of 5 uL; the P1000 in steps of 1 uL; each has its range. |
7 · Limits
Not claimed:
- the crystal size or the onset of crystallisation (the model's size is likely too large; neither is measured here)
- that the hidden laboratory is a real one: it is a simulated laboratory whose differences from the model come from the literature
- the model's table validated at this composition: agreement in a virtual laboratory tests the procedure and the decision, not the table
- any yield at another temperature, charge or antisolvent than the goal's
- any physical measurement
Assumed in the virtual laboratory (each with what would measure it):
- vessel_tare: a 15 mL PP tube without its cap weighs 6.85-6.95 g; a 25 mL glass weighing bottle 14.5-15.5 g: weigh a few
- spatula_sd_mg: a charge weighed in by spatula lands within about 1 mg of the target (sd): the analyst's practice
- drying: a pellet's or residue's solvent leaves as exp(-E/tau), tau 15-30 min, with an hour in the fume hood counting a quarter of an oven hour and a hot plate as much as the oven: weigh a pellet after each hour of drying
- occluded: 0-0.2 % of the crystals' mass is mother liquor the oven does not remove: bounded by the USP loss-on-drying limit
- pellet: pellet porosity 0.35-0.55 (random packings of cubes, Baker & Kudrolli 2010), a film of 20-60 uL left by the decant, 0-1 % of the crystals poured off: weigh wet and dry pellets
- bench_cooling: a tube left on the bench cools 1.5 times slower than in the desiccator
- spin_temperature: a tube keeps the block's temperature through a 10 min spin in a refrigerated chamber set to the same point: measure with a probe tube spun alongside
8 · The record and its evidence package
The record (JSON) holds both scenarios' notebooks; anyone can run them again (POST /api/world/bench) and check the evidence packages:
References
- gum: JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G
- mcm: JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)
- conformity: JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
- guard: Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification
- zeta: ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal
- solubility: Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates
- density: Lange's Handbook of Chemistry, 10th ed. (1967): density of aqueous ethanol (vialbot/examples/world/density_water_ethanol.json)
- constant_mass: drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention)
- minimum_weight: USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'
- instruments: vialbot/instruments/labkit.py: every instrument's specification and its source