{"record_version": "1", "case": {"id": "vcw-001", "title": "VCW-001: recover 60 % of a KCl charge as dry crystals", "summary": "300 mg of KCl in 1,000 uL of water at 25.5 C; ethanol from the bottle labelled 'Ethanol absolute' as the antisolvent; the volume of ethanol is the recipe. The crystals are spun down, the pellet dried to constant mass and weighed, corrected for the salt its mother liquor carried. Goal: 180.0 mg of dry KCl crystals within 6 %.", "requirements": "the recipe is the model's (the published solubility table, interpolated); the measurement is the laboratory's own (gravimetric, with its uncertainty); the decision rule is fixed before any tube is made"}, "intent": {"system": "kcl-water-ethanol", "solute": "KCl", "charge_mg": 300.0, "water_uL": 1000.0, "antisolvent": "ethanol ('Ethanol absolute' bottle)", "target_mg": 180.0, "tol_rel": 0.06, "coverage": 0.95, "T_C": 25.5, "T_basis": "the block is set 0.5 C above the table's lowest isotherm (25.0 C) so that its +/-0.5 C accuracy keeps the solution inside the model's domain; the tolerance of 6 % is the smallest the published table's own uncertainty lets the model promise at 95 %", "vessel": "15 mL polypropylene conical tube", "assay": {"method": "gravimetric", "drying": "105 C to constant mass (two weighings within 0.5 mg), after evaporating the ethanol in the fume hood", "correction": "minus the KCl the retained mother liquor carried, from the decanted supernatant's measured mass fraction"}, "tubes": ["T1", "T2", "T3"], "bottles": ["S1", "S2", "S3"]}, "intent_digest": "35901e3371c97d1b", "decision": {"rule": "accept the tube when its measured yield lies in [180.0 - 10.8 + w, 180.0 + 10.8 - w], reject it outside [169.2 - w, 190.8 + w], and say it cannot be decided between, w = 1.645 u(Y) at the limits (guarded acceptance, P = 0.95)", "agreement": "zeta scores of the measured yield and of the supernatant's mass fraction against the prediction for the tube as made: |zeta| <= 2 agree, 2-3 a warning, >= 3 an action signal", "zeta_warning": 2.0, "zeta_action": 3.0, "cause": "in order: an instrument refused or aborted -> execution; constant mass not reached or a tube weighed warm -> drying, weigh again; a fill (weighed against planned addition) at an action signal -> execution; the yield low and the supernatant rich (or the reverse) at an action signal -> the solvent or the temperature: read the ethanol bottle with the densitometer and the block with the thermometer; the one at an action signal against its belief is the cause; neither -> the model's table at this composition (reported, not corrected; the loop stops); the yield alone at an action signal -> handling, repeat", "learning": "a cause found by a check replaces the belief it contradicts (the ethanol's water content from its density through the published table; the block's offset from the thermometer); a belief never changes without a check", "recipe": "the next tube's ethanol volume (or block setting) is the model's for the goal with the new belief, rounded to the instrument's step", "tubes": ["T1", "T2", "T3"], "frozen": "2026-10-23"}, "decision_digest": "67945473e9ac8f64", "limits": {"p": 0.95, "z": 1.644854, "u_mg": 0.218225, "yield_mg": {"tolerance": [169.2, 190.8], "w": [0.358947, 0.358947], "accept": [169.558947, 190.441053], "reject_outside": [168.841053, 191.158947]}}, "model": {"charge_half_width_mg": 2.0, "charge_basis": "ASSUMED: a charge weighed in by spatula within +/-2 mg of the target", "kcl_moisture": {"lo": 0.0, "hi": 0.01, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C"}, "kcl_purity": {"lo": 0.99, "hi": 1.0, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %"}, "loss_rel": {"lo": 0.0, "hi": 0.01, "basis": "ASSUMED: up to 1 % of the crystals lost on decanting"}, "occluded_rel": {"lo": 0.0, "hi": 0.002, "basis": "ASSUMED: up to 0.2 % of the crystals' mass is mother liquor the oven leaves (USP LOD bound)"}, "form": "the interpolation's own error is part of the table's stated uncertainty (its leave-one-out roughness term between the points, and the half-difference term between the isotherms): no separate component, which would count it twice", "block_check": "before the first tube the block is checked with the reference thermometer (a probe in a water-filled tube at the set point): its offset is then a measured belief, not the datasheet's +/-0.5 C", "mc_seed": 20261023, "room_C": 22.0}, "label_belief": {"ethanol_water": {"provenance": "ASSUMED", "dist": "rect", "lo": 0.0, "hi": 0.002, "basis": "the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)"}, "block_offset_K": {"provenance": "ASSUMED", "dist": "rect", "lo": -0.5, "hi": 0.5, "basis": "the thermoblock's specification: +/-0.5 C at 20-45 C"}}, "nominal_recipe": {"charge_mg": 300.0, "water_uL": 1000.0, "water_pipette": "P1000", "ethanol_by": "pipette", "ethanol_pipette": "P5000", "technique": "reverse", "block_C": 25.5, "ethanol_uL": 1580.0, "spin": {"rpm": 4400.0, "min": 10.0, "set_C": 25.5}}, "instruments": {"balance": {"readability_mg": 0.1, "repeatability_sd_mg": 0.08, "linearity_limit_mg": 0.2, "linearity_typical_mg": 0.06, "capacity_g": 220.0, "min_weight_mg": 160.0, "min_weight_optimum_mg": 82.0, "electrostatic_sd_mg": {"low_rh_below_45_pct": 2.0, "normal": 0.2, "ionizer": 0.02}, "warm_bias_mg_per_K": -0.05, "source": "Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class); NIST SOP 2 (NISTIR 6969, 2019) and OIML R 111-1: conventional mass against 8,000 kg/m3 weights; Mettler Toledo white paper 'Electrostatic charges in weighing': errors of a few mg on plastic vessels, worst below 45 % RH", "assumed": {"warm_bias_mg_per_K": "ASSUMED: a tube weighed warmer than the air reads light (convection); measure by weighing a tube straight from the oven and again after 30 min in the desiccator"}}, "pipettes": {"models": {"P1000": {"range_uL": [100.0, 1000.0], "nominal_uL": 1000.0, "step_uL": 1.0, "sys_mpe_uL": 8.0, "rand_mpe_uL": 3.0}, "P5000": {"range_uL": [500.0, 5000.0], "nominal_uL": 5000.0, "step_uL": 5.0, "sys_mpe_uL": 40.0, "rand_mpe_uL": 15.0}}, "volatile_belief": {"forward": {"mean": -0.02, "half_width": 0.01}, "reverse": {"mean": -0.0015, "half_width": 0.0015}}, "source": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL; Thermo Scientific application note D00178 'Effect of liquid properties in pipetting': ethanol's vapour pressure makes an air-displacement pipette under-deliver with the forward technique; reverse pipetting reduces it (magnitude ASSUMED)", "assumed": {"volatile": "ASSUMED magnitudes (1-3 % forward, 0-0.3 % reverse) for ethanol; measure with 10 forward and 10 reverse gravimetric strokes of absolute ethanol", "step_uL": "ASSUMED setting increments"}}, "syringe_pump": {"accuracy_rel": 0.005, "reproducibility_rel": 0.0005, "syringe": "B-D 5 mL, 11.99 mm", "syringe_uL": 5000.0, "rate_range_mL_h": [4.748e-06, 345.5], "source": "Harvard Apparatus Pump 11 Elite (distributor's specification): accuracy +/-0.5 %, reproducibility +/-0.05 %; New Era NE-1000 syringe table (B-D 5 mL, 11.99 mm); start-up lag from slack: Neff et al., Eur. J. Anaesthesiol. 2007", "assumed": {"slack_uL": "ASSUMED 10-60 uL taken up by slack when the line is not primed; measure by weighing the first minute's delivery"}}, "thermoblock": {"accuracy_K": 0.5, "range_C": [5.0, 100.0], "positions": 8, "vessel": "15 mL conical (SmartBlock 15 mL)", "fluctuation_sd_K": 0.1, "source": "Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes", "assumed": {"fluctuation_sd_K": "ASSUMED; a stability figure for the 15 mL block was not found"}}, "centrifuge": {"rotor": "A-4-38 swing-bucket, 15 mL conical", "r_max_cm": 13.7, "max_rpm": 4400.0, "rpm_step": 100.0, "ramp_up_s": 25.0, "ramp_down_s": 25.0, "imbalance_limit_g": 2.0, "refrigerated": true, "temp_alarm_K": 3.0, "temp_control_K": 1.0, "source": "Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm", "assumed": {"imbalance_limit_g": "ASSUMED 2 g: the manual gives no number; weigh the two tubes to within 0.1 g", "temp_control_K": "ASSUMED +/-1 C control at 25 C"}}, "drying_oven": {"set_C": 105.0, "uniformity_K": 2.0, "max_for_pp_tubes_C": 121.0, "constant_mass_mg": 0.5, "cycle_min": 60.0, "source": "drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096)", "assumed": {"uniformity_K": "ASSUMED +/-2 C for a natural-convection oven"}}, "desiccator": {"tau_min": 7.0, "source": "practice: cool in a desiccator before weighing (USP <731>)", "assumed": {"tau_min": "ASSUMED; measure with a probe in a tube"}}, "densitometer": {"accuracy_g_cm3": 0.001, "repeatability_sd_g_cm3": 0.0005, "resolution_g_cm3": 0.0001, "temp_accuracy_K": 0.2, "source": "Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C"}, "thermometer": {"class": "IEC 60751 class AA", "resolution_K": 0.01, "source": "IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C"}, "sources": {"balance": "Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class)", "buoyancy": "NIST SOP 2 (NISTIR 6969, 2019) and OIML R 111-1: conventional mass against 8,000 kg/m3 weights", "electrostatics": "Mettler Toledo white paper 'Electrostatic charges in weighing': errors of a few mg on plastic vessels, worst below 45 % RH", "pipette": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "volatile": "Thermo Scientific application note D00178 'Effect of liquid properties in pipetting': ethanol's vapour pressure makes an air-displacement pipette under-deliver with the forward technique; reverse pipetting reduces it (magnitude ASSUMED)", "pump": "Harvard Apparatus Pump 11 Elite (distributor's specification): accuracy +/-0.5 %, reproducibility +/-0.05 %; New Era NE-1000 syringe table (B-D 5 mL, 11.99 mm); start-up lag from slack: Neff et al., Eur. J. Anaesthesiol. 2007", "block": "Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes", "centrifuge": "Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm", "oven": "drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096)", "densitometer": "Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C", "thermometer": "IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C"}, "provenance": "simulated: specifications from the sources named; every unit's hidden characteristics drawn inside them"}, "assumed": {"vessel_tare": "a 15 mL PP tube without its cap weighs 6.85-6.95 g; a 25 mL glass weighing bottle 14.5-15.5 g: weigh a few", "spatula_sd_mg": "a charge weighed in by spatula lands within about 1 mg of the target (sd): the analyst's practice", "drying": "a pellet's or residue's solvent leaves as exp(-E/tau), tau 15-30 min, with an hour in the fume hood counting a quarter of an oven hour and a hot plate as much as the oven: weigh a pellet after each hour of drying", "occluded": "0-0.2 % of the crystals' mass is mother liquor the oven does not remove: bounded by the USP loss-on-drying limit", "pellet": "pellet porosity 0.35-0.55 (random packings of cubes, Baker & Kudrolli 2010), a film of 20-60 uL left by the decant, 0-1 % of the crystals poured off: weigh wet and dry pellets", "bench_cooling": "a tube left on the bench cools 1.5 times slower than in the desiccator", "spin_temperature": "a tube keeps the block's temperature through a 10 min spin in a refrigerated chamber set to the same point: measure with a probe tube spun alongside"}, "room": {"t_c": 22.0, "p_kpa": 101.325, "rh_pct": 50.0}, "references": {"gum": "JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G", "mcm": "JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)", "conformity": "JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection", "guard": "Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification", "zeta": "ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal", "solubility": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates", "density": "Lange's Handbook of Chemistry, 10th ed. (1967): density of aqueous ethanol (vialbot/examples/world/density_water_ethanol.json)", "constant_mass": "drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention)", "minimum_weight": "USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'", "instruments": "vialbot/instruments/labkit.py: every instrument's specification and its source"}, "seed": 4242, "scenarios": {"nominal": {"words": {"title": "The laboratory as it should be", "lead": "Every bottle is what its label says. Only the ordinary scatter of reagents, instruments and hands remains, inside what the model's beliefs allow."}, "hidden": false, "actions": [{"do": "read_temperature", "where": "block", "set_C": 25.5}, {"do": "tare", "vessel": "T1"}, {"do": "tare", "vessel": "S1"}, {"do": "weigh_in", "vessel": "T1", "target_mg": 300.0}, {"do": "pipette", "vessel": "T1", "liquid": "water", "uL": 1000.0, "pipette": "P1000", "technique": "forward"}, {"do": "weigh", "vessel": "T1"}, {"do": "vortex", "vessel": "T1", "s": 60}, {"do": "block", "vessel": "T1", "set_C": 25.5, "min": 10}, {"do": "pipette", "vessel": "T1", "liquid": "ethanol", "uL": 1575.0, "pipette": "P5000", "technique": "reverse"}, {"do": "weigh", "vessel": "T1"}, {"do": "vortex", "vessel": "T1", "s": 10}, {"do": "block", "vessel": "T1", "set_C": 25.5, "min": 60}, {"do": "counterweight", "vessel": "CW", "match": "T1"}, {"do": "spin", "tubes": ["T1", "CW"], "rpm": 4400, "min": 10, "set_C": 25.5}, {"do": "decant", "vessel": "T1", "into": "S1"}, {"do": "weigh", "vessel": "S1"}, {"do": "weigh", "vessel": "T1"}, {"do": "evaporate", "vessel": "S1", "where": "hotplate", "min": 120}, {"do": "evaporate", "vessel": "T1", "where": "hood", "min": 120}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}], "bench_digest": "2ce3a7a7dac981e1", "attempts": [{"n": 1, "tube": "T1", "bottle": "S1", "recipe": {"tube": "T1", "charge_mg": 300.0, "water_uL": 1000.0, "water_pipette": "P1000", "ethanol_uL": 1575.0, "ethanol_by": "pipette", "block_C": 25.5, "ethanol_pipette": "P5000", "technique": "reverse", "spin": {"rpm": 4400.0, "min": 10.0, "set_C": 25.5}}, "belief": {"ethanol_water": {"provenance": "ASSUMED", "dist": "rect", "lo": 0.0, "hi": 0.002, "basis": "the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)"}, "block_offset_K": {"provenance": "MEASURED", "dist": "normal", "mean": -0.15, "sd": 0.082763, "from_attempt": 0, "set_C": 25.5, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)"}}, "made": {"charge_mg": 299.9, "water_mg": 994.0, "ethanol_mg": 1231.7, "u_mg": 0.123558, "gross_mg": {"tare": 6867.8, "charge": 7167.7, "water": 8161.7, "ethanol": 9393.4}}, "gates": [{"id": "G1", "name": "Recipe meets the goal", "verdict": "PASS", "reason": "With 1575 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg.", "basis": "the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief", "numbers": {"yield_mg": 180.023526, "gap_mg": 0.023526, "ethanol_uL": 1575.0}}, {"id": "G2", "name": "Feasible in this laboratory", "verdict": "PASS", "reason": "Every step is within the instruments' specifications and the model's domain.", "basis": "vialbot/instruments/labkit.py specifications; the table's domain", "numbers": {"problems": []}}, {"id": "G3", "name": "Performed as ordered", "verdict": "PASS", "reason": "Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles.", "basis": "the instruments' own logs, never the hidden laboratory", "numbers": {"missing": [], "recovered": []}}, {"id": "G4", "name": "Prediction and its uncertainty", "verdict": "PASS", "reason": "The model expects 180.0 mg with a standard uncertainty of 5.4 mg (95.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 76 %, eth_sys_uL 5 %, charge_mg 4 %. The Monte Carlo method (269,994 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 169.4-190.5 mg, 95.5 % within the tolerance.", "basis": "JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)", "numbers": {"y_mg": 180.023526, "u_mg": 5.400433, "interval_mg": [169.438873, 190.60818], "p_within": 0.95448, "w_sat": 0.049302, "u_w_sat": 0.002218, "as_made": false}}, {"id": "G5", "name": "The instruments' evidence", "verdict": "PASS", "reason": "Measured yield 184.88 mg (U = 0.46 mg, k = 2); the supernatant held 4.889 % KCl by mass. The supernatant residue (109.8 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %.", "basis": "drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'", "numbers": {"y_mg": 184.876303, "u_mg": 0.229054, "w": 0.048886, "u_w": 8.4e-05, "masses_mg": {"wet_pellet": 286.4, "dry_pellet": 190.2, "supernatant": 2239.3, "residue": 109.8}}}, {"id": "D", "name": "Decision", "verdict": "PASS", "reason": "Measured 184.88 mg (u 0.23 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. ACCEPT. Against the prediction for the tube as made: zeta +1.0 for the yield, -0.3 for the supernatant.", "basis": "JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection; ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal", "numbers": {"outcome": "accept", "y_mg": 184.876303, "u_mg": 0.229054, "p_conform": 1.0, "acceptance_mg": [169.558947, 190.441053], "rejection_below_mg": 168.841053, "rejection_above_mg": 191.158947, "zeta_yield": 0.953848, "zeta_w": -0.286082, "predicted_mg": 180.172525, "u_predicted_mg": 4.926048, "predicted_w": 0.049436, "fills": [{"liquid": "water", "weighed_uL": 994.0, "u_weighed_uL": 0.123558, "predicted_uL": 997.772977, "u_predicted_uL": 5.495305, "zeta": -0.686408, "agree": true}, {"liquid": "ethanol", "weighed_uL": 1231.7, "u_weighed_uL": 0.123558, "predicted_uL": 1239.104047, "u_predicted_uL": 21.724076, "zeta": -0.340817, "agree": true}], "cause": null}}], "prediction": {"planned": {"first_order": {"y_mg": 180.023526, "u_mg": 5.400433, "k": 1.959964, "interval_mg": [169.438873, 190.60818], "p_within": 0.95448, "w_sat": 0.049302, "u_w_sat": 0.002218, "w_water": 0.446981, "crystallised_mg": 180.928167}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.038835, "unit": "1", "c_yield": -121.489054, "u_yield_mg": 4.718041, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.763249}, {"input": "eth_sys_uL", "dist": "rect", "u": 23.094011, "unit": "uL", "c_yield": 0.054452, "u_yield_mg": 1.257519, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.054222}, {"input": "charge_mg", "dist": "rect", "u": 1.154701, "unit": "mg", "c_yield": 0.984068, "u_yield_mg": 1.136304, "basis": "ASSUMED: a charge weighed in by spatula within +/-2 mg of the target", "difference": "central", "share": 0.044272}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -357.429387, "u_yield_mg": 1.03181, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.036504}, {"input": "water_sys_uL", "dist": "rect", "u": 4.618802, "unit": "uL", "c_yield": -0.200958, "u_yield_mg": 0.928184, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.02954}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 297.0075, "u_yield_mg": 0.857387, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.025206}, {"input": "eth_rand_uL", "dist": "normal", "u": 15.0, "unit": "uL", "c_yield": 0.054452, "u_yield_mg": 0.816776, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.022874}, {"input": "water_rand_uL", "dist": "normal", "u": 3.0, "unit": "uL", "c_yield": -0.200958, "u_yield_mg": 0.602873, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.012462}], "mcm": {"trials": 269994, "batches": 27, "batch_size": 10000, "seed": 20261023, "stable": true, "tolerance_mg": 0.05, "refused_draws": 6, "y_mg": 179.995815, "u_mg": 5.399429, "interval_mg": [169.361458, 190.465582], "p_within": 0.955095, "p_se": 0.000399}, "validation": {"tolerance_mg": 0.05, "d_low_mg": 0.077415, "d_high_mg": 0.142599, "validated": false, "basis": "JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)"}}, "as_made": {"first_order": {"y_mg": 180.172525, "u_mg": 4.926048, "k": 1.959964, "interval_mg": [170.517649, 189.827401], "p_within": 0.971554, "w_sat": 0.049436, "u_w_sat": 0.001921, "w_water": 0.447527, "crystallised_mg": 181.077915}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.038782, "unit": "1", "c_yield": -121.245836, "u_yield_mg": 4.702128, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.911154}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -357.32293, "u_yield_mg": 1.031502, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.043847}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 296.908498, "u_yield_mg": 0.857101, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.030274}, {"input": "loss_rel", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -181.077915, "u_yield_mg": 0.522727, "basis": "ASSUMED: up to 1 % of the crystals lost on decanting", "difference": "central", "share": 0.01126}, {"input": "eth_water", "dist": "rect", "u": 0.000577, "unit": "1", "c_yield": -333.748531, "u_yield_mg": 0.19269, "basis": "the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)", "difference": "central", "share": 0.00153}, {"input": "block_offset_K", "dist": "normal", "u": 0.082763, "unit": "K", "c_yield": -2.143361, "u_yield_mg": 0.177391, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)", "difference": "central", "share": 0.001297}, {"input": "charge_mg", "dist": "normal", "u": 0.123558, "unit": "mg", "c_yield": 0.984068, "u_yield_mg": 0.12159, "basis": "the charge weighed by difference on the balance", "difference": "central", "share": 0.000609}, {"input": "water_g", "dist": "normal", "u": 0.000124, "unit": "g", "c_yield": -201.448595, "u_yield_mg": 0.024891, "basis": "the water weighed by difference", "difference": "central", "share": 2.6e-05}]}}, "measurement": {"verdict": "PASS", "checks": [{"id": "spun", "name": "Spun down before the decant", "verdict": "PASS", "reason": "the centrifuge completed a run with this tube before the decant"}, {"id": "constant", "name": "Dried to constant mass", "verdict": "PASS", "reason": "the last two weighings differ by 0.4 mg (pellet) and 0.2 mg (residue); the criterion is 0.5 mg"}, {"id": "cooled", "name": "Cooled before weighing", "verdict": "PASS", "reason": "the vessels were weighed at room temperature"}, {"id": "consistent", "name": "The masses are consistent", "verdict": "PASS", "reason": "the wet pellet is heavier than the dry one, the residue lighter than the supernatant"}], "flags": ["The supernatant residue (109.8 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %"], "y_mg": 184.876303, "u_mg": 0.229054, "U_mg": 0.458108, "w": 0.048886, "u_w": 8.4e-05, "masses_mg": {"wet_pellet": 286.4, "dry_pellet": 190.2, "supernatant": 2239.3, "residue": 109.8}, "corrections": {"buoyancy_kcl": 1.000452, "drying_mg": 0.25, "occluded_rel": 0.001}, "k": 2, "method": "Y = m_dry - (m_wet - m_dry) w / (1 - w), w = m_residue / m_supernatant; masses by difference on the analytical balance, converted from conventional to true mass; the constant-mass criterion and the occluded liquor as rectangular corrections", "budget": [{"input": "solvent left in the pellet (constant-mass criterion, rectangular 0-0.5 mg)", "u": 0.144338, "u_yield_mg": 0.151673, "share": 0.438472}, {"input": "mother liquor occluded in the crystals (ASSUMED 0-0.2 %)", "u": 0.000577, "u_yield_mg": 0.115357, "share": 0.253635}, {"input": "the dry pellet's last weighing", "u": 0.087369, "u_yield_mg": 0.091809, "share": 0.160656}, {"input": "the tube's tare", "u": 0.087369, "u_yield_mg": 0.087316, "share": 0.145317}, {"input": "solvent left in the residue (constant-mass criterion)", "u": 0.144338, "u_yield_mg": 0.006884, "share": 0.000903}, {"input": "the wet pellet's weighing", "u": 0.087369, "u_yield_mg": 0.004493, "share": 0.000385}]}, "next": null}], "readings": [{"k": 0, "do": "read_temperature", "t_min": 0.0, "instrument": "thermometer", "display": "25.35 C", "temperature_C": 25.35, "message": "a probe in a water-filled tube in the block set to 25.5 C"}, {"k": 1, "do": "tare", "t_min": 2.0, "vessel": "T1", "instrument": "balance", "display": "6.8678 g", "mass_mg": 6867.8, "stable": true, "warm": false}, {"k": 2, "do": "tare", "t_min": 3.0, "vessel": "S1", "instrument": "balance", "display": "14.5263 g", "mass_mg": 14526.3, "stable": true, "warm": false}, {"k": 3, "do": "weigh_in", "t_min": 4.0, "vessel": "T1", "instrument": "balance", "display": "7.1677 g", "mass_mg": 7167.7, "stable": true, "message": "KCl weighed in on the balance (target 300 mg)"}, {"k": 4, "do": "pipette", "t_min": 7.0, "vessel": "T1", "instrument": "P1000", "message": "1000 uL of water set on the P1000, forward technique"}, {"k": 5, "do": "weigh", "t_min": 8.0, "vessel": "T1", "instrument": "balance", "display": "8.1617 g", "mass_mg": 8161.7, "stable": true, "warm": false}, {"k": 6, "do": "vortex", "t_min": 9.0, "vessel": "T1", "instrument": "vortex", "message": "60 s"}, {"k": 7, "do": "block", "t_min": 10.0, "vessel": "T1", "instrument": "thermoblock", "display": "25.5 C", "message": "10 min at 25.5 C, shaking 500 rpm"}, {"k": 8, "do": "pipette", "t_min": 20.0, "vessel": "T1", "instrument": "P5000", "message": "1575 uL of ethanol set on the P5000, reverse technique"}, {"k": 9, "do": "weigh", "t_min": 21.0, "vessel": "T1", "instrument": "balance", "display": "9.3934 g", "mass_mg": 9393.4, "stable": true, "warm": false}, {"k": 10, "do": "vortex", "t_min": 22.0, "vessel": "T1", "instrument": "vortex", "message": "10 s"}, {"k": 11, "do": "block", "t_min": 22.17, "vessel": "T1", "instrument": "thermoblock", "display": "25.5 C", "message": "60 min at 25.5 C, shaking 500 rpm"}, {"k": 12, "do": "counterweight", "t_min": 82.17, "vessel": "CW", "instrument": "balance", "display": "9.4413 g", "mass_mg": 9441.3, "message": "filled with water to the mass of T1"}, {"k": 13, "do": "spin", "t_min": 85.17, "tubes": ["T1", "CW"], "instrument": "centrifuge", "log": "4400 rpm, 10 min, 2,965 x g at r_max 13.7 cm", "message": "run complete"}, {"k": 14, "do": "decant", "t_min": 96.0, "vessel": "T1", "into": "S1", "message": "supernatant poured off into the weighing bottle"}, {"k": 15, "do": "weigh", "t_min": 97.0, "vessel": "S1", "instrument": "balance", "display": "16.7656 g", "mass_mg": 16765.6, "stable": true, "warm": false}, {"k": 16, "do": "weigh", "t_min": 98.0, "vessel": "T1", "instrument": "balance", "display": "7.1542 g", "mass_mg": 7154.2, "stable": true, "warm": false}, {"k": 17, "do": "evaporate", "t_min": 99.0, "vessel": "S1", "instrument": "hot plate (in the fume hood)", "message": "120 min, open"}, {"k": 18, "do": "evaporate", "t_min": 219.0, "vessel": "T1", "instrument": "fume hood", "message": "120 min, open"}, {"k": 19, "do": "dry", "t_min": 339.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 20, "do": "desiccate", "t_min": 399.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 21, "do": "weigh", "t_min": 429.0, "vessel": "T1", "instrument": "balance", "display": "7.0613 g", "mass_mg": 7061.3, "stable": true, "warm": false}, {"k": 22, "do": "weigh", "t_min": 430.0, "vessel": "S1", "instrument": "balance", "display": "14.6386 g", "mass_mg": 14638.6, "stable": true, "warm": false}, {"k": 23, "do": "dry", "t_min": 431.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 24, "do": "desiccate", "t_min": 491.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 25, "do": "weigh", "t_min": 521.0, "vessel": "T1", "instrument": "balance", "display": "7.0584 g", "mass_mg": 7058.4, "stable": true, "warm": false}, {"k": 26, "do": "weigh", "t_min": 522.0, "vessel": "S1", "instrument": "balance", "display": "14.6363 g", "mass_mg": 14636.3, "stable": true, "warm": false}, {"k": 27, "do": "dry", "t_min": 523.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 28, "do": "desiccate", "t_min": 583.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 29, "do": "weigh", "t_min": 613.0, "vessel": "T1", "instrument": "balance", "display": "7.0580 g", "mass_mg": 7058.0, "stable": true, "warm": false}, {"k": 30, "do": "weigh", "t_min": 614.0, "vessel": "S1", "instrument": "balance", "display": "14.6361 g", "mass_mg": 14636.1, "stable": true, "warm": false}], "status": {"state": "accepted", "attempt": 1, "text": "tube 1 (T1) accepted: 184.9 mg"}, "g6": {"gate": {"id": "G6", "name": "The evidence package", "verdict": "PASS", "reason": "The package verifies (VALID): every section matches its digest, the goal and the rule are the frozen ones, and running the notebook again gives the same attempts, status and claims.", "basis": "web/kclrecord.py verify()", "numbers": {"status": "VALID"}}, "verification": {"status": "VALID", "checks": [{"id": "digest:inputs", "ok": true, "what": "the section 'inputs' matches its digest"}, {"id": "digest:goal", "ok": true, "what": "the section 'goal' matches its digest"}, {"id": "digest:rule", "ok": true, "what": "the section 'rule' matches its digest"}, {"id": "digest:method", "ok": true, "what": "the section 'method' matches its digest"}, {"id": "digest:attempts", "ok": true, "what": "the section 'attempts' matches its digest"}, {"id": "digest:status", "ok": true, "what": "the section 'status' matches its digest"}, {"id": "digest:claims", "ok": true, "what": "the section 'claims' matches its digest"}, {"id": "digest:not_claimed", "ok": true, "what": "the section 'not_claimed' matches its digest"}, {"id": "digest:references", "ok": true, "what": "the section 'references' matches its digest"}, {"id": "digest:reproduce", "ok": true, "what": "the section 'reproduce' matches its digest"}, {"id": "digest:engine", "ok": true, "what": "the section 'engine' matches its digest"}, {"id": "manifest", "ok": true, "what": "the manifest matches its digest"}, {"id": "goal", "ok": true, "what": "the goal is the frozen one"}, {"id": "rule", "ok": true, "what": "the rule is the frozen one"}, {"id": "reproduce:attempts", "ok": true, "what": "running the notebook again gives the same attempts"}, {"id": "reproduce:status", "ok": true, "what": "running the notebook again gives the same status"}, {"id": "reproduce:claims", "ok": true, "what": "running the notebook again gives the same claims"}]}, "claims": [{"claim": "Tube 1 met the goal: 184.9 mg of dry KCl crystals (U = 0.5 mg, k = 2) from a 299.9 mg charge, accepted by the frozen rule.", "rests_on": ["G0", "G1", "G2", "G3", "G5", "D"], "qualifier": "virtually qualified: a simulated laboratory and its virtual instruments; nothing measured"}, {"claim": "The model's prediction for tube 1 as made (180.2 mg) agrees with its measurement (zeta +1.0).", "rests_on": ["G4", "G5", "D"], "qualifier": "virtually qualified: a simulated laboratory and its virtual instruments; nothing measured"}], "not_claimed": ["the crystal size or the onset of crystallisation (the model's size is likely too large; neither is measured here)", "that the hidden laboratory is a real one: it is a simulated laboratory whose differences from the model come from the literature", "the model's table validated at this composition: agreement in a virtual laboratory tests the procedure and the decision, not the table", "any yield at another temperature, charge or antisolvent than the goal's", "any physical measurement"], "manifest_digest": "032b36047f2b9fc708f5327541c51b22822a811edc018bf5522e26953275da0e"}}, "wrong-ethanol": {"words": {"title": "The ethanol bottle is not what its label says (hidden)", "lead": "The bottle labelled 'Ethanol absolute' holds Ethanol 96 % (Ph. Eur.), 5-7 % water by mass: a common mix-up. The analyst is not told; the readings must reveal it."}, "hidden": true, "actions": [{"do": "read_temperature", "where": "block", "set_C": 25.5}, {"do": "tare", "vessel": "T1"}, {"do": "tare", "vessel": "S1"}, {"do": "weigh_in", "vessel": "T1", "target_mg": 300.0}, {"do": "pipette", "vessel": "T1", "liquid": "water", "uL": 1000.0, "pipette": "P1000", "technique": "forward"}, {"do": "weigh", "vessel": "T1"}, {"do": "vortex", "vessel": "T1", "s": 60}, {"do": "block", "vessel": "T1", "set_C": 25.5, "min": 10}, {"do": "pipette", "vessel": "T1", "liquid": "ethanol", "uL": 1575.0, "pipette": "P5000", "technique": "reverse"}, {"do": "weigh", "vessel": "T1"}, {"do": "vortex", "vessel": "T1", "s": 10}, {"do": "block", "vessel": "T1", "set_C": 25.5, "min": 60}, {"do": "counterweight", "vessel": "CW", "match": "T1"}, {"do": "spin", "tubes": ["T1", "CW"], "rpm": 4400, "min": 10, "set_C": 25.5}, {"do": "decant", "vessel": "T1", "into": "S1"}, {"do": "weigh", "vessel": "S1"}, {"do": "weigh", "vessel": "T1"}, {"do": "evaporate", "vessel": "S1", "where": "hotplate", "min": 120}, {"do": "evaporate", "vessel": "T1", "where": "hood", "min": 120}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}, {"do": "dry", "vessels": ["T1", "S1"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T1", "S1"], "min": 30}, {"do": "weigh", "vessel": "T1"}, {"do": "weigh", "vessel": "S1"}, {"do": "read_density", "sample": "ethanol"}, {"do": "tare", "vessel": "T2"}, {"do": "tare", "vessel": "S2"}, {"do": "weigh_in", "vessel": "T2", "target_mg": 300.0}, {"do": "pipette", "vessel": "T2", "liquid": "water", "uL": 1000.0, "pipette": "P1000", "technique": "forward"}, {"do": "weigh", "vessel": "T2"}, {"do": "vortex", "vessel": "T2", "s": 60}, {"do": "block", "vessel": "T2", "set_C": 25.5, "min": 10}, {"do": "pipette", "vessel": "T2", "liquid": "ethanol", "uL": 2170.0, "pipette": "P5000", "technique": "reverse"}, {"do": "weigh", "vessel": "T2"}, {"do": "vortex", "vessel": "T2", "s": 10}, {"do": "block", "vessel": "T2", "set_C": 25.5, "min": 60}, {"do": "counterweight", "vessel": "CW", "match": "T2"}, {"do": "spin", "tubes": ["T2", "CW"], "rpm": 4400, "min": 10, "set_C": 25.5}, {"do": "decant", "vessel": "T2", "into": "S2"}, {"do": "weigh", "vessel": "S2"}, {"do": "weigh", "vessel": "T2"}, {"do": "evaporate", "vessel": "S2", "where": "hotplate", "min": 120}, {"do": "evaporate", "vessel": "T2", "where": "hood", "min": 120}, {"do": "dry", "vessels": ["T2", "S2"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T2", "S2"], "min": 30}, {"do": "weigh", "vessel": "T2"}, {"do": "weigh", "vessel": "S2"}, {"do": "dry", "vessels": ["T2", "S2"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T2", "S2"], "min": 30}, {"do": "weigh", "vessel": "T2"}, {"do": "weigh", "vessel": "S2"}, {"do": "dry", "vessels": ["T2", "S2"], "set_C": 105, "min": 60}, {"do": "desiccate", "vessels": ["T2", "S2"], "min": 30}, {"do": "weigh", "vessel": "T2"}, {"do": "weigh", "vessel": "S2"}], "bench_digest": "0c34a934ea800c6e", "attempts": [{"n": 1, "tube": "T1", "bottle": "S1", "recipe": {"tube": "T1", "charge_mg": 300.0, "water_uL": 1000.0, "water_pipette": "P1000", "ethanol_uL": 1575.0, "ethanol_by": "pipette", "block_C": 25.5, "ethanol_pipette": "P5000", "technique": "reverse", "spin": {"rpm": 4400.0, "min": 10.0, "set_C": 25.5}}, "belief": {"ethanol_water": {"provenance": "ASSUMED", "dist": "rect", "lo": 0.0, "hi": 0.002, "basis": "the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)"}, "block_offset_K": {"provenance": "MEASURED", "dist": "normal", "mean": -0.15, "sd": 0.082763, "from_attempt": 0, "set_C": 25.5, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)"}}, "made": {"charge_mg": 299.9, "water_mg": 994.0, "ethanol_mg": 1263.4, "u_mg": 0.123558, "gross_mg": {"tare": 6867.8, "charge": 7167.7, "water": 8161.7, "ethanol": 9425.1}}, "gates": [{"id": "G1", "name": "Recipe meets the goal", "verdict": "PASS", "reason": "With 1575 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg.", "basis": "the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief", "numbers": {"yield_mg": 180.023526, "gap_mg": 0.023526, "ethanol_uL": 1575.0}}, {"id": "G2", "name": "Feasible in this laboratory", "verdict": "PASS", "reason": "Every step is within the instruments' specifications and the model's domain.", "basis": "vialbot/instruments/labkit.py specifications; the table's domain", "numbers": {"problems": []}}, {"id": "G3", "name": "Performed as ordered", "verdict": "PASS", "reason": "Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles.", "basis": "the instruments' own logs, never the hidden laboratory", "numbers": {"missing": [], "recovered": []}}, {"id": "G4", "name": "Prediction and its uncertainty", "verdict": "PASS", "reason": "The model expects 180.0 mg with a standard uncertainty of 5.4 mg (95.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 76 %, eth_sys_uL 5 %, charge_mg 4 %. The Monte Carlo method (269,994 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 169.4-190.5 mg, 95.5 % within the tolerance.", "basis": "JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)", "numbers": {"y_mg": 180.023526, "u_mg": 5.400433, "interval_mg": [169.438873, 190.60818], "p_within": 0.95448, "w_sat": 0.049302, "u_w_sat": 0.002218, "as_made": false}}, {"id": "G5", "name": "The instruments' evidence", "verdict": "PASS", "reason": "Measured yield 161.77 mg (U = 0.45 mg, k = 2); the supernatant held 5.737 % KCl by mass. The supernatant residue (132.4 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %.", "basis": "drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'", "numbers": {"y_mg": 161.768445, "u_mg": 0.224106, "w": 0.05737, "u_w": 8.2e-05, "masses_mg": {"wet_pellet": 255.6, "dry_pellet": 167.5, "supernatant": 2301.8, "residue": 132.4}}}, {"id": "D", "name": "Decision", "verdict": "FAIL", "reason": "Measured 161.77 mg (u 0.22 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. REJECT. Against the prediction for the tube as made: zeta -4.2 for the yield, +5.1 for the supernatant. The densitometer reads 0.8095 g/cm3 where absolute ethanol gives 0.7878 (zeta +27): the bottle holds 7.6 +/- 0.3 % water. The next tube takes 2170 uL of it.", "basis": "JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection", "numbers": {"outcome": "reject", "y_mg": 161.768445, "u_mg": 0.224106, "p_conform": 0.0, "acceptance_mg": [169.558947, 190.441053], "rejection_below_mg": 168.841053, "rejection_above_mg": 191.158947, "zeta_yield": -4.237308, "zeta_w": 5.062953, "predicted_mg": 182.361418, "u_predicted_mg": 4.854749, "predicted_w": 0.047894, "fills": [{"liquid": "water", "weighed_uL": 994.0, "u_weighed_uL": 0.123558, "predicted_uL": 997.772977, "u_predicted_uL": 5.495305, "zeta": -0.686408, "agree": true}, {"liquid": "ethanol", "weighed_uL": 1263.4, "u_weighed_uL": 0.123558, "predicted_uL": 1239.104047, "u_predicted_uL": 21.724076, "zeta": 1.11837, "agree": true}], "density_check": {"k": 31, "density_g_cm3": 0.8095, "temperature_C": 22.1, "expected_g_cm3": 0.787829, "u_g_cm3": 0.00079, "zeta": 27.415241, "water_mass_fraction": 0.075532, "u_water": 0.002812, "signal": "action"}, "thermometer_check": {"k": null, "temperature_C": 25.35, "offset_K": -0.15, "u_K": 0.082763, "zeta": 0.0, "signal": "agree", "basis": "measured before the tube (the block check)"}, "cause": "ethanol", "standardised": {"water_mass_fraction": 0.075532, "u": 0.002812}}}], "prediction": {"planned": {"first_order": {"y_mg": 180.023526, "u_mg": 5.400433, "k": 1.959964, "interval_mg": [169.438873, 190.60818], "p_within": 0.95448, "w_sat": 0.049302, "u_w_sat": 0.002218, "w_water": 0.446981, "crystallised_mg": 180.928167}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.038835, "unit": "1", "c_yield": -121.489054, "u_yield_mg": 4.718041, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.763249}, {"input": "eth_sys_uL", "dist": "rect", "u": 23.094011, "unit": "uL", "c_yield": 0.054452, "u_yield_mg": 1.257519, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.054222}, {"input": "charge_mg", "dist": "rect", "u": 1.154701, "unit": "mg", "c_yield": 0.984068, "u_yield_mg": 1.136304, "basis": "ASSUMED: a charge weighed in by spatula within +/-2 mg of the target", "difference": "central", "share": 0.044272}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -357.429387, "u_yield_mg": 1.03181, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.036504}, {"input": "water_sys_uL", "dist": "rect", "u": 4.618802, "unit": "uL", "c_yield": -0.200958, "u_yield_mg": 0.928184, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.02954}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 297.0075, "u_yield_mg": 0.857387, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.025206}, {"input": "eth_rand_uL", "dist": "normal", "u": 15.0, "unit": "uL", "c_yield": 0.054452, "u_yield_mg": 0.816776, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.022874}, {"input": "water_rand_uL", "dist": "normal", "u": 3.0, "unit": "uL", "c_yield": -0.200958, "u_yield_mg": 0.602873, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.012462}], "mcm": {"trials": 269994, "batches": 27, "batch_size": 10000, "seed": 20261023, "stable": true, "tolerance_mg": 0.05, "refused_draws": 6, "y_mg": 179.995815, "u_mg": 5.399429, "interval_mg": [169.361458, 190.465582], "p_within": 0.955095, "p_se": 0.000399}, "validation": {"tolerance_mg": 0.05, "d_low_mg": 0.077415, "d_high_mg": 0.142599, "validated": false, "basis": "JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)"}}, "as_made": {"first_order": {"y_mg": 182.361418, "u_mg": 4.854749, "k": 1.959964, "interval_mg": [172.846285, 191.876551], "p_within": 0.95556, "w_sat": 0.047894, "u_w_sat": 0.00187, "w_water": 0.44126, "crystallised_mg": 183.277807}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.038961, "unit": "1", "c_yield": -118.750441, "u_yield_mg": 4.626663, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.908243}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -357.176809, "u_yield_mg": 1.031081, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.045108}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 296.908498, "u_yield_mg": 0.857101, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.03117}, {"input": "loss_rel", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -183.277807, "u_yield_mg": 0.529077, "basis": "ASSUMED: up to 1 % of the crystals lost on decanting", "difference": "central", "share": 0.011877}, {"input": "eth_water", "dist": "rect", "u": 0.000577, "unit": "1", "c_yield": -341.230217, "u_yield_mg": 0.197009, "basis": "the label: ethanol absolute, water <= 0.2 % by mass (ACS reagent specification)", "difference": "central", "share": 0.001647}, {"input": "block_offset_K", "dist": "normal", "u": 0.082763, "unit": "K", "c_yield": -2.114107, "u_yield_mg": 0.17497, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)", "difference": "central", "share": 0.001299}, {"input": "charge_mg", "dist": "normal", "u": 0.123558, "unit": "mg", "c_yield": 0.98407, "u_yield_mg": 0.12159, "basis": "the charge weighed by difference on the balance", "difference": "central", "share": 0.000627}, {"input": "water_g", "dist": "normal", "u": 0.000124, "unit": "g", "c_yield": -200.961362, "u_yield_mg": 0.02483, "basis": "the water weighed by difference", "difference": "central", "share": 2.6e-05}]}}, "measurement": {"verdict": "PASS", "checks": [{"id": "spun", "name": "Spun down before the decant", "verdict": "PASS", "reason": "the centrifuge completed a run with this tube before the decant"}, {"id": "constant", "name": "Dried to constant mass", "verdict": "PASS", "reason": "the last two weighings differ by 0.4 mg (pellet) and 0.3 mg (residue); the criterion is 0.5 mg"}, {"id": "cooled", "name": "Cooled before weighing", "verdict": "PASS", "reason": "the vessels were weighed at room temperature"}, {"id": "consistent", "name": "The masses are consistent", "verdict": "PASS", "reason": "the wet pellet is heavier than the dry one, the residue lighter than the supernatant"}], "flags": ["The supernatant residue (132.4 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %"], "y_mg": 161.768445, "u_mg": 0.224106, "U_mg": 0.448212, "w": 0.05737, "u_w": 8.2e-05, "masses_mg": {"wet_pellet": 255.6, "dry_pellet": 167.5, "supernatant": 2301.8, "residue": 132.4}, "corrections": {"buoyancy_kcl": 1.000452, "drying_mg": 0.25, "occluded_rel": 0.001}, "k": 2, "method": "Y = m_dry - (m_wet - m_dry) w / (1 - w), w = m_residue / m_supernatant; masses by difference on the analytical balance, converted from conventional to true mass; the constant-mass criterion and the occluded liquor as rectangular corrections", "budget": [{"input": "solvent left in the pellet (constant-mass criterion, rectangular 0-0.5 mg)", "u": 0.144338, "u_yield_mg": 0.153038, "share": 0.466329}, {"input": "mother liquor occluded in the crystals (ASSUMED 0-0.2 %)", "u": 0.000577, "u_yield_mg": 0.102485, "share": 0.209129}, {"input": "the dry pellet's last weighing", "u": 0.087369, "u_yield_mg": 0.092635, "share": 0.170863}, {"input": "the tube's tare", "u": 0.087369, "u_yield_mg": 0.087316, "share": 0.151802}, {"input": "solvent left in the residue (constant-mass criterion)", "u": 0.144338, "u_yield_mg": 0.006245, "share": 0.000777}, {"input": "the wet pellet's weighing", "u": 0.087369, "u_yield_mg": 0.00532, "share": 0.000563}]}, "next": {"cause": "ethanol", "recipe": {"tube": "T1", "charge_mg": 300.0, "water_uL": 1000.0, "water_pipette": "P1000", "ethanol_uL": 2170.0, "ethanol_by": "pipette", "block_C": 25.5, "ethanol_pipette": "P5000", "technique": "reverse", "spin": {"rpm": 4400.0, "min": 10.0, "set_C": 25.5}}, "belief": {"ethanol_water": {"provenance": "MEASURED", "dist": "normal", "mean": 0.075532, "sd": 0.002812, "from_attempt": 1, "basis": "the bottle's density 0.8095 g/cm3 at 22.1 C, through the published table"}, "block_offset_K": {"provenance": "MEASURED", "dist": "normal", "mean": -0.15, "sd": 0.082763, "from_attempt": 0, "set_C": 25.5, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)"}}, "plan": {"ethanol_uL": 2170.0, "exact_uL": 2169.777599, "step_uL": 5.0, "expected_mg": 180.008497}, "why": "the bottle holds 7.6 % water, not <= 0.2 %"}}, {"n": 2, "tube": "T2", "bottle": "S2", "recipe": {"tube": "T2", "charge_mg": 300.0, "water_uL": 1000.0, "water_pipette": "P1000", "ethanol_uL": 2170.0, "ethanol_by": "pipette", "block_C": 25.5, "ethanol_pipette": "P5000", "technique": "reverse", "spin": {"rpm": 4400.0, "min": 10.0, "set_C": 25.5}}, "belief": {"ethanol_water": {"provenance": "MEASURED", "dist": "normal", "mean": 0.075532, "sd": 0.002812, "from_attempt": 1, "basis": "the bottle's density 0.8095 g/cm3 at 22.1 C, through the published table"}, "block_offset_K": {"provenance": "MEASURED", "dist": "normal", "mean": -0.15, "sd": 0.082763, "from_attempt": 0, "set_C": 25.5, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)"}}, "made": {"charge_mg": 297.9, "water_mg": 994.6, "ethanol_mg": 1733.7, "u_mg": 0.123558, "gross_mg": {"tare": 6937.7, "charge": 7235.6, "water": 8230.2, "ethanol": 9963.9}}, "gates": [{"id": "G1", "name": "Recipe meets the goal", "verdict": "PASS", "reason": "With 2170 uL of ethanol the model expects 180.0 mg of dry crystals; the goal is 180 mg.", "basis": "the frozen world model (Pinho & Macedo table, linear interpolation) on the recipe and the current belief", "numbers": {"yield_mg": 180.008497, "gap_mg": 0.008497, "ethanol_uL": 2170.0}}, {"id": "G2", "name": "Feasible in this laboratory", "verdict": "PASS", "reason": "Every step is within the instruments' specifications and the model's domain.", "basis": "vialbot/instruments/labkit.py specifications; the table's domain", "numbers": {"problems": []}}, {"id": "G3", "name": "Performed as ordered", "verdict": "PASS", "reason": "Every instrument acknowledged its step: the charge weighed, the liquids added, the block held, the centrifuge completed its run, the oven its cycles.", "basis": "the instruments' own logs, never the hidden laboratory", "numbers": {"missing": [], "recovered": []}}, {"id": "G4", "name": "Prediction and its uncertainty", "verdict": "PASS", "reason": "The model expects 180.0 mg with a standard uncertainty of 3.9 mg (99.4 % probability of landing within 180 mg +/- 6 %); the largest components: table_rel 53 %, eth_water 10 %, charge_mg 8 %. The Monte Carlo method (209,996 draws) does not validate the first-order interval; its own result stands: 180.0 mg, interval 172.3-187.6 mg, 99.5 % within the tolerance.", "basis": "JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G; JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)", "numbers": {"y_mg": 180.008497, "u_mg": 3.908867, "interval_mg": [172.347259, 187.669736], "p_within": 0.994272, "w_sat": 0.040458, "u_w_sat": 0.001265, "as_made": false}}, {"id": "G5", "name": "The instruments' evidence", "verdict": "PASS", "reason": "Measured yield 181.93 mg (U = 0.45 mg, k = 2); the supernatant held 4.063 % KCl by mass. The supernatant residue (111.9 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %.", "basis": "drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention); USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'", "numbers": {"y_mg": 181.933488, "u_mg": 0.226121, "w": 0.040634, "u_w": 6.9e-05, "masses_mg": {"wet_pellet": 280.6, "dry_pellet": 186.3, "supernatant": 2745.7, "residue": 111.9}}}, {"id": "D", "name": "Decision", "verdict": "PASS", "reason": "Measured 181.93 mg (u 0.23 mg): accept between 169.56 and 190.44 mg, reject below 168.84 or above 191.16 mg. ACCEPT. Against the prediction for the tube as made: zeta +1.2 for the yield, -0.2 for the supernatant.", "basis": "JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection; ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal", "numbers": {"outcome": "accept", "y_mg": 181.933488, "u_mg": 0.226121, "p_conform": 1.0, "acceptance_mg": [169.558947, 190.441053], "rejection_below_mg": 168.841053, "rejection_above_mg": 191.158947, "zeta_yield": 1.188756, "zeta_w": -0.240985, "predicted_mg": 177.603936, "u_predicted_mg": 3.63506, "predicted_w": 0.040908, "fills": [{"liquid": "water", "weighed_uL": 994.6, "u_weighed_uL": 0.123558, "predicted_uL": 997.772977, "u_predicted_uL": 5.495305, "zeta": -0.577252, "agree": true}, {"liquid": "ethanol", "weighed_uL": 1733.7, "u_weighed_uL": 0.123558, "predicted_uL": 1754.168151, "u_predicted_uL": 22.346136, "zeta": -0.915945, "agree": true}], "cause": null}}], "prediction": {"planned": {"first_order": {"y_mg": 180.008497, "u_mg": 3.908867, "k": 1.959964, "interval_mg": [172.347259, 187.669736], "p_within": 0.994272, "w_sat": 0.040458, "u_w_sat": 0.001265, "w_water": 0.411038, "crystallised_mg": 180.913063}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.023588, "unit": "1", "c_yield": -120.384546, "u_yield_mg": 2.839672, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.527758}, {"input": "eth_water", "dist": "normal", "u": 0.002812, "unit": "1", "c_yield": -438.405642, "u_yield_mg": 1.232811, "basis": "the bottle's density 0.8095 g/cm3 at 22.1 C, through the published table", "difference": "central", "share": 0.09947}, {"input": "charge_mg", "dist": "rect", "u": 1.154701, "unit": "mg", "c_yield": 0.984082, "u_yield_mg": 1.13632, "basis": "ASSUMED: a charge weighed in by spatula within +/-2 mg of the target", "difference": "central", "share": 0.084508}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -356.578041, "u_yield_mg": 1.029352, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.069347}, {"input": "water_sys_uL", "dist": "rect", "u": 4.618802, "unit": "uL", "c_yield": -0.198126, "u_yield_mg": 0.915106, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.054808}, {"input": "eth_sys_uL", "dist": "rect", "u": 23.094011, "unit": "uL", "c_yield": 0.038208, "u_yield_mg": 0.88237, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.050957}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 297.0075, "u_yield_mg": 0.857387, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.048112}, {"input": "water_rand_uL", "dist": "normal", "u": 3.0, "unit": "uL", "c_yield": -0.198126, "u_yield_mg": 0.594379, "basis": "ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL", "difference": "central", "share": 0.023122}], "mcm": {"trials": 209996, "batches": 21, "batch_size": 10000, "seed": 20261023, "stable": true, "tolerance_mg": 0.05, "refused_draws": 4, "y_mg": 179.990457, "u_mg": 3.913171, "interval_mg": [172.307661, 187.591254], "p_within": 0.994509, "p_se": 0.000161}, "validation": {"tolerance_mg": 0.05, "d_low_mg": 0.039598, "d_high_mg": 0.078482, "validated": false, "basis": "JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)"}}, "as_made": {"first_order": {"y_mg": 177.603936, "u_mg": 3.63506, "k": 1.959964, "interval_mg": [170.479348, 184.728523], "p_within": 0.989467, "w_sat": 0.040908, "u_w_sat": 0.001132, "w_water": 0.412867, "crystallised_mg": 178.496418}, "budget": [{"input": "table_rel", "dist": "normal", "u": 0.025443, "unit": "1", "c_yield": -120.791285, "u_yield_mg": 3.073245, "basis": "Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates: the table's expanded uncertainty (k = 2) at the end composition, which includes the interpolation's error", "difference": "central", "share": 0.714778}, {"input": "eth_water", "dist": "normal", "u": 0.002812, "unit": "1", "c_yield": -461.455739, "u_yield_mg": 1.297629, "basis": "the bottle's density 0.8095 g/cm3 at 22.1 C, through the published table", "difference": "central", "share": 0.127432}, {"input": "moisture", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -354.125719, "u_yield_mg": 1.022273, "basis": "USP Potassium Chloride: loss on drying <= 1.0 % at 105 C", "difference": "central", "share": 0.079088}, {"input": "purity", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": 294.928448, "u_yield_mg": 0.851385, "basis": "USP/ACS assay 99.0-100.5 % on the dried basis, capped at 100 %", "difference": "central", "share": 0.054857}, {"input": "loss_rel", "dist": "rect", "u": 0.002887, "unit": "1", "c_yield": -178.496418, "u_yield_mg": 0.515275, "basis": "ASSUMED: up to 1 % of the crystals lost on decanting", "difference": "central", "share": 0.020093}, {"input": "block_offset_K", "dist": "normal", "u": 0.082763, "unit": "K", "c_yield": -2.23321, "u_yield_mg": 0.184827, "basis": "the reference thermometer read 25.35 C in the block set to 25.5 C before the first tube (IEC 60751 class AA)", "difference": "central", "share": 0.002585}, {"input": "charge_mg", "dist": "normal", "u": 0.123558, "unit": "mg", "c_yield": 0.984081, "u_yield_mg": 0.121591, "basis": "the charge weighed by difference on the balance", "difference": "central", "share": 0.001119}, {"input": "water_g", "dist": "normal", "u": 0.000124, "unit": "g", "c_yield": -198.715245, "u_yield_mg": 0.024553, "basis": "the water weighed by difference", "difference": "central", "share": 4.6e-05}]}}, "measurement": {"verdict": "PASS", "checks": [{"id": "spun", "name": "Spun down before the decant", "verdict": "PASS", "reason": "the centrifuge completed a run with this tube before the decant"}, {"id": "constant", "name": "Dried to constant mass", "verdict": "PASS", "reason": "the last two weighings differ by 0.3 mg (pellet) and 0.2 mg (residue); the criterion is 0.5 mg"}, {"id": "cooled", "name": "Cooled before weighing", "verdict": "PASS", "reason": "the vessels were weighed at room temperature"}, {"id": "consistent", "name": "The masses are consistent", "verdict": "PASS", "reason": "the wet pellet is heavier than the dry one, the residue lighter than the supernatant"}], "flags": ["The supernatant residue (111.9 mg) is below the balance's typical USP <41> minimum weight (160 mg): its relative uncertainty is larger than 0.10 %"], "y_mg": 181.933488, "u_mg": 0.226121, "U_mg": 0.452242, "w": 0.040634, "u_w": 6.9e-05, "masses_mg": {"wet_pellet": 280.6, "dry_pellet": 186.3, "supernatant": 2745.7, "residue": 111.9}, "corrections": {"buoyancy_kcl": 1.000452, "drying_mg": 0.25, "occluded_rel": 0.001}, "k": 2, "method": "Y = m_dry - (m_wet - m_dry) w / (1 - w), w = m_residue / m_supernatant; masses by difference on the analytical balance, converted from conventional to true mass; the constant-mass criterion and the occluded liquor as rectangular corrections", "budget": [{"input": "solvent left in the pellet (constant-mass criterion, rectangular 0-0.5 mg)", "u": 0.144338, "u_yield_mg": 0.150368, "share": 0.442213}, {"input": "mother liquor occluded in the crystals (ASSUMED 0-0.2 %)", "u": 0.000577, "u_yield_mg": 0.112016, "share": 0.245403}, {"input": "the dry pellet's last weighing", "u": 0.087369, "u_yield_mg": 0.091019, "share": 0.162027}, {"input": "the tube's tare", "u": 0.087369, "u_yield_mg": 0.087317, "share": 0.149114}, {"input": "solvent left in the residue (constant-mass criterion)", "u": 0.144338, "u_yield_mg": 0.005409, "share": 0.000572}, {"input": "the wet pellet's weighing", "u": 0.087369, "u_yield_mg": 0.003702, "share": 0.000268}]}, "next": null}], "readings": [{"k": 0, "do": "read_temperature", "t_min": 0.0, "instrument": "thermometer", "display": "25.35 C", "temperature_C": 25.35, "message": "a probe in a water-filled tube in the block set to 25.5 C"}, {"k": 1, "do": "tare", "t_min": 2.0, "vessel": "T1", "instrument": "balance", "display": "6.8678 g", "mass_mg": 6867.8, "stable": true, "warm": false}, {"k": 2, "do": "tare", "t_min": 3.0, "vessel": "S1", "instrument": "balance", "display": "14.5263 g", "mass_mg": 14526.3, "stable": true, "warm": false}, {"k": 3, "do": "weigh_in", "t_min": 4.0, "vessel": "T1", "instrument": "balance", "display": "7.1677 g", "mass_mg": 7167.7, "stable": true, "message": "KCl weighed in on the balance (target 300 mg)"}, {"k": 4, "do": "pipette", "t_min": 7.0, "vessel": "T1", "instrument": "P1000", "message": "1000 uL of water set on the P1000, forward technique"}, {"k": 5, "do": "weigh", "t_min": 8.0, "vessel": "T1", "instrument": "balance", "display": "8.1617 g", "mass_mg": 8161.7, "stable": true, "warm": false}, {"k": 6, "do": "vortex", "t_min": 9.0, "vessel": "T1", "instrument": "vortex", "message": "60 s"}, {"k": 7, "do": "block", "t_min": 10.0, "vessel": "T1", "instrument": "thermoblock", "display": "25.5 C", "message": "10 min at 25.5 C, shaking 500 rpm"}, {"k": 8, "do": "pipette", "t_min": 20.0, "vessel": "T1", "instrument": "P5000", "message": "1575 uL of ethanol set on the P5000, reverse technique"}, {"k": 9, "do": "weigh", "t_min": 21.0, "vessel": "T1", "instrument": "balance", "display": "9.4251 g", "mass_mg": 9425.1, "stable": true, "warm": false}, {"k": 10, "do": "vortex", "t_min": 22.0, "vessel": "T1", "instrument": "vortex", "message": "10 s"}, {"k": 11, "do": "block", "t_min": 22.17, "vessel": "T1", "instrument": "thermoblock", "display": "25.5 C", "message": "60 min at 25.5 C, shaking 500 rpm"}, {"k": 12, "do": "counterweight", "t_min": 82.17, "vessel": "CW", "instrument": "balance", "display": "9.4730 g", "mass_mg": 9473.0, "message": "filled with water to the mass of T1"}, {"k": 13, "do": "spin", "t_min": 85.17, "tubes": ["T1", "CW"], "instrument": "centrifuge", "log": "4400 rpm, 10 min, 2,965 x g at r_max 13.7 cm", "message": "run complete"}, {"k": 14, "do": "decant", "t_min": 96.0, "vessel": "T1", "into": "S1", "message": "supernatant poured off into the weighing bottle"}, {"k": 15, "do": "weigh", "t_min": 97.0, "vessel": "S1", "instrument": "balance", "display": "16.8281 g", "mass_mg": 16828.1, "stable": true, "warm": false}, {"k": 16, "do": "weigh", "t_min": 98.0, "vessel": "T1", "instrument": "balance", "display": "7.1234 g", "mass_mg": 7123.4, "stable": true, "warm": false}, {"k": 17, "do": "evaporate", "t_min": 99.0, "vessel": "S1", "instrument": "hot plate (in the fume hood)", "message": "120 min, open"}, {"k": 18, "do": "evaporate", "t_min": 219.0, "vessel": "T1", "instrument": "fume hood", "message": "120 min, open"}, {"k": 19, "do": "dry", "t_min": 339.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 20, "do": "desiccate", "t_min": 399.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 21, "do": "weigh", "t_min": 429.0, "vessel": "T1", "instrument": "balance", "display": "7.0384 g", "mass_mg": 7038.4, "stable": true, "warm": false}, {"k": 22, "do": "weigh", "t_min": 430.0, "vessel": "S1", "instrument": "balance", "display": "14.6613 g", "mass_mg": 14661.3, "stable": true, "warm": false}, {"k": 23, "do": "dry", "t_min": 431.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 24, "do": "desiccate", "t_min": 491.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 25, "do": "weigh", "t_min": 521.0, "vessel": "T1", "instrument": "balance", "display": "7.0357 g", "mass_mg": 7035.7, "stable": true, "warm": false}, {"k": 26, "do": "weigh", "t_min": 522.0, "vessel": "S1", "instrument": "balance", "display": "14.6590 g", "mass_mg": 14659.0, "stable": true, "warm": false}, {"k": 27, "do": "dry", "t_min": 523.0, "vessels": ["T1", "S1"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 28, "do": "desiccate", "t_min": 583.0, "vessels": ["T1", "S1"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 29, "do": "weigh", "t_min": 613.0, "vessel": "T1", "instrument": "balance", "display": "7.0353 g", "mass_mg": 7035.3, "stable": true, "warm": false}, {"k": 30, "do": "weigh", "t_min": 614.0, "vessel": "S1", "instrument": "balance", "display": "14.6587 g", "mass_mg": 14658.7, "stable": true, "warm": false}, {"k": 31, "do": "read_density", "t_min": 615.0, "instrument": "densitometer", "display": "0.8095 g/cm3 at 22.1 C", "density_g_cm3": 0.8095, "temperature_C": 22.1, "message": "the 'Ethanol absolute' bottle, at room temperature"}, {"k": 32, "do": "tare", "t_min": 617.0, "vessel": "T2", "instrument": "balance", "display": "6.9377 g", "mass_mg": 6937.7, "stable": true, "warm": false}, {"k": 33, "do": "tare", "t_min": 618.0, "vessel": "S2", "instrument": "balance", "display": "14.7826 g", "mass_mg": 14782.6, "stable": true, "warm": false}, {"k": 34, "do": "weigh_in", "t_min": 619.0, "vessel": "T2", "instrument": "balance", "display": "7.2356 g", "mass_mg": 7235.6, "stable": true, "message": "KCl weighed in on the balance (target 300 mg)"}, {"k": 35, "do": "pipette", "t_min": 622.0, "vessel": "T2", "instrument": "P1000", "message": "1000 uL of water set on the P1000, forward technique"}, {"k": 36, "do": "weigh", "t_min": 623.0, "vessel": "T2", "instrument": "balance", "display": "8.2302 g", "mass_mg": 8230.2, "stable": true, "warm": false}, {"k": 37, "do": "vortex", "t_min": 624.0, "vessel": "T2", "instrument": "vortex", "message": "60 s"}, {"k": 38, "do": "block", "t_min": 625.0, "vessel": "T2", "instrument": "thermoblock", "display": "25.5 C", "message": "10 min at 25.5 C, shaking 500 rpm"}, {"k": 39, "do": "pipette", "t_min": 635.0, "vessel": "T2", "instrument": "P5000", "message": "2170 uL of ethanol set on the P5000, reverse technique"}, {"k": 40, "do": "weigh", "t_min": 636.0, "vessel": "T2", "instrument": "balance", "display": "9.9639 g", "mass_mg": 9963.9, "stable": true, "warm": false}, {"k": 41, "do": "vortex", "t_min": 637.0, "vessel": "T2", "instrument": "vortex", "message": "10 s"}, {"k": 42, "do": "block", "t_min": 637.17, "vessel": "T2", "instrument": "thermoblock", "display": "25.5 C", "message": "60 min at 25.5 C, shaking 500 rpm"}, {"k": 43, "do": "counterweight", "t_min": 697.17, "vessel": "CW", "instrument": "balance", "display": "9.9133 g", "mass_mg": 9913.3, "message": "filled with water to the mass of T2"}, {"k": 44, "do": "spin", "t_min": 700.17, "tubes": ["T2", "CW"], "instrument": "centrifuge", "log": "4400 rpm, 10 min, 2,965 x g at r_max 13.7 cm", "message": "run complete"}, {"k": 45, "do": "decant", "t_min": 711.0, "vessel": "T2", "into": "S2", "message": "supernatant poured off into the weighing bottle"}, {"k": 46, "do": "weigh", "t_min": 712.0, "vessel": "S2", "instrument": "balance", "display": "17.5283 g", "mass_mg": 17528.3, "stable": true, "warm": false}, {"k": 47, "do": "weigh", "t_min": 713.0, "vessel": "T2", "instrument": "balance", "display": "7.2183 g", "mass_mg": 7218.3, "stable": true, "warm": false}, {"k": 48, "do": "evaporate", "t_min": 714.0, "vessel": "S2", "instrument": "hot plate (in the fume hood)", "message": "120 min, open"}, {"k": 49, "do": "evaporate", "t_min": 834.0, "vessel": "T2", "instrument": "fume hood", "message": "120 min, open"}, {"k": 50, "do": "dry", "t_min": 954.0, "vessels": ["T2", "S2"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 51, "do": "desiccate", "t_min": 1014.0, "vessels": ["T2", "S2"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 52, "do": "weigh", "t_min": 1044.0, "vessel": "T2", "instrument": "balance", "display": "7.1273 g", "mass_mg": 7127.3, "stable": true, "warm": false}, {"k": 53, "do": "weigh", "t_min": 1045.0, "vessel": "S2", "instrument": "balance", "display": "14.8973 g", "mass_mg": 14897.3, "stable": true, "warm": false}, {"k": 54, "do": "dry", "t_min": 1046.0, "vessels": ["T2", "S2"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 55, "do": "desiccate", "t_min": 1106.0, "vessels": ["T2", "S2"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 56, "do": "weigh", "t_min": 1136.0, "vessel": "T2", "instrument": "balance", "display": "7.1243 g", "mass_mg": 7124.3, "stable": true, "warm": false}, {"k": 57, "do": "weigh", "t_min": 1137.0, "vessel": "S2", "instrument": "balance", "display": "14.8947 g", "mass_mg": 14894.7, "stable": true, "warm": false}, {"k": 58, "do": "dry", "t_min": 1138.0, "vessels": ["T2", "S2"], "instrument": "oven", "display": "105 C", "message": "60 min at 105 C"}, {"k": 59, "do": "desiccate", "t_min": 1198.0, "vessels": ["T2", "S2"], "instrument": "desiccator", "message": "30 min over silica gel"}, {"k": 60, "do": "weigh", "t_min": 1228.0, "vessel": "T2", "instrument": "balance", "display": "7.1240 g", "mass_mg": 7124.0, "stable": true, "warm": false}, {"k": 61, "do": "weigh", "t_min": 1229.0, "vessel": "S2", "instrument": "balance", "display": "14.8945 g", "mass_mg": 14894.5, "stable": true, "warm": false}], "status": {"state": "accepted", "attempt": 2, "text": "tube 2 (T2) accepted: 181.9 mg"}, "g6": {"gate": {"id": "G6", "name": "The evidence package", "verdict": "PASS", "reason": "The package verifies (VALID): every section matches its digest, the goal and the rule are the frozen ones, and running the notebook again gives the same attempts, status and claims.", "basis": "web/kclrecord.py verify()", "numbers": {"status": "VALID"}}, "verification": {"status": "VALID", "checks": [{"id": "digest:inputs", "ok": true, "what": "the section 'inputs' matches its digest"}, {"id": "digest:goal", "ok": true, "what": "the section 'goal' matches its digest"}, {"id": "digest:rule", "ok": true, "what": "the section 'rule' matches its digest"}, {"id": "digest:method", "ok": true, "what": "the section 'method' matches its digest"}, {"id": "digest:attempts", "ok": true, "what": "the section 'attempts' matches its digest"}, {"id": "digest:status", "ok": true, "what": "the section 'status' matches its digest"}, {"id": "digest:claims", "ok": true, "what": "the section 'claims' matches its digest"}, {"id": "digest:not_claimed", "ok": true, "what": "the section 'not_claimed' matches its digest"}, {"id": "digest:references", "ok": true, "what": "the section 'references' matches its digest"}, {"id": "digest:reproduce", "ok": true, "what": "the section 'reproduce' matches its digest"}, {"id": "digest:engine", "ok": true, "what": "the section 'engine' matches its digest"}, {"id": "manifest", "ok": true, "what": "the manifest matches its digest"}, {"id": "goal", "ok": true, "what": "the goal is the frozen one"}, {"id": "rule", "ok": true, "what": "the rule is the frozen one"}, {"id": "reproduce:attempts", "ok": true, "what": "running the notebook again gives the same attempts"}, {"id": "reproduce:status", "ok": true, "what": "running the notebook again gives the same status"}, {"id": "reproduce:claims", "ok": true, "what": "running the notebook again gives the same claims"}]}, "claims": [{"claim": "Tube 1 missed the goal (161.8 mg) and the cause was found: the bottle labelled 'Ethanol absolute' holds 7.6 +/- 0.3 % water by mass (its density through the published table).", "rests_on": ["G5", "D"], "qualifier": "virtually qualified: a simulated laboratory and its virtual instruments; nothing measured"}, {"claim": "Tube 2 met the goal: 181.9 mg of dry KCl crystals (U = 0.5 mg, k = 2) from a 297.9 mg charge, accepted by the frozen rule.", "rests_on": ["G0", "G1", "G2", "G3", "G5", "D"], "qualifier": "virtually qualified: a simulated laboratory and its virtual instruments; nothing measured"}, {"claim": "The model's prediction for tube 2 as made (177.6 mg) agrees with its measurement (zeta +1.2).", "rests_on": ["G4", "G5", "D"], "qualifier": "virtually qualified: a simulated laboratory and its virtual instruments; nothing measured"}], "not_claimed": ["the crystal size or the onset of crystallisation (the model's size is likely too large; neither is measured here)", "that the hidden laboratory is a real one: it is a simulated laboratory whose differences from the model come from the literature", "the model's table validated at this composition: agreement in a virtual laboratory tests the procedure and the decision, not the table", "any yield at another temperature, charge or antisolvent than the goal's", "any physical measurement"], "manifest_digest": "79c07d03048437c39ffd529e5ab39d144ace3f80d9c4aba9d76415e52b2b395e"}}}, "engine": {"version": "0.13.0", "engine_source_digest": "22a4cfa592f66fc883e56066a9576a368b1de83cc78fff13a184ce520352de19", "world_model_digest": "a8a57a3e1c3e7de2"}, "simulated": "Everything here is simulated: a virtual laboratory whose differences from the model come from the literature, read by virtual instruments with their datasheets' behaviour. The results are virtually qualified; nothing is a measurement of a physical device.", "digest": "ede873f90b3e1c14", "created": "2026-11-24"}