Scientific basis · VCW-001
VCW-001: KCl crystallised by ethanol, by hand: the scientific basis
Scientific foundations, present capabilities, partial replacement of physical experiments, and the validation roadmap.
The structure and the positioning follow a colleague's technical assessment (24 October 2026); every number on this page is the credibility engine's or the reference record's, so it stays true to the code. · print this page for a PDF.
Status: a scientific computational prototype with a functioning KCl reference case; not an experimentally qualified, general-purpose predictive digital twin.
Suggested public description: VialBot is a physics-informed virtual laboratory for screening, testing and refining supported experiments before physical execution.
What 'replacement' means here: VialBot can replace some preliminary calculations, in-silico screening runs, workflow rehearsal and equipment-practice trials, within its models. It cannot replace the new physical measurements needed to establish a crystal phase, calibrate an instrument or validate the model's accuracy.
1. Why this is a world model rather than a 3D animation
A scientific world model represents the state of the experiment and evolves it when an action is performed; its predictions generate what the displays and the instruments show:
s(t + dt) = F_theta( s(t), a(t), dt ) y(t) = H_phi( s(t) ) + noise(t)Here the state s(t) holds chemical amounts, solvent masses, the solid phases, the temperature and, at the kinetic fidelity, the moments of the crystal-size distribution; an action a(t) is to add a solid, add a liquid, mix, add the antisolvent, wait in the block, centrifuge, decant, dry; the virtual instruments read the state (balance, densitometer, thermometer, centrifuge, oven). F is the physical and chemical process model, H the simulated instruments. Hence the room is not programmed to 'show crystals at 1.5 mL of ethanol': the solubility table, re-evaluated after the ethanol changes the solvent's composition, decides how much salt leaves the liquid, and the room shows that state. It is neither a universal molecular simulation nor a calibrated digital twin of specific equipment: its validity is stated part by part, quantity by quantity (below).
2. What it can replace, reduce, and cannot presently replace
| Research activity | Current practical benefit | Appropriate claim |
|---|---|---|
| Calculate stoichiometry, balances and stock use | replaces manual calculations and catches impossible recipes | available for the modelled species |
| Screen KCl / water / ethanol compositions | equilibrium outcomes before any chemical is used | computational screening, inside the table's domain |
| Rehearse the crystallisation by hand | instruments with their limits and refusals, a wrong label, the corrective step | virtual process training; hardware qualification separate |
| Explore the crystallisation's time, onset or crystal size | the kinetic model responds to time and supersaturation (reference runs only, not the room) | exploratory: estimated parameters |
| Obtain balance, density or centrifuge readings | what the modelled instrument would report, with its errors | synthetic readings, not measurements |
| Validate a real instrument or a batch of product | needs calibration and measurements on real hardware | physical work still required |
| Predict an arbitrary new synthesis | no general thermodynamic or kinetic mechanism | unsupported |
Scope of replacement: the defensible contribution is the partial substitution of exploratory physical iterations by computationally justified screening, training and protocol checking, not the elimination of real experiments. Quantified reductions of cost, waste or runs need a comparative study; none is established yet.
3. What happens when a researcher changes the experiment?
The model starts with 300 mg of KCl in 1,000 uL of water and varies the ethanol. Its equilibrium predictions, frozen before any bench run (VCW-001-V):
| ethanol (uL) | water at the end | crystallised KCl (mg) ± U95 | still dissolved (mg) | left in the liquid (mM) ± U95 |
|---|---|---|---|---|
| 1,000 | 55.9 % | 146.35 ± 5.6 | 153.65 | 1,035 ± 35 |
| 1,500 | 45.8 % | 181.17 ± 6.1 | 118.83 | 648 ± 31 |
| 2,000 | 38.8 % | 205.97 ± 3.7 | 94.03 | 429 ± 16 |
These are model predictions, not measured yields: they follow from the published solubility data and the conservation of matter. What is recovered after the spin and the handling may differ from what crystallised; the room's measurement corrects for the mother liquor the pellet keeps. Varying the feed's duration engages the kinetic model, whose course is exploratory (estimated coefficients); the room computes every tube at equilibrium. The room's reference loop is a synthetic fault experiment: a nominal laboratory gives about 184.9 mg; a hidden bottle of wetter ethanol gives about 161.8 mg, the decision finds the cause, and the corrected tube 2 gives about 181.9 mg. The hidden laboratory is another simulator, not measured equipment.
4. The scientific principles and equations, part by part
Each part distinguishes the physical principle, its actual computation and its evidence. A genuine equation does not guarantee that its parameters fit the experiment studied: the evidence level says how far it does.
The goal and the decision rule, frozen established
Principle: pre-registration: the goal and the rule are fixed, and their digests recorded, before any tube is made, so a result cannot redefine its own target.
Where: Goal · why: a tube is judged against what was asked for, not against its own recipe.
Implementation: web/kclcase.py
Confidence: established a stated goal and a decision rule after JCGM 106 and ILAC G8, frozen by sha256 digests (G0).
Limitations:
- the tolerance (6 %) is the smallest the published table's uncertainty lets the model promise at 95 %; 5 % is not attainable
Amounts and conservation of matter exact law
Principle: conservation of mass and of every element: what was fed is inside the vessel or recorded as removed.
n_k,fed = n_k,in vessel + n_k,removed, for every element kWhere: Plan · why: every other number is built on amounts that add up.
Implementation: vialbot/chemistry/substances.py, vialbot/chemistry/state.py
Confidence: exact law exact bookkeeping; molar masses from IUPAC CIAAW atomic weights (2021).
Limitations:
- a species missing from the model cannot be conserved: the books are only as complete as the substances listed
Solubility of KCl in water + ethanol checked
Principle: a published table of measured solubilities, interpolated piecewise-linearly in composition and linearly in temperature between isotherms; never extrapolated (a query outside the table is refused).
w* = w_sat(T, w_water); S = w / w*Where: Plan · why: the model's recipe (the ethanol volume) and its predicted yield rest on it.
Implementation: vialbot/chemistry/solubility.py, vialbot/examples/world/solubility_kcl_water_ethanol.json, vialbot/examples/validation/kcl_water_ethanol_independent.json
Confidence: checked 44 of 46 points from three independent laboratories agree within their uncertainty; where the case lives the sharpest independent point differs by +1.2 % (/world/overview, level 2).
Limitations:
- the table's stated uncertainty is one laboratory's: in water-rich solvent independent laboratories differ by up to 2.3 % where it claims 0.3 %
- between 20 and 55 % water no precise independent point exists; the case lives at 38-57 % water
- only three isotherms (25, 50, 75 C): between them the value is linear in temperature
Density of the liquid checked
Principle: water + ethanol from a published density table by mass fraction (volumes are not additive: the mixture contracts); the dissolved salt at its apparent molar volume (Masson's form).
V = m_solvent / rho(T, w_ethanol) + n_KCl V_phi, V_phi = V0_phi + b sqrt(molality)Where: Plan · why: the concentration left in the liquid and the volumes of the decant.
Implementation: vialbot/chemistry/density.py, vialbot/examples/world/density_water_ethanol.json
Confidence: checked the saturated solutions' densities of an independent laboratory: 8 of 10 agree at 25 C, all within 1.2 %.
Limitations:
- outside 10-30 C the volumes fall back to additive pure densities (assumed): the run says which one it used
- two independent points (63 and 71 % water, much salt) differ by 0.6 and 1.2 %
Equilibrium crystallisation established
Principle: a mass balance on the solubility table: the excess salt over saturation leaves the liquid as crystals.
m_crystals = max(0, m_KCl - w*/(1 - w*) m_solvent)Where: Plan, Crystallise · why: the room computes every tube this way (equilibrium fidelity); it is the predicted yield.
Implementation: vialbot/chemistry/simulate.py, web/kcllab.py, web/kclcase.py
Confidence: established a mass balance (exact) on a checked table; that equilibrium is reached within the hour in the block is an assumption, robust in the kinetic model (see the quantity).
Limitations:
- the 3D room runs the equilibrium fidelity only: the kinetic model is not live in the room
- that an hour in the block reaches equilibrium is assumed; the bench campaign tests it
The prediction and its uncertainty (G4) established
Principle: the GUM's law of propagation, validated by the adaptive Monte Carlo method; every input with its source.
Where: Plan · why: the recipe is chosen so that the goal is met with P >= 0.95, and the measurement is compared with it.
Implementation: web/kclcase.py
Confidence: established JCGM 100 and 101 methods; their inputs carry their own evidence (the budget below).
The checks that find a cause established
Principle: an independent measurement replaces an assumed belief (a standardisation): the ethanol's water from its density through the published table; the block's offset from the thermometer.
Where: Prepare, Decide · why: the cause of a rejection is found, not guessed, and the next tube uses it.
Implementation: web/kclcase.py
Confidence: established standardisation (Eurachem/CITAC QUAM Example A2); the density table (checked).
Mixing and the antisolvent feed estimate
Principle: one well-mixed compartment with a time-varying volume; a mix op is homogeneous after the method's blend time.
Where: Make · why: the solubility applies to the whole liquid.
Implementation: vialbot/chemistry/mixing.py
Confidence: estimate one well-mixed compartment with a time-varying volume; the blend time only says when a mix op is homogeneous: a vortexed 2.5 mL tube blends in seconds, so one well-mixed compartment is an estimate, not a measurement.
Limitations:
- feed-zone (micromixing) supersaturation: the engulfment model of Bałdyga & Bourne is v2 (ledger R-W04)
Crystallisation kinetics placeholder
Principle: a population balance by moments; classical nucleation theory in two regimes; first-order growth; Noyes-Whitney dissolution; RK4 with step doubling.
J = A_het exp(-B_het/ln^2 S) + A_hom exp(-B_hom/ln^2 S); G = k_g (S-1)^g (S >= 1), -k_d (1-S) (S < 1); dM_j/dt = j G M_(j-1) + J L0^j VWhere: Crystallise · why: the onset, the supersaturation's course and the crystal size in the reference runs and the campaign's recorded (not judged) quantities.
Implementation: vialbot/chemistry/crystallization.py, vialbot/chemistry/ode.py, vialbot/examples/world/kinetics_kcl.json
Confidence: placeholder the nucleation exponents are published; the pre-factors are calibrated to published induction times; the growth and dissolution coefficients, the nucleus size, the powder's size distribution and the shape factor are assumed and say so; the 4 induction times not fitted are reproduced within 22 %; they come from the same study and method as the fitted two, in water: an internal check, not an independent laboratory, and the transfer to the ethanol mixture is assumed.
Limitations:
- not run in the 3D room
- the growth and dissolution coefficients are estimates from the literature and the powder's size distribution a placeholder: the onset and the crystal size are labelled uncertain
- the nucleation pre-factors were fitted to induction times in water; the ethanol mixture is assumed to behave alike
Centrifugation and decanting established
Principle: RCF = 1.118e-5 r rpm^2; Stokes cut size d_c = sqrt(18 mu h / ((rho_p - rho_l) a t)); the mass fraction above d_c pellets, the rest leaves with the supernatant; the pellet keeps liquid at an assumed porosity.
RCF = 1.118e-5 r_cm rpm^2; d_c = sqrt(18 mu h / ((rho_p - rho_l) RCF g t))Where: Separate · why: the pellet's retained liquid carries dissolved salt the measurement corrects for.
Implementation: vialbot/chemistry/separation.py
Confidence: established the relative centrifugal force (exact) and Stokes settling (established); the viscosity, the pellet's porosity and the film the decant leaves are assumed.
Limitations:
- the liquid's viscosity as a function of composition
- hindered settling at high solid fractions
- the conical bottom of a tube
- resuspension and pellet compaction
The gravimetric measurement (G5) established
Principle: drying to constant mass, weighing by difference, and the supernatant's measured mass fraction for the correction.
Y = m_dry - (m_wet - m_dry) w/(1 - w), w = residue/supernatantWhere: Decide · why: the tube's result, as the laboratory's own measurement (never the model's).
Implementation: web/kclcase.py
Confidence: established Ph. Eur. constant mass; USP <41> minimum weight; GUM budget.
The decision (D) established
Principle: guarded acceptance (the conformance probability at the limits) and zeta scores of the yield and the supernatant against the prediction for the tube as made.
Where: Decide · why: a frozen rule decides, not a person after the fact.
Implementation: web/kclcase.py, web/decision.py
Confidence: established JCGM 106 conformity assessment; Eurachem/CITAC guard bands; ISO 13528 zeta.
The evidence package (G6) exact law
Principle: content hashes and a deterministic engine: the same notebook gives the same digests.
Where: Decide · why: anyone can check that the record is what the engine produces.
Implementation: web/kclrecord.py
Confidence: exact law sha256 digests and reproduction.
The virtual laboratory (the hidden truth) test scenario
Principle: its solubility lies within the spread of the independent datasets; the charge's moisture and purity, the ethanol's water, the pellet, the decant and the drying are drawn within stated ranges; in one scenario the bottle labelled absolute is the 96 % grade.
Where: — · why: to test the loop (plan, make, measure, decide, find the cause) against a laboratory that is not the model.
Implementation: web/kcllab.py, vialbot/chemistry/truth.py
Confidence: test scenario simulated by design: the test bench, not the model; its ranges are stated (kcllab.ASSUMED), its values are never shown.
Limitations:
- it tests the procedure, not the chemistry: a pass here is not a measurement
- its differences from the model are the ones written in; a real laboratory has others
5. The instruments
Virtual instruments respond to the simulated state plus the modelled instrument's errors; they never display the target. Their readings are synthetic.
| Instrument | Evidence | Specification from | Estimated parameters (and placeholders) |
|---|---|---|---|
| Reference thermometer (IEC 60751 class AA) checks the block before the first tube | established | IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C | — |
| Thermoblock (ThermoMixer C class) holds the tube at the crystallisation temperature | established | Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes | fluctuation_sd_K (placeholder) |
| Analytical balance (Sartorius Quintix class) net masses by difference: the charge, every addition, the pellets and residues | established | Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class) | warm_bias_mg_per_K (placeholder) |
| Air-displacement pipettes P1000, P5000 the water and the ethanol | established | ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL | volatile (placeholder); step_uL |
| Syringe pump (slow ethanol feed) a slow ethanol feed | established | Harvard Apparatus Pump 11 Elite (distributor's specification): accuracy +/-0.5 %, reproducibility +/-0.05 %; New Era NE-1000 syringe table (B-D 5 mL, 11.99 mm); start-up lag from slack: Neff et al., Eur. J. Anaesthesiol. 2007 | slack_uL (placeholder) |
| Vortex mixer and fume hood mixing; evaporating the ethanol before the oven | placeholder | — | |
| Benchtop centrifuge (5702 R class, A-4-38 rotor) packs the crystals into a pellet | established | Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm | imbalance_limit_g (placeholder); temp_control_K (placeholder) |
| Drying oven dries the pellet and the residue to constant mass | established | drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096) | uniformity_K (placeholder) |
| Desiccator cools a vessel before it is weighed | established | practice: cool in a desiccator before weighing (USP <731>) | tau_min (placeholder) |
| Densitometer (DMA 35 class) reads the ethanol bottle's density | established | Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C | — |
What the instruments showed in the reference run
the reference run, scenario 'wrong-ethanol', seed 4242: each reading set against the published data or the specification that gives it meaning; the numbers are under each figure.
- the thermometer's reading ± its class AA tolerance (±0.14 °C)
- the block's datasheet: ±0.5 °C around the set point
The instrument showed 25.35 C; the block holds the tube 0.15 °C below its set point, inside its datasheet's ±0.5 °C; taken before the first tube, this measured offset replaces the datasheet's band in the plan.
The numbers
| °C | |
|---|---|
| set point | 25.50 |
| reading | 25.35 |
| offset | -0.15 |
| thermometer tolerance (class AA) | ±0.14 |
| block datasheet | ±0.50 |
IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C; Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes · simulated: a virtual instrument's reading of a hidden virtual laboratory
- published solubility at 25.35 °C (line); measured points at 25.0 °C (dots)
- T1's path as the ethanol goes in
- the tubes, as weighed (the ethanol's water as the plan believed it)
- the same tube with the water the densitometer found in the ethanol
T1 holds 11.6 % KCl overall where its liquid holds 4.8 % at saturation: the excess crystallises (the model: 182 mg); with the water the densitometer found in the ethanol the tube sat further right, where the liquid holds more KCl: fewer crystals.
The numbers
| tube | water in the solvent (%) | KCl overall (%) | saturation there (%) | crystals, the model (mg) |
|---|---|---|---|---|
| T1 (weighed; ethanol's water: the label) | 44.1 | 11.62 | 4.79 | 182.4 |
| T2 (weighed; ethanol's water: the densitometer) | 41.3 | 9.75 | 4.09 | 177.6 |
| T1 (weighed; ethanol's water: the densitometer, read afterwards) | 48.3 | 11.62 | 5.82 | 156.7 |
Pinho & Macedo, J. Chem. Eng. Data 50 (2005) 29 (NIST ThermoML): the table the model interpolates; the tubes' compositions from the notebook's weighings and the case's model (web/kclcase.py) · published data; the compositions are the model's, from simulated weighings
- liquid at 2.4 mPa·s (the mixture's viscosity peak; not modelled)
- liquid at 1.0 mPa·s (the model's assumed viscosity)
The instrument showed 4400 rpm, 10 min, 2,965 x g at r_max 13.7 cm; after 10 min at 2,965 × g every KCl crystal larger than 0.12-0.19 µm has reached the bottom of the 1.5 cm column; the room does not track crystal sizes (equilibrium fidelity): every crystal is taken as larger.
The numbers
| speed, time | 4,400 rpm, 10 min |
| relative centrifugal force at r_max 13.7 cm | 2,965 × g |
| T1 against the counterweight (imbalance limit 2 g, assumed) | 9.4251 g / 9.4730 g: 0.048 g apart |
| liquid column in the tube (taken as a cylinder, assumed) | 1.53 cm |
| liquid density (the published table) | 0.8962 g/mL |
| smallest KCl crystal at the bottom after 10 min | 0.12 µm (1.0 mPa·s) to 0.19 µm (2.4 mPa·s) |
Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm; Stokes settling: Svarovsky, Solid-Liquid Separation, 4th ed., 2000, ch. 7 (sedimenting centrifuges, cut size, sigma theory) · the instrument's program; the settling is the world model's (Stokes), not a reading
- liquid at 2.4 mPa·s (the mixture's viscosity peak; not modelled)
- liquid at 1.0 mPa·s (the model's assumed viscosity)
The instrument showed 4400 rpm, 10 min, 2,965 x g at r_max 13.7 cm; after 10 min at 2,965 × g every KCl crystal larger than 0.13-0.20 µm has reached the bottom of the 1.9 cm column; the room does not track crystal sizes (equilibrium fidelity): every crystal is taken as larger.
The numbers
| speed, time | 4,400 rpm, 10 min |
| relative centrifugal force at r_max 13.7 cm | 2,965 × g |
| T2 against the counterweight (imbalance limit 2 g, assumed) | 9.9639 g / 9.9133 g: 0.051 g apart |
| liquid column in the tube (taken as a cylinder, assumed) | 1.86 cm |
| liquid density (the published table) | 0.8897 g/mL |
| smallest KCl crystal at the bottom after 10 min | 0.13 µm (1.0 mPa·s) to 0.20 µm (2.4 mPa·s) |
Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm; Stokes settling: Svarovsky, Solid-Liquid Separation, 4th ed., 2000, ch. 7 (sedimenting centrifuges, cut size, sigma theory) · the instrument's program; the settling is the world model's (Stokes), not a reading
- the pellet in T1
- the residue in S1
- constant mass: two weighings within 0.5 mg (Ph. Eur.)
The instrument showed T1 7.0353 g, S1 14.6587 g; the pellet in T1: the last two weighings 0.4 mg apart; the residue in S1: the last two weighings 0.3 mg apart.
The numbers
| weighing | T1 (gross) | T1 net (mg) | S1 (gross) | S1 net (mg) |
|---|---|---|---|---|
| tare | 6.8678 g | 14.5263 g | ||
| after the decant (wet pellet; supernatant) | 7.1234 g | 255.6 | 16.8281 g | 2301.8 |
| after oven cycle 1 (60 min at 105 °C) | 7.0384 g | 170.6 | 14.6613 g | 135.0 |
| after oven cycle 2 (120 min at 105 °C) | 7.0357 g | 167.9 | 14.6590 g | 132.7 |
| after oven cycle 3 (180 min at 105 °C) | 7.0353 g | 167.5 | 14.6587 g | 132.4 |
Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class); drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096) · simulated: a virtual instrument's reading of a hidden virtual laboratory
- the pellet in T2
- the residue in S2
- constant mass: two weighings within 0.5 mg (Ph. Eur.)
The instrument showed T2 7.1240 g, S2 14.8945 g; the pellet in T2: the last two weighings 0.3 mg apart; the residue in S2: the last two weighings 0.2 mg apart.
The numbers
| weighing | T2 (gross) | T2 net (mg) | S2 (gross) | S2 net (mg) |
|---|---|---|---|---|
| tare | 6.9377 g | 14.7826 g | ||
| after the decant (wet pellet; supernatant) | 7.2183 g | 280.6 | 17.5283 g | 2745.7 |
| after oven cycle 1 (60 min at 105 °C) | 7.1273 g | 189.6 | 14.8973 g | 114.7 |
| after oven cycle 2 (120 min at 105 °C) | 7.1243 g | 186.6 | 14.8947 g | 112.1 |
| after oven cycle 3 (180 min at 105 °C) | 7.1240 g | 186.3 | 14.8945 g | 111.9 |
Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class); drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096) · simulated: a virtual instrument's reading of a hidden virtual laboratory
- published water + ethanol densities at 22.1 °C (dots: the table's 1 % rows)
- the reading, through the table: the bottle's water (± 2u)
- the label: ethanol absolute, ≤ 0.2 % water
The instrument showed 0.8095 g/cm3 at 22.1 C; the table puts the bottle's water at 7.6 ± 0.6 % by mass (k = 2): not the label's ≤ 0.2 %.
The numbers
| value | |
|---|---|
| reading | 0.8095 g/cm³ |
| at | 22.1 °C |
| u of the reading (accuracy, repeatability, temperature) | 0.0008 g/cm³ |
| water implied (the table inverted) | 7.55 % |
| u of the water | 0.28 % |
| the label's claim | ≤ 0.2 % |
Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C; the table: N. S. Osborne, E. C. McKelvy, H. W. Bearce, Density and thermal expansion of ethyl alcohol and of its mixtures with water, Bull. Bur. Stand. 9 (1913) 327-474, as compiled in N. A. Lange, Lange's Handbook of Chemistry, 10th ed., ed. J. A. Dean, McGraw-Hill, 1967 · published data; a simulated reading
6. Evidence audit: what is actually supported?
| Evidence | Present? | What it supports |
|---|---|---|
| The sources and the equations can be inspected | yes | every part names its method, equation and code (above) |
| Automated tests | yes | 0 test functions in 0 files (test_world_*.py, test_kcl_*.py, test_labkit.py, test_credibility.py); the whole suite runs before every commit |
| Numerical convergence and conservation | yes | W1 matter conserved to rounding; W6 a hundredfold tighter integration moves nothing that matters |
| Comparison with independent published data | yes | KCl solubility: 44 of 46 independent points agree within their uncertainty (1 disagrees); the saturated solutions' density: 8 of 10 at 25 C, all within 1.2 % (/world/overview) |
| Synthetic hidden-laboratory trials, a fault and its correction | yes | both scenarios' loops; their evidence packages verify (G6) |
| Predictions frozen before experiments (unseen conditions) | yes | VCW-001-V: 18 runs pre-registered with the model's digest |
| A characterised real KCl experiment run in this workflow | no | physical end-to-end validation not established |
| Device-specific calibration of the real instruments | no | physical-equipment fidelity not established |
And the credibility engine's verdict on each reported number (its weakest governing part; the share of its uncertainty that rests on estimates or placeholders):
| Number | Weakest part | On estimates | Verdict |
|---|---|---|---|
| the predicted yield (G4) | established | 4 % | supported |
| the measured yield (G5) | established | 21 % | supported |
| the decision (D) | established | 21 % | supported |
| the cause found (the ethanol's water) | established | — | supported |
| the crystal size and the onset (reference runs only) | placeholder | — | uncertain |
Rule: a quantity is supported when its weakest governing part is at least calibrated and estimates or placeholders carry less than half of its uncertainty; uncertain when a governing part is only an estimate or a placeholder, or when they carry half or more of its uncertainty; unsupported when a part it needs is not modelled.
7. Strengths and hard boundaries, stated precisely
Strengths demonstrated in software
- A shared material state, not a scripted animation: the chemistry is computed on the server and the room shows it.
- Physical transparency: solubility, density, kinetics and separation are identifiable equations with sources and parameter provenance.
- Scope limits: a temperature or composition outside the table is refused, not extrapolated.
- Two fidelities: equilibrium for the supported screening, kinetics for the exploratory course of the crystallisation.
- Instrument-aware training: wrong reagents, instrument limits, noise and corrective experiments explored safely.
- Reproducible evidence: frozen goals and predictions, model digests, records and packages that verify by re-running.
Current boundaries
- the tolerance (6 %) is the smallest the published table's uncertainty lets the model promise at 95 %; 5 % is not attainable
- a species missing from the model cannot be conserved: the books are only as complete as the substances listed
- the table's stated uncertainty is one laboratory's: in water-rich solvent independent laboratories differ by up to 2.3 % where it claims 0.3 %
- between 20 and 55 % water no precise independent point exists; the case lives at 38-57 % water
- only three isotherms (25, 50, 75 C): between them the value is linear in temperature
- outside 10-30 C the volumes fall back to additive pure densities (assumed): the run says which one it used
- two independent points (63 and 71 % water, much salt) differ by 0.6 and 1.2 %
- the 3D room runs the equilibrium fidelity only: the kinetic model is not live in the room
- that an hour in the block reaches equilibrium is assumed; the bench campaign tests it
- a virtual instrument: its readings are synthetic
- evaporation during handling is not modelled
- feed-zone (micromixing) supersaturation: the engulfment model of Bałdyga & Bourne is v2 (ledger R-W04)
- not run in the 3D room
- the growth and dissolution coefficients are estimates from the literature and the powder's size distribution a placeholder: the onset and the crystal size are labelled uncertain
- the nucleation pre-factors were fitted to induction times in water; the ethanol mixture is assumed to behave alike
- the liquid's viscosity as a function of composition
- hindered settling at high solid fractions
- the conical bottom of a tube
- resuspension and pellet compaction
- it tests the procedure, not the chemistry: a pass here is not a measurement
- its differences from the model are the ones written in; a real laboratory has others
- Visual appearance, instrument displays and a PASS in a simulated laboratory are never measured evidence.
8. How to turn 'partially replacing experiments' into a measurable claim
The publishable proposition is not a digital replica of every chemical. A testable question: for a defined class of liquid-handling and crystallisation experiments, how many exploratory physical trials, operator errors and unproductive runs can a verified chemistry world model avoid, without compromising the quality of the eventual physical outcome?
The comparison uses matched objectives and budgets: (A) conventional recipe preparation and trial-and-error; (B) virtual screening plus physical confirmation. Pre-register the scope, the materials and the endpoints; report the physical experiment count, the material consumed, the operator and instrument time, the success fraction, the endpoint error and the corrective actions. Simulated savings are never cited as laboratory savings.
Minimum next demonstration: Hold out several KCl compositions, keep the model and its uncertainty frozen, run the supervised bench experiments, measure the dry yield and the dissolved KCl, compare with the predictions, then connect the accepted protocol to qualified hardware with the same record. The 18-run pilot pre-registered in docs/validation/vcw-001-v-campaign.md is a practical start, not a universal proof.
Afterwards: Cs3Sb2I9 (LARP-001) is a second, much harder challenge, for when independent phase-identified Raman or XRD and untrained synthesis conditions exist: success on KCl does not validate phase selection in metal-halide mixtures.
9. Recommended positioning
Headline: VialBot: explore chemistry before entering the laboratory.
One sentence: VialBot is a physics-informed virtual laboratory for screening, testing and refining supported experiments before physical execution.
Two sentences: VialBot combines chemical-state models, simulated laboratory instruments and independently checked workflows to move part of early experimental screening into a virtual environment. It helps researchers reduce avoidable physical trial-and-error while clearly reporting which outcomes are established, estimated or still require bench validation.
Manuscript framing: A provenance-aware chemistry world model for the partial substitution of laboratory trial-and-error through in-silico experiment design, simulation-based fault diagnosis, instrument-aware workflows and quantitatively benchmarked sim-to-real transfer.
Avoid: 'Replaces chemistry experiments'; 'Predicts the exact phase of any material'; 'A fully realistic equipment digital twin'; 'Verified 95 % accurate experimental outcomes' or a quantified waste reduction before measurements exist.
10. References
- JCGM 106:2012, The role of measurement uncertainty in conformity assessment: 7.4 expression (11), the conformance probability for a normal PDF; 8.3.2 guarded acceptance, 8.3.3 guarded rejection
- Eurachem/CITAC Guide, Use of uncertainty information in compliance assessment, 2nd ed. (2021): 4.3 a guard band of 1.64u gives a 5 % probability of a wrong decision at the limit; 4.4 conditional or inconclusive results; 5 and 7, the uncertainty at the limit; Annex B, Example 1, g = 1.64u on a two-sided specification
- IUPAC CIAAW, abridged standard atomic weights (2021)
- Pinho, S. P., Macedo, E. A.[Eugenia A.], J. Chem. Eng. Data 2005, 50, 29-32, DOI 10.1021/je049922y; NIST TRC ThermoML record https://trc.nist.gov/ThermoML/10.1021/je049922y.json (sha256 f82f5add71ea)
- Chiavone-Filho & Rasmussen 1993 (doi:10.1021/je00011a009); Farelo et al. 2004 (doi:10.1021/je049782s); Galleguillos et al. 2003 (doi:10.1021/je020173z): the independent checks
- N. A. Lange, Lange's Handbook of Chemistry, 10th ed., ed. J. A. Dean, McGraw-Hill, 1967
- N. S. Osborne, E. C. McKelvy, H. W. Bearce, Density and thermal expansion of ethyl alcohol and of its mixtures with water, Bull. Bur. Stand. 9 (1913) 327-474
- Millero, Chem. Rev. 71 (1971) 147: the apparent molar volume of KCl
- Mullin, Crystallization, 4th ed. (2001): solubility and the crystal yield
- JCGM 100:2008 (GUM): law of propagation of uncertainty, 5.1.2; coverage factor, 6 and G
- JCGM 101:2008: the adaptive Monte Carlo procedure (7.9) and the validation of the GUM framework (8.1, 8.2)
- IEC 60751 class AA platinum resistance thermometer: tolerance +/-(0.1 + 0.0017 |t|) C
- Eppendorf ThermoMixer C: temperature accuracy +/-0.5 C at 20-45 C; SmartBlock 15 mL: 8 conical tubes
- Sartorius Quintix 224-1x datasheet (repeatability typ. 0.08 mg, linearity <= 0.2 mg, typ. 0.06 mg); USP <41> Balances (2014): minimum weight 2000 s, s >= 0.41 d (82 mg for d = 0.1 mg; 160 mg typical for this class)
- Eurachem/CITAC Guide QUAM:2012, Example A2
- Lange's Handbook of Chemistry, 10th ed. (1967): density of aqueous ethanol (vialbot/examples/world/density_water_ethanol.json)
- ISO 8655-2:2022 maximum permissible errors, single-channel air-displacement pipettes (Eppendorf customer information, Table 1): 1,000 uL nominal +/-8 uL systematic, 3 uL random; 5,000 uL nominal +/-40 uL, 15 uL
- Harvard Apparatus Pump 11 Elite (distributor's specification): accuracy +/-0.5 %, reproducibility +/-0.05 %; New Era NE-1000 syringe table (B-D 5 mL, 11.99 mm); start-up lag from slack: Neff et al., Eur. J. Anaesthesiol. 2007
- Grenville & Nienow (Handbook of Industrial Mixing 2004) for a stirred vessel; nothing published for a hand-mixed vial
- Randolph & Larson 1988; Mullin 2001 ch. 5-6; Wang et al. 2023 (induction times); Lopes & Farelo 2006 (growth order)
- Eppendorf Centrifuge 5702 R operating manual: rotor A-4-38, 15 mL conical tube r_max 13.7 cm, 4,400 rpm max (3,000 x g), start-up < 25 s, imbalance stop 'Inb', speed limit, refrigerated set point with a +/-3 C alarm
- Svarovsky, Solid-Liquid Separation, 4th ed., 2000, ch. 7 (sedimenting centrifuges, cut size, sigma theory); Beckman Coulter rotor tables (the RCF formula)
- drying at 105 C: USP Potassium Chloride monograph (loss on drying 105 C, 2 h); constant mass: two consecutive weighings within 0.5 mg, the second after further drying (Ph. Eur. convention, from secondary sources); PP tubes stable to 121 C (Falcon 352096)
- practice: cool in a desiccator before weighing (USP <731>)
- drying to constant mass: two consecutive weighings, the second after further drying, within 0.5 mg (Ph. Eur. convention)
- USP <41>: minimum weight 2000 s; a net mass below it is not 'accurately weighed'
- ISO 13528:2022, 9.6 zeta scores and 9.4.2 their interpretation, as given in the Eurachem Guide on proficiency testing (EPT 2021:P3, 7.2.4): |zeta| <= 2 satisfactory, 2 < |zeta| < 3 a warning signal, |zeta| >= 3 an action signal
- Anton Paar DMA 35 (reseller specifications): density accuracy +/-0.001 g/cm3, repeatability s.d. 0.0005 g/cm3, resolution 0.0001 g/cm3, temperature accuracy 0.2 C
- ASME V&V 10-2019 and V&V 20-2009 (verification, then validation; E = S - D)
- ASME V&V 40-2018 (context of use, credibility factors)
- Shao, G.; Hightower, J.; Schindel, W. Credibility consideration for digital twins in manufacturing. Manufacturing Letters 2023, 35, 24-28
- JCGM 100:2008 (GUM), JCGM 101:2008 (Monte Carlo), JCGM 106:2012 (conformity assessment)
- ISO 13528:2022 (zeta scores)
- VialBot World Model: scientific foundations, present capabilities, partial replacement of physical experiments, and validation roadmap. A technical assessment of commit 5e1922b by a colleague, 24 October 2026 (the structure and the positioning of this report)
Assessment boundary: everything marked simulated, fitted, estimated, a placeholder or pending remains so; no physical measurement was made for this report.