Reference experiment · VCW-002
VCW-002 · sodium bromide precipitated by ethanol PASS
1,000 mg of NaBr, 150 uL of ethanol and 1,000 uL of water (the solvent must hold some ethanol: in pure water below 323 K the stable solid is the dihydrate, outside the model), a minute's vortex, 1,500 uL of ethanol over a minute, ten minutes' wait, 10 minutes at 4,400 rpm (3,000 x g), the supernatant decanted, the pellet weighed. The same mechanics as VCW-001 on another salt of the same published table, at equilibrium fidelity only: no kinetics are published for it here.
Goal: the mass of NaBr that precipitates when the solvent reaches a water mass fraction of 0.435 at 298.15 K, within the table's expanded uncertainty
Revision: re-frozen 2026-10-23: the spin made attainable, as for VCW-001
Frozen 2026-10-23 · protocol digest 1e020fd115efe53f · record 183614199691d65d · the record as JSON
1 · The protocol
- add 1000 mg of NaBr
- add 150 uL of ethanol
- add 1000 uL of water
- mix 60 s (vortex)
- add 1500 uL of ethanol over 60 s
- wait 600 s
- centrifuge 4400 rpm, 10 min
- decant (keep 20 uL)
- read the balance
2 · The runs
| fidelity | solid at the end (mg) | the balance read (mg, wet pellet) | run digest |
|---|---|---|---|
| equilibrium | 172.65 | 232.1 | 6a4370948955 |
3 · The checks
| Check | Verdict | What was found | |
|---|---|---|---|
| W0 | Protocol frozen | PASS | The protocol's digest is 1e020fd115efe53f; the frozen one is 1e020fd115efe53f. sha256 of the checked protocol, truncated; frozen on 2026-10-23 |
| W1 | Matter conserved | PASS | Through every operation, fed = inside + removed to 1.2e-14 mmol and 4.4e-16 g. the run's ledger, element by element and as mass |
| W2 | Solubility inside the table | PASS | At 298.15 K and a water mass fraction of 0.4349 the table gives a saturation mass fraction of 0.26519 ± 0.00697 (95 %), linear between the table's points at 298.15 K, w_water 0.4 and 0.5. Pinho & Macedo 2005 (NIST ThermoML), interpolated; the uncertainty is the table's plus the interpolation's |
| W3 | Equilibrium reached before separation | PASS | At the end of the wait the supersaturation is 1.000000: the liquid is saturated (equilibrium fidelity; this system has no kinetics). S = w / w* from the model's composition and the table |
| W4 | The balance agrees with the model | PASS | The balance read 232.1 mg for the pellet with its liquid; the model's mass in air is 232.08 mg: a difference of +0.02 mg against 0.25 mg (the readability plus three standard deviations of the noise). the virtual balance's reading of the vessel's contents (OIML R 111-1 buoyancy, readability, noise) against the material state |
| W5 | Yield with the table's uncertainty | PASS | The equilibrium fidelity precipitates 172.65 mg; the table's expanded uncertainty on that yield is 29.61 mg (dm/dw* = m_solvent/(1 - w*)^2). This system has no kinetics: the agreement of the two fidelities is not tested here. the solid at the end of the wait; the uncertainty propagated from the table's U95 at the final composition |
| W6 | Numerical convergence | PASS | Equilibrium fidelity only: the end state is an algebraic mass balance on the table, no integrator. The conservation error, 4.4e-13 mg, lies 11 orders of magnitude below the balance's readability (0.1 mg). no kinetics for this system: nothing to converge; the ledger's conservation error against the balance's readability |
| W7 | Refuses what it cannot model | PASS | 5 of 5 probes were refused or answered 'unavailable': T_below (OutOfValidity); T_above (OutOfValidity); pure_water (OutOfValidity); unknown_op (ContractError); raman_without_library (reading marked unavailable). each probe is the reference protocol with one change the model has no basis for; the only acceptable answer is the one expected of it: a refusal with its reason (OutOfValidity, ContractError), or, for a reading without a library, a reading marked unavailable |
Not evaluated yet: W8 Validation against the bench: the pre-registered campaign VCW-001-V, to be run in a laboratory, W9 The evidence package.
4 · How the outputs respond to the assumed parameters
| quantity | reference | range over the plan | largest deviation | driven by | label |
|---|---|---|---|---|---|
| solid before separation (mg) | 172.648 | 165.486 … 172.648 | 4.1 % | T_K = 299.15 | robust |
| concentration left in the liquid (mmol/L) | 2,852.274 | 2,852.274 … 2,872.263 | 0.7 % | T_K = 299.15 | robust |
| dry mass of the pellet (mg) | 188.433 | 180.722 … 196.325 | 4.2 % | pellet_porosity = 0.6 | robust |
| wet pellet with its liquid (mg) | 232.170 | 212.329 … 261.931 | 12.8 % | pellet_porosity = 0.6 | moderate |
| onset of crystallisation | — | — | — | — | not given by this fidelity |
| highest supersaturation (1) | 1.000 | 1.000 … 1.000 | 0.0 % | pellet_porosity = 0.6 | robust |
| mean crystal size (d43) | — | — | — | — | not given by this fidelity |
| fraction pelleted (1) | 1.000 | 1.000 … 1.000 | 0.0 % | — | robust |
one parameter at a time at the values tried, every other parameter at its stated value; the relative deviation of each quantity from the reference run; robust below 5 %, sensitive above 20 %, moderate between.
The plan: each parameter, its stated value, the values tried and why
- pellet_porosity (pellet_porosity): stated 0.5000, tried 0.4, 0.6 · ASSUMED liquid fraction of a loose pellet: 0.4 to 0.6 brackets dense to loose packings
- viscosity_pa_s (viscosity_pa_s): stated 0.0010, tried 0.001, 0.0025 · the liquid's viscosity is ASSUMED 1.0 mPa s; an ethanol-water mixture at 25 C peaks near 2.4 mPa s around 40 % ethanol by mass
- T_K (protocol): stated 298.1500, tried 299.15 · a bath at 25 +/- 1 C; the table's lowest isotherm is 298.15 K and the model refuses below it, so only +1 K is tried
5 · What the model is credible for (context of use)
Question: what the protocol VCW-002 yields and how the crystallisation proceeds, under the stated models.
Intended use: a virtual experiment whose numbers a scientist may use for the quantities marked supported, treat as indicative for those marked uncertain, and must not use for those marked unsupported; none is a measurement.
| quantity | value | governed by | sensitivity | decision | why |
|---|---|---|---|---|---|
| solid before separation (mg) | 172.648 ± 29.61 | solubility (predictive) | robust | supported | governed by solubility (predictive); robust to the assumed parameters (4.1 % at most) evidence: W2, W3, W5, W6 · on the bench: the pellet dried to constant mass, minus the solute its retained liquid held (dry_mass_mg is the direct comparison) |
| concentration left in the liquid (mmol/L) | 2,852.274 ± 75.01 | solubility (predictive) | robust | supported | governed by solubility (predictive); robust to the assumed parameters (0.7 % at most) evidence: W2, W3 · on the bench: the supernatant's concentration (gravimetry of an aliquot, conductivity or titration) |
| dry mass of the pellet (mg) | 188.433 ± 29.61 | solubility (predictive), separation (assumed) | robust | uncertain | governed by an assumed module (solubility (predictive), separation (assumed)) evidence: W2, W5 · on the bench: the pellet dried to constant mass |
| wet pellet with its liquid (mg) | 232.170 | solubility (predictive), separation (assumed) | moderate | uncertain | governed by an assumed module (solubility (predictive), separation (assumed)) evidence: W4 · on the bench: the tube weighed after decanting, minus its tare |
| onset of crystallisation (s) | — | crystallization (not modelled), mixing (assumed) | undefined | unsupported | this fidelity does not give it evidence: W6, metrology: calibration · on the bench: the first turbidity or particle count after the feed starts, at the same threshold |
| highest supersaturation (1) | 1.000 | crystallization (not modelled), mixing (assumed) | robust | unsupported | a module it needs is not modelled (not_modelled) evidence: W6 · on the bench: not directly measurable; inferred from concentration and the table |
| mean crystal size (d43) (um) | — | crystallization (not modelled) | undefined | unsupported | this fidelity does not give it evidence: W6 · on the bench: laser diffraction or image analysis of the product |
| fraction pelleted (1) | 1.000 | separation (assumed), crystallization (not modelled) | robust | unsupported | a module it needs is not modelled (not_modelled) evidence: none · on the bench: the solid left in the decanted supernatant |
Rule: supported = every governing module predictive or calibrated inside its domain and the quantity robust to the assumed parameters; unsupported = a governing module not modelled or the fidelity silent; uncertain otherwise, with the driver named. Framework: ASME V&V 40-2018 (context of use; verification, validation and applicability as credibility factors) and the VVUQ view of digital-twin credibility in Shao, Hightower & Schindel, Manufacturing Letters 35 (2023) 24-28.
6 · What it refuses
- T_below: the protocol at 297.15 K, below the table's lowest isotherm → refused (OutOfValidity) 297.15 K is outside the table's isotherms 298.15-348.15 K (NaBr)
- T_above: the protocol at 360.15 K, above the table's highest isotherm → refused (OutOfValidity) 360.15 K is outside the table's isotherms 298.15-348.15 K (NaBr)
- pure_water: the protocol with water as the only solvent (water mass fraction 1, outside this isotherm's range [0.0, 0.9]) → refused (OutOfValidity) w_water 1 is outside the table's range [0, 0.9] at 298.15 K (NaBr, crystal NaBr)
- unknown_op: a 'heat' operation: no module models heating → refused (ContractError) op 1: must be an object with 'op' one of ('add_solid', 'add_liquid', 'mix', 'add_antisolvent', 'wait', 'centrifuge', 'decant', 'read')
- raman_without_library: a Raman reading of a system without a measured reference library → unavailable (reading marked unavailable) no Raman reference library for the system nabr-water-ethanol: a reading cannot be produced
each probe is the reference protocol with one change the model has no basis for; the only acceptable answer is the one expected of it: a refusal with its reason (OutOfValidity, ContractError), or, for a reading without a library, a reading marked unavailable.
7 · What this establishes
rests on W0, W1, W2, W3 · virtually qualified: a simulation consistent with its stated models and the published table; nothing measured
Not claimed:
- the time course: this system has no kinetics, only the equilibrium end state
- the pellet's liquid content (an assumed porosity)
- any physical measurement
8 · Calibration, accuracy, precision
- Calibration: none: the equilibrium fidelity uses the published table alone;
- Accuracy: the yield is 172.65 ± 29.61 mg (17.2 %, the table's expanded uncertainty propagated); not cross-checked: the compilation's sodium bromide row lists nine values for eleven temperatures and does not say which solid (the dihydrate below 51 C, the anhydrous salt above) each refers to: no value can be matched to the primary table without guessing
- Precision: over 30 seeds the balance's reading has a standard deviation of 0.057 mg (expected 0.058 mg from its noise and readability): consistent.
calibration, accuracy and precision in the metrological sense, of a simulation: they say how well the model agrees with the published numbers it rests on and how repeatable its virtual instrument is, not how a bench run would come out
9 · What the model can say (validity)
Overall: assumed; the weakest module decides: a result is no better than its most assumed parameter.
| Module | Step | Status | Basis |
|---|---|---|---|
| solubility | dissolution and antisolvent addition: how much can stay dissolved | predictive | every value is interpolated between the published table's own points; outside them the model refuses (OutOfValidity) every node reproduced exactly; the equilibrium yield of VCW-001 equals the table's mass balance (tests/test_world_solubility.py, test_world_conservation.py) |
| density | the liquid's volume (concentrations, levels, the pellet's liquid) | predictive | the solvent's density interpolated in a published table (10-30 C; outside it the state falls back to additive volumes and says so); the salt at its apparent molar volume in water, ASSUMED to hold in the mixture (at most 1.5 % of the liquid's volume) the table's pure ends against Tanaka 2001 and the CRC Handbook, its interior against a second compilation; the mixture contracts by about 3 % near equal masses (tests/test_world_density.py) |
| crystallization | nucleation, growth, dissolution | not modelled | this system has no kinetics file: only the equilibrium fidelity |
| mixing | mixing and the antisolvent feed | assumed | one well-mixed compartment with a time-varying volume; the blend time only says when a mix op is homogeneous |
| separation | centrifugation and decanting | assumed | the cut size follows Stokes settling in the stated field; the viscosity, the pellet's porosity and the tube's geometry are assumed RCF and the cut size against the formulas by hand (tests/test_world_separation.py) |
| temperature | heating and cooling | not modelled | the temperature is one constant per protocol; no heat transfer, no evaporation |
| aging | aging (ripening, polymorph transformation) | not modelled | Ostwald ripening and polymorph transformation are absent: a wait only lets nucleation and growth finish |
predictive: an established method with literature or measured parameters, inside their domain · calibrated: an established method with a parameter fitted to a named observation; valid in the fitted domain only · assumed: an established method with a stated, unmeasured parameter: what the model does, not what the bench will do · not modelled: absent from the model; the ledger says so